Sl exam pt i

2,878
views

Published on


1 Comment
3 Likes
Statistics
Notes
No Downloads
Views
Total Views
2,878
On Slideshare
0
From Embeds
0
Number of Embeds
0
Actions
Shares
0
Downloads
235
Comments
1
Likes
3
Embeds 0
No embeds

Report content
Flagged as inappropriate Flag as inappropriate
Flag as inappropriate

Select your reason for flagging this presentation as inappropriate.

Cancel
No notes for slide

Transcript

  • 1. INSTRUCTION COURSE SLIT LAMP EXAMINATION MP STATE OPHTHALMIC CONFERENCE UJJAIN 2008
  • 2. Presented by
    • Dr. Sanjay Shrivastava
    • Dr. Nikhilesh Trivedi
    • Dr. Kavita Kumar
  • 3. Slit Lamp Examination
    • The science of examination with a slit lamp is called Biomicroscopy as it allows in vivo study of living tissues at high magnification.
  • 4. SLITLAMP EXAMINATION
    • Permits detailed examination of various structures of and posterior segment of eye
    • Angle of anterior chamber with gonioscope
    • Measurement of intra ocular pressure
    • Endothelial cell counts
  • 5. SLITLAMP EXAMINATION
    • Photographic representation of every above mentioned part of the eye
    • Modern slit lamp and its auxiliary devices provide magnified view of parts of eye from cornea to retina
    • Measurement of Corneal thickness is also possible
  • 6.
    • Slit lamp is a misnomer
    • In 1925 MAVAS introduced a term bio- microscopy defined as examination of the living eye by means of the corneal microscope and slit lamp.
    • LITTMAN in1950 introduced the rotatory magnification charger based on the principle of GALILEAN telescope is the forerunner of current ZEISS slit lamp .
  • 7.  
  • 8. PARTS
    • A. OBSERVTION SYSTEM
    • B. ILLUMINATION SYSTEM
    • C. MECHANICAL SYSTEM
  • 9. Slit lamp
  • 10. The Instrument
  • 11. The Instrument
  • 12.  
  • 13.  
  • 14.  
  • 15.  
  • 16.  
  • 17.  
  • 18.  
  • 19.  
  • 20. Posterior Segment Examination with +78 D Lens
  • 21. OBSERVATION SYSTEM
    • a. Objective lenses -22 diopters
    • b. Eye piece- 10 diopters
  • 22. ILLUMINATION SYSTEM
    • a. Light source
    • b. Condenser lens system
    • c. Slit and other diaphgram
    • d. Filters
    • e. Projection lens
    • f. Reflecting mirror/prism2.
  • 23. MECHANICAL SYSTEM
    • A. Joy stick arrangement
    • B. Up and down arrangement
    • C. Patient support arrangement
    • D. Mechanical coupling
  • 24. Slit Lamp Examination -Method
    • Prerequisites – Preferably a dark room and slit lamp biomicroscope
    • Position – Patient is seated comfortably on an adjustable stool, with chin resting against chin rest and forehead against head rest bar.
  • 25. PREPARATION
    • A. Preliminary examination of eye ball
    • B. Illumination of room.
    • C. Preparation and management of patients
    • D. Adjustment of instruments.
    • E. Stereopsis
  • 26. METHODS OF ILLUMINATION
    • Diffuse illumination
    • Sclerotic scatter
    • Direct focal illumination
    • Indirect illumination
    • Retro-illumination
    • Specular reflection
  • 27.
    • SET UP
    • 1. Angle between microscope and illuminating system – 30 deg to 45 deg. 2. Slit width - widest.
    • 3. Filters - diffusing filter
    • 4. Magnification - low to medium 5. Illumination – low to high.
  • 28. USE
    • 1. Topography of pathological changes.
    • 2. Entire surface of cornea, iris, lens can be viewed.
    • 3. Folds of descemets membrane.
    • 4. Corneal scar.
    • 5. Whole configuration of lens.
    • 6. Surface of adult nucleus.
  • 29. SCLEROTIC SCATTER
            • Also called indirect lateral illumination-it is possible by concentrating the beam upon the spot at the side of the object it is proposed to examine
            • Projecting the light upon sclera near the limbus,it is easy to examine the limbal portion of cornea,and if the incident ray is sufficient to induce total internal reflection,the whole of cornea including the opposite limbus ,is lit up.
  • 30. USE
    • 1. Corneal opacities
    • 2. Keratocele
    • 3. Interstitial deposits
    • 4. Perforating scars
  • 31. DIRECT FOCAL ILLUMINATION BASIS OF ALL METHODS.
  • 32. SET UP
    • Slit beam is regulated until it coincides with exact focus of microscope.
    • Light - narrow slit at an oblique angle 30 deg to 45 deg
  • 33. THREE SLIT BEAM EFFECTS
    • OPTICAL SECTION
    • PARALLELOPIPED CORNEAL SECTION
    • CONICAL BEAM EFFECT
  • 34.
    • I. OPTICAL SECTION
    • Produced by a very narrow slit beam focused obliquely.
    • It resemble a knife like histological section of tissue focused.
    • Corneal section uses
    • changes in corneal curvature.
    • change in corneal thickness.
    • depth of corneal pathologies; foreign body localization.
  • 35. OPTICAL SECTION Corneal optic section showing; Tear layer-anterior most bright zone Epithelial layer-dark line immediately behind it. Bowmen’s membrane-a bright line. Stroma-wider granular and greyer zone. Descemets membrane and endothelial layers-posterior most bright zone.
  • 36. II. PARELLOPIPED CORNEA
    • 2-3 mm. wide focused slit beam.
    • Because beam traverse cornea obliquely, shape of resulting rectangular block is paralellopiped prism. the ext. and internal surface of which slightly curved owing to anatomic shape of cornea
  • 37. USES 1. Pathologies of epithelium and Stroma. 2. Corneal scar or infiltrates. Appear brighter. 3. Water clefts have low density so appear black.
  • 38. III. CONICAL BEAM Small circular beam Used to detect the aqueous flare
  • 39. INDIRECT ILLUMINATION
        • Slit beam is focused at a point just behind a area to be examined.
  • 40. SET UP
    • Angle between slit beam and microscope is 30 deg to 45 deg
    • - Beam width - moderate
    • - Illumination - low medium or high
  • 41.
    • USES
      • Corneal infiltration
      • Corneal microcysts
      • Corneal vacuoles
      • Epithelial cells
  • 42. RETROILLUMINATION
    • Examination of normal or pathological structures in light reflected from tissue situated more posteriorly
  • 43. RETROILLUMINATION
  • 44.
    • Direct retro illumination
    • Observer in direct pathway of light
    • Light is reflected from structures so pathology seen against illuminated background
  • 45.
    • Indirect retro illumination
    • Observer is right angle to observed structures not in line so pathology seen against dark non illuminated background
  • 46.
    • Pathologies can be
    • 1. Obstructive – seen as dark against light background eg pigment or blood filled vessels
    • 2. Respersive –these scatter light but not obstruct completely seen as bright against dark background eg epithelial edema, precipitates.
  • 47. Pathologies can be
    • 3. Refractile – they distort the view of junction of illuminated and dark area because there refractive index is different from surroundings eg Vacuoles.
  • 48. SPECULAR REFLECTION
    • It is one of the most difficult methods. Not possible to use on all kind of slit lamps
    • Reflection of light occurs when beam of light is incident on an optical surface called zone of discontinuity, such zones may be found in cornea and lens
    • When observer is placed in pathway of reflected light a dazzling reflex seen called specular reflection
  • 49.
    • SET UP
    • Illuminating and viewing arms each forms and angle of about 30 deg to the central perpendicular
    • Slit beam height of 2 mm and focused onto the central corneal endothelium
    • Golden sheen with darker lines outline the hexagonal endothelial cells
    • Magnification of atleast 40X is needs to see endothelial pattern
  • 50.
    • USES
    • Surface pathologies scatter the light irregularly. therefore create dark area in reflex.
    • Using an eye piece reticule endothelial cells can be measured and counted and also used to study the tear film details .
  • 51. Oscillating illumination of Koppe
    • Slit beam is given an oscillatory movement by which it is often possible to see minute objects and filaments especially in the aqueous which would otherwise escapes detection
  • 52.
    • SPECIALISED USES
    1. Gonioscopy 2. Fundus examination with focal illumination 3. Pachymatery 4. Applanation tonometry 5. Ophthalmodynamometry
  • 53. SPECIALISED USES
    • 6. Slit lamp photography
    • 7. As a delivery system for argon , diode and Nd YAG laser
    • 8. Laser interferometry
    • 9. Potential acuity meter test
  • 54. END OF PART ONE THANKS

×