Proline Promag L 400-Electromagnetic Flowmeter

  1. 1. Application• Economical bidirectional measurement of liquids witha minimum conductivity of ³ 5 mS/cm in water/wastewater applications.• The electromagnetic measuring principle is unaffectedby pressure and temperature. Additionally the flowprofile has a minimal effect on the measurementresults.Device properties• Nominal diameter: DN 50 to 2400 (2 to 90")• Liner (HG, PU, PTFE) with worldwide drinking waterapprovals: KTW, WRAS, NSF, ACS• Process pressure: max. 16 bar (232 psi)• Durable polycarbonate transmitter housing• Graphical local display with operation from the outside(Touch Control)• Up to 3 outputs and 1 input including 4-20 mA HARTcommunicationYour benefitsThe versatile flowmeter for water, optimized for easyinstallationSizing - correct product selectionApplicator - the reliable, easy-to-use tool for selecting andsizing measuring devices for every applicationInstallation - simple and efficient• Economical compact design• Flexible mounting by lap joint flanges• Compact and remote version perfectly suited to meetthe requirements of the water/wastewater industryCommissioning - reliable and intuitive• Guided parameterization - "Make-it-run" wizards• Integrated web server for fast commissioningOperation - increased measurement availability• Simultaneous measurement: volume flow andconductivity• No pressure loss, no moving parts, immune tovibrations• Diagnostics; automatic data restore by HistoROMCost-effective life cycle management by W@MTechnical InformationProline Promag L 400Electromagnetic flow measuring systemThe no-frills flowmeter for water: Robust – economical – preciseTI01045D/06/EN/01.1271183813
  2. 2. Table of contents Proline Promag L 4002 Endress+HauserTable of contentsDocument information . . . . . . . . . . . . . . . . . . . . . . . . 3Symbols used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Function and system design . . . . . . . . . . . . . . . . . . . . 4Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Measured variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Operable flow range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Input signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Signal on alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Low flow cut off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Galvanic isolation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Protocol-specific data . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Terminal assignment . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Supply voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12Power consumption . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12Current consumption . . . . . . . . . . . . . . . . . . . . . . . . . . . 12Power supply failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . . 13Potential equalization . . . . . . . . . . . . . . . . . . . . . . . . . . . 15Terminals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17Cable entries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17Cable specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18Performance characteristics . . . . . . . . . . . . . . . . . . . 19Reference operating conditions . . . . . . . . . . . . . . . . . . . . . 19Maximum measured error . . . . . . . . . . . . . . . . . . . . . . . . 19Repeatability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20Influence of ambient temperature . . . . . . . . . . . . . . . . . . . 20Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20Mounting location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20Orientation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22Inlet and outlet runs . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22Adapters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23Length of connecting cable . . . . . . . . . . . . . . . . . . . . . . . . 23Installing the wall-mount housing . . . . . . . . . . . . . . . . . . . 24Special mounting instructions . . . . . . . . . . . . . . . . . . . . . . 25Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25Ambient temperature range . . . . . . . . . . . . . . . . . . . . . . . 25Storage temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25Atmosphere . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26Degree of protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26Shock resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26Vibration resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26Mechanical load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26Electromagnetic compatibility (EMC) . . . . . . . . . . . . . . . . . 26Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26Medium temperature range . . . . . . . . . . . . . . . . . . . . . . . 26Conductivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26Pressure-temperature ratings . . . . . . . . . . . . . . . . . . . . . . 26Pressure tightness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29Flow limit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30Pressure loss . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30System pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30Vibrations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30Mechanical construction . . . . . . . . . . . . . . . . . . . . . 31Design, dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41Measuring tube specification . . . . . . . . . . . . . . . . . . . . . . . 46Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48Fitted electrodes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49Process connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50Surface roughness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50Operability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50Operating concept . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50Local operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51Remote operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51Certificates and approvals . . . . . . . . . . . . . . . . . . . . 52CE mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52C-Tick symbol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52Ex approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52Drinking water approval . . . . . . . . . . . . . . . . . . . . . . . . . 52Other standards and guidelines . . . . . . . . . . . . . . . . . . . . . 52Ordering information . . . . . . . . . . . . . . . . . . . . . . . . 53Application packages . . . . . . . . . . . . . . . . . . . . . . . 53Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53Diagnostics functions . . . . . . . . . . . . . . . . . . . . . . . . . . . 53Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54Device-specific accessories . . . . . . . . . . . . . . . . . . . . . . . . 54Communication-specific accessories . . . . . . . . . . . . . . . . . . 54Service-specific accessories . . . . . . . . . . . . . . . . . . . . . . . . 55System components . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56Standard documentation . . . . . . . . . . . . . . . . . . . . . . . . . 56Supplementary device-dependent documentation . . . . . . . . . . 56Registered trademarks . . . . . . . . . . . . . . . . . . . . . . . 56
  3. 3. Proline Promag L 400Endress+Hauser 3Document informationSymbols used Electrical symbolsSymbol MeaningA0011197Direct currentA terminal to which DC voltage is applied or through which direct current flows.A0011198Alternating currentA terminal to which alternating voltage is applied or through which alternating current flows.A0017381Direct current and alternating current• A terminal to which alternating voltage or DC voltage is applied.• A terminal through which alternating current or direct current flows.)A0011200Ground connectionA grounded terminal which, as far as the operator is concerned, is grounded via a grounding system.*A0011199Protective ground connectionA terminal which must be connected to ground prior to establishing any other connections.A0011201Equipotential connectionA connection that has to be connected to the plant grounding system: This may be a potential equalizationline or a star grounding system depending on national or company codes of practice.Symbols for certain types of informationSymbol MeaningA0011182AllowedIndicates procedures, processes or actions that are allowed.A0011183PreferredIndicates procedures, processes or actions that are preferred.A0011184ForbiddenIndicates procedures, processes or actions that are forbidden.A0011193TipIndicates additional information.A0011194Reference to documentationRefers to the corresponding device documentation.A0011195Reference to pageRefers to the corresponding page number.A0011196Reference to graphicRefers to the corresponding graphic number and page number.Symbols in graphicsSymbol Meaning1, 2, 3,... Item numbers, …, Series of stepsA, B, C, ... ViewsA-A, B-B, C-C, ... SectionsA0013441Flow direction
  4. 4. Proline Promag L 4004 Endress+HauserSymbol Meaning-A0011187Hazardous areaIndicates a hazardous area..A0011188Safe area (non-hazardous area)Indicates a non-hazardous area.Function and system designMeasuring principle Following Faradays law of magnetic induction, a voltage is induced in a conductor moving through a magneticfield.In the electromagnetic measuring principle, the flowing medium is the moving conductor. The voltage inducedis proportional to the flow velocity and is supplied to the amplifier by means of two measuring electrodes. Theflow volume is calculated via the pipe cross-sectional area. The DC magnetic field is created through a switcheddirect current of alternating polarity.ILBIUeA0017035å 1 Ue = B · L · v ; Q = A · vUe Induced voltageB Magnetic induction (magnetic field)L Electrode spacingv Flow velocityQ Volume flowA Piping cross-sectionI Current
  5. 5. Proline Promag L 400Endress+Hauser 5Measuring system The device consists of a transmitter and a sensor.Two device versions are available:• Compact version – the transmitter and sensor form a mechanical unit.• Remote version – the transmitter and sensor are mounted separately from one another.TransmitterPromag 400 Device versions and materials:• Compact housing: polycarbonate plastic• Wall-mount housing: polycarbonate plasticConfiguration:• External operation via four-line, illuminated local display with touch controland guided menus ("Make-it-run" wizards) for applications• Via operating tools (e.g. FieldCare)• Via Web browser (e.g. Microsoft Internet Explorer)A0017117SensorPromag L Nominal diameter range: DN 50 to 2400 (2 to 90")Materials:• Sensor housing: aluminum coated AlSi10Mg, carbon steel with protectivevarnish• Sensor connection housing: aluminum coated AlSi10Mg• Measuring tube: stainless steel 1.4301/304, 1.4306/304L, 202, 304• Liner: hard rubber, polyurethane, PTFE• Seals: as per DIN EN 1514-1• Electrodes: 1.4435/304L, Alloy C-22• Ground disks: 1.4435/316L, Alloy C-22Lap joint flange; lap joint flange,stamped plate: DN 50 to 300 (2 to 12")A0017040Fixed flange: DN 350 to 2400 (14 to90")A0017041InputMeasured variable Direct measured variables• Volume flow (proportional to induced voltage)• Electrical conductivityCalculated measured variablesMass flow
  6. 6. Proline Promag L 4006 Endress+HauserMeasuring range Typically v = 0.01 to 10 m/s (0.03 to 33 ft/s) with the specified accuracyFlow characteristic values in SI unitsNominaldiameterRecommendedflowFactory settingsmin./max. full scale value(v ~ 0.3/10 m/sFull scale value currentoutput(v ~ 2.5 m/s)Pulse value(~ 2 pulse/s)Low flow cut off(v ~ 0.04 m/s)[mm] [in] [m3/h] [m3/h] [m3] [m3/h]50 2 35 to 1100 dm3/min 300 dm3/min 2.5 dm3 5 dm3/min65 – 60 to 2000 dm3/min 500 dm3/min 5 dm3 8 dm3/min80 3 90 to 3000 dm3/min 750 dm3/min 5 dm3 12 dm3/min100 4 145 to 4700 dm3/min 1200 dm3/min 10 dm3 20 dm3/min125 – 220 to 7500 dm3/min 1850 dm3/min 15 dm3 30 dm3/min150 6 20 to 600 150 0.025 2.5200 8 35 to 1100 300 0.05 5250 10 55 to 1700 500 0.05 7.5300 12 80 to 2400 750 0.1 10350 14 110 to 3300 1000 0.1 15375 15 140 to 4200 1200 0.15 20400 16 140 to 4200 1200 0.15 20450 18 180 to 5400 1500 0.25 25500 20 220 to 6600 2000 0.25 30600 24 310 to 9600 2500 0.3 40700 28 420 to 13500 3500 0.5 50750 30 480 to 15000 4000 0.5 60800 32 550 to 18000 4500 0.75 75900 36 690 to 22500 6000 0.75 1001000 40 850 to 28000 7000 1 125– 42 950 to 30000 8000 1 1251200 48 1250 to 40000 10000 1.5 150– 54 1550 to 50000 13000 1.5 2001400 – 1700 to 55000 14000 2 225– 60 1950 to 60000 16000 2 2501600 – 2200 to 70000 18000 2.5 300– 66 2500 to 80000 20500 2.5 3251800 72 2850 to 90000 23000 3 350– 78 3300 to 100000 28500 3.5 4502000 – 3400 to 110000 28500 3.5 450– 84 3700 to 125000 31000 4.5 5002200 – 4100 to 136000 34000 4.5 540– 90 4300 to 143000 36000 5 5702400 – 4800 to 162000 40000 5.5 650
  7. 7. Proline Promag L 400Endress+Hauser 7Flow characteristic values in US unitsNominaldiameterRecommendedflowFactory settingsmin./max. full scale value(v ~ 0.3/10 m/sFull scale value currentoutput(v ~ 2.5 m/s)Pulse value(~ 2 pulse/s)Low flow cut off(v ~ 0.04 m/s)[in] [mm] [gal/min] [gal/min] [gal] [gal/min]2 50 10 to 300 75 0.5 1.25– 65 16 to 500 130 1 23 80 24 to 800 200 2 2.54 100 40 to 1250 300 2 4– 125 60 to 1950 450 5 76 150 90 to 2650 600 5 128 200 155 to 4850 1200 10 1510 250 250 to 7500 1500 15 3012 300 350 to 10600 2400 25 4514 350 500 to 15000 3600 30 6015 375 600 to 19000 4800 50 6016 400 600 to 19000 4800 50 6018 450 800 to 24000 6000 50 9020 500 1000 to 30000 7500 75 12024 600 1400 to 44000 10500 100 18028 700 1900 to 60000 13500 125 21030 750 2150 to 67000 16500 150 27032 800 2450 to 80000 19500 200 30036 900 3100 to 100000 24000 225 36040 1000 3800 to 125000 30000 250 48042 – 4200 to 135000 33000 250 60048 1200 5500 to 175000 42000 400 60054 – 9 to 300 Mgal/day 75 Mgal/day 0.0005 Mgal 1.3 Mgal/day– 1400 10 to 340 Mgal/day 85 Mgal/day 0.0005 Mgal 1.3 Mgal/day60 – 12 to 380 Mgal/day 95 Mgal/day 0.0005 Mgal 1.3 Mgal/day– 1600 13 to 450 Mgal/day 110 Mgal/day 0.0008 Mgal 1.7 Mgal/day66 – 14 to 500 Mgal/day 120 Mgal/day 0.0008 Mgal 2.2 Mgal/day72 1800 16 to 570 Mgal/day 140 Mgal/day 0.0008 Mgal 2.6 Mgal/day78 – 18 to 650 Mgal/day 175 Mgal/day 0.0010 Mgal 3.0 Mgal/day– 2000 20 to 700 Mgal/day 175 Mgal/day 0.0010 Mgal 2.9 Mgal/day84 – 24 to 800 Mgal/day 190 Mgal/day 0.0011 Mgal 3.2 Mgal/day– 2200 26 to 870 Mgal/day 210 Mgal/day 0.0012 Mgal 3.4 Mgal/day90 – 27 to 910 Mgal/day 220 Mgal/day 0.0013 Mgal 3.6 Mgal/day– 2400 31 to 1030 Mgal/day 245 Mgal/day 0.0014 Mgal 4.1 Mgal/dayTo calculate the measuring range, use the Applicator sizing tool (® ä 55)
  8. 8. Proline Promag L 4008 Endress+HauserRecommended measuring range"Flow limit" section (® ä 30)Operable flow range Over 1000 : 1Input signal Status inputMaximum input values • DC 30 V• 6 mAResponse time Adjustable: 5 to 200 msInput signal level • Low signal: DC –3 to +5 V• High signal: DC 12 to 30 VAssignable functions • Off• Reset totalizers 1-3 separately• Reset all totalizers• Flow overrideOutputOutput signal Current outputCurrent output Can be set as:• 4-20 mA HART (active)• 0-20 mA (active)Maximum output values • DC 24 V (when idle)• 22.5 mALoad 0 to 700 ΩResolution 0.5 µADamping Adjustable: 0 to 999 sAssignable measuredvariables• Volume flow• Conductivity• Mass flowPulse/frequency/switch outputFunction • With the order code for "Output; Input", option H: output 2 can be set as a pulse orfrequency output• With the order code for "Output; Input", option I: output 2 and 3 can be set as a pulse,frequency or switch outputVersion Passive, open collectorMaximum input values • DC 30 V• 250 mAVoltage drop For 25 mA: £ DC 2 VPulse outputPulse width Adjustable:0.05 to 2000 msMaximum pulse rate 10000 Impulse/sPulse value AdjustableAssignable measuredvariables• Volume flow• Mass flowFrequency outputOutput frequency Adjustable: 0 to 10000 Hz
  9. 9. Proline Promag L 400Endress+Hauser 9Damping Adjustable: 0 to 999 sPulse/pause ratio 1:1Assignable measuredvariables• Volume flow• Conductivity• Mass flowSwitch outputSwitching behavior Binary, conductive or non-conductiveSwitching delay Adjustable: 0 to 100 sNumber of switching cycles UnlimitedAssignable functions • Off• On• Diagnostic behavior• Limit value:– Volume flow– Conductivity– Mass flow– Totalizer 1-3• Flow direction monitoring• Status– Empty pipe detection– Low flow cut offSignal on alarm Depending on the interface, failure information is displayed as follows:Current output4-20 mAFailure mode Selectable (as per NAMUR recommendation NE 43):• Minimum alarm: 3.6 mA• Maximum alarm: 22 mA• Adjustable value: 3.59 to 22.5 mA0-20 mAFailure mode Choose from:• Maximum alarm: 22 mA• Adjustable value: 0 to 22.5 mAHARTDevice diagnostics Device condition can be read out via HART Command 48Pulse/frequency/switch outputPulse outputFailure mode Choose from:• Actual value• No pulsesFrequency outputFailure mode Choose from:• Actual value• Defined value: 0 to 12500 Hz• 0 Hz
  10. 10. Proline Promag L 40010 Endress+HauserSwitch outputFailure mode Choose from:• Current status• Open• ClosedLocal displayPlain text display With information on cause and remedial measuresBacklight Red backlighting indicates a device error.Status signal as per NAMUR recommendation NE 107Operating tool• Via digital communication: HART protocol• Via service interfacePlain text display With information on cause and remedial measuresAdditional information on remote operation (® ä 51)Web browserPlain text display With information on cause and remedial measuresLow flow cut off The switch points for low flow cut off are user-selectable.Galvanic isolation The following connections are galvanically isolated from each other:• Inputs• Outputs• Voltage supplyProtocol-specific data HARTManufacturer ID 0x11Device type ID 0x47HART protocol revision 6.0Device description files (DTM,DD)Information and files under:www.endress.comHART load Min. 250 ΩDynamic variables The measured variables can be freely assigned to the dynamic variables.Measured variables for PV (primary dynamic variable)• Volume flow• Conductivity• Mass flowMeasured variables for SV, TV, QV (secondary, tertiary and quaternarydynamic variable)• Volume flow• Conductivity• Mass flow• Totalizer 1• Totalizer 2• Totalizer 3
  11. 11. Proline Promag L 400Endress+Hauser 11Power supplyTerminal assignment Transmitter0-20 mA/4-20 mA HART connection version with additional outputs and inputs20 2122 2324 2526 271 2L+/LL-/N+ +- -+ +- -A0017112Supply voltageOrder code for"Power supply"Terminal numbers1 (L+/L) 2 (L-/N)Option A AC 100 to 230 VOption B AC/DC 24 VSignal transmissionOrder code for"Output; input"Terminal numbersOutput 1 Output 2 Output 3 Input 426 (+) 27 (-) 24 (+) 25 (-) 22 (+) 23 (-) 20 (+) 21 (-)Option H • 4-20 mA HART(active)• 0-20 mA (active)Pulse/frequencyoutput (passive)switch output (passive) -Option I • 4-20 mA HART(active)• 0-20 mA (active)Pulse/frequency/switch output (passive)Pulse/frequency/switch output (passive)Status input
  12. 12. Proline Promag L 40012 Endress+HauserRemote versionE1E2GNDES1E1E2S2GNDES5 7 4 37 42 416 5 7 8 4 37 36n.c. n.c.dcc dbn.c.42 41aA0017275å 2 Connecting the remote versiona Transmitter: main electronics module with terminalsb Sensor: connection modulec Electrode cabled Coil current cablen.c. Not connected, insulated cable shieldsTerminal No. and cable colors: 6/5 = brown; 7/8 = white; 4 = green; 36/37 = yellowSupply voltage Order code for"Power supply"Terminal voltage Frequency rangeOption A: AC 100 to 230 V AC 85 to 260 V 47 to 63 HzOption B: AC/DC 24 V AC/DC 18 to 30 V 44 to 66 Hz/-Power consumption TransmitterOrder code for"Power supply"Maximum power consumptionOption A: AC 100 to 230 V 30 VAOption B: AC/DC 24 V 30 VA/8 WCurrent consumption TransmitterOrder code for"Power supply"Maximum currentconsumptionMaximum switch-oncurrentOption A: AC 100 to 230 V 145 mA 25 A (< 5 ms)Option B: AC/DC 24 V 350 mA 27 A (< 5 ms)Power supply failure • Totalizers stop at the last value measured.• Configuration is retained in the plug-in memory (HistoROM DAT).• Error messages (incl. total operated hours) are stored.
  13. 13. Proline Promag L 400Endress+Hauser 13Electrical connection Connecting the transmitter2 2 1 4 3 2 2 1A BA0017113A Compact versionB Remote version1 Cable entry for supply voltage2 Cable entry for signal transmission3 Cable entry for coil current cable4 Cable entry for electrode cableRemote version connection1 2A00172671 Electrode cable2 Coil current cable• Fix the cable run or route it in an armored conduit.Cable movements can influence the measuring signal especially in the case of low fluid conductivities.• Route the cable well clear of electrical machines and switching elements.• Ensure potential equalization between sensor and transmitter (® ä 15).Connection examples0/4...20 mA+–213+_A0017162å 3 Connection example for 0-20 mA current output (active) and 4-20 mA current output (active)1 Automation system with current input (e.g. PLC)2 Analog display unit: observe maximum load (® ä 8)3 Transmitter
  14. 14. Proline Promag L 40014 Endress+Hauser44...20 mA+–521 36+_A0016800å 4 Connection example for 4-20 mA HART current output (active)1 Automation system with current input (e.g. PLC)2 Observe cable specification (® ä 18)3 Connection for HART operating devices (® ä 51)4 Resistor for HART communication (³ 250 W): observe maximum load (® ä 8)5 Analog display unit: observe maximum load (® ä 8)6 Transmitter1+_123452+–+–3A0016801å 5 Connection example for pulse/frequency output (passive)1 Automation system with pulse/frequency input (e.g. PLC)2 Voltage supply3 Transmitter: observe input values (® ä 8)1+_+_2+_ 3A0016802å 6 Connection example for switch output (passive)1 Automation system with switch input (e.g. PLC)2 Voltage supply3 Transmitter: observe input values (® ä 8)
  15. 15. Proline Promag L 400Endress+Hauser 151+ _+_2+_3A0017163å 7 Connection example for status input1 Automation system with status output (e.g. PLC)2 Voltage supply3 Transmitter: observe input values (® ä 8)Potential equalization RequirementsPlease consider the following to ensure correct measurement:• Same electrical potential for the fluid and sensor• Company-internal grounding concepts• Pipe material and groundingConnection examples for standard situationsMetal, grounded pipeA0016315å 8 Potential equalization via measuring tubeConnection example in special situationsUnlined and ungrounded metal pipeThis connection method also applies in situations where:• The customary potential equalization is not used• Equalizing currents are present
  16. 16. Proline Promag L 40016 Endress+HauserDN 300£ DN 350³A0016317å 9 Potential equalization via ground terminal and pipe flangesNote the following when installing:• Connect both sensor flanges to the pipe flange via a ground cable and ground them.• Connect the connection housing of the transmitter or sensor to ground potential by means of the groundterminal provided for the purpose.To mount the ground cable:– If DN £ 300 (12"): Mount the ground cable directly on the conductive flange coating of the sensor withthe flange screws.– If DN ³ 350 (14"): Mount the ground cable directly on the metal transport bracket.Ground cable Copper wire, at least 6 mm2 (0.0093 in2)For remote device versions, the ground terminal in the example always refers to the sensor and not tothe transmitter.The necessary ground cable can be ordered from Endress+Hauser (® ä 54).Plastic pipe or pipe with insulating linerThis connection method also applies in situations where:• The customary potential equalization is not used• Equalizing currents are presentA0016318å 10 Potential equalization via ground terminal and ground disksNote the following when installing:The ground disks must be connected to the ground terminal via the ground cable and be connected to groundpotential.Ground cable Copper wire, at least 6 mm2 (0.0093 in2)For remote device versions, the ground terminal in the example always refers to the sensor and not tothe transmitter.
  17. 17. Proline Promag L 400Endress+Hauser 17The ground cable and ground disks can be ordered from Endress+Hauser (® ä 54).Pipe with a cathodic protection unitThis connection method is only used if the following two conditions are met:• Metal pipe without liner or pipe with electrically conductive liner• Cathodic protection is integrated in the personal protection equipment1 /+–2 /2 /3 /A0016319å 11 Potential equalization and cathodic protection1 Isolation transformer power supply2 Electrically isolated from the pipe3 CapacitorGround cable Copper wire, at least 6 mm2 (0.0093 in2)1. Connect the measuring device to the power supply such that it is floating in relation to the protectiveground.2. Install the sensor in the pipe in a way that provides electrical insulation.3. Connect the two flanges of the pipe to one another via a ground cable.4. Guide the shield of the signal lines through a capacitor.For remote device versions, the ground terminal in the example always refers to the sensor and not tothe transmitter.The necessary ground cable can be ordered from Endress+Hauser (® ä 54).Terminals Transmitter• Supply voltage cable: plug-in screw terminals for wire cross-sections 0.5 to 2.5 mm2 (20 to 14 AWG)• Signal cable: plug-in screw terminals for wire cross-sections 0.5 to 2.5 mm2 (20 to 14 AWG)• Electrode cable: plug-in screw terminals for wire cross-sections 0.5 to 2.5 mm2 (20 to 14 AWG)• Coil current cable: screw terminals for wire cross-sections 0.5 to 2.5 mm2 (20 to 14 AWG)SensorScrew terminals for wire cross-sections 0.5 to 2.5 mm2 (20 to 14 AWG)Cable entries Transmitter and sensor• Cable gland:– For standard cable: M20 × 1.5 with cable Æ 6 to 12 mm (0.24 to 0.47 in)– For reinforced cable: M20 × 1.5 with cable Æ 9.5 to 16 mm (0.37 to 0.63 in)• Thread for cable entry:– NPT ½"– G ½"– M20If metal cable entries are used, use a grounding plate.
  18. 18. Proline Promag L 40018 Endress+HauserCable specification Permitted temperature range• –40 °C (–40 °F)...³ 80 °C (176 °F)• Minimum requirement: cable temperature range ³ ambient temperature + 20 KPower supply cableStandard installation cable is sufficient.Signal cableCurrent output• For 0-20 mA and 4-20 mA: standard installation cable is sufficient.• For 4-20 mA HART: Shielded cable recommended. Observe grounding concept of the plant.Pulse/frequency/switch outputStandard installation cable is sufficient.Status inputStandard installation cable is sufficient.Connecting cable for remote versionElectrode cableStandard cable 3 ´ 0.38 mm2 (20 AWG) with common, braided copper shield (Æ ~ 7 mm (0.28")) andindividually shielded coresCable for empty pipedetection (EPD)4 ´ 0.38 mm2 (20 AWG) with common, braided copper shield (Æ ~ 7 mm (0.28")) andindividually shielded coresConductor resistance £ 50 W/km (0.015 W/ft)Capacitance: core/shield £ 420 pF/m (128 pF/ft)Operating temperature –20 to +80 °C (–68 to +176 °F)Coil current cableStandard cable 2 ´ 0.75 mm2 (18 AWG) with common, braided copper shield (Æ ~ 7 mm (0.28")) andindividually shielded coresConductor resistance £ 37 W/km (0.011 W/ft)Capacitance: core/core,shield grounded£ 120 pF/m (37 pF/ft)Operating temperature –20 to +80 °C (–68 to +176 °F)Test voltage for cableinsulation£ AC 1433 V r.m.s. 50/60 Hz or ³ DC 2026 V
  19. 19. Proline Promag L 400Endress+Hauser 191234567a bA0003194å 12 Cable cross-sectiona Electrode cableb Coil current cable1 Core2 Core insulation3 Core shield4 Core jacket5 Core reinforcement6 Cable shield7 Outer jacketReinforced connecting cableUse in the following situations:• When laying the cable directly in the ground• Where there is a risk of damage from rodentsThe reinforced connecting cable with an additional, reinforcing metal braid can be ordered from Endress+Hauser (® ä 54).Operation in zones of severe electrical interferenceThe measuring system meets the general safety requirements (® ä 52) and EMC specifications(® ä 26).Grounding is by means of the ground terminal provided for the purpose inside the connection housing. Thestripped and twisted lengths of cable shield to the ground terminal must be as short as possible.Performance characteristicsReference operatingconditionsTo DIN EN 29104• Fluid temperature: +28 ± 2 °C (+82 ± 4 °F)• Ambient temperature range: +22 ± 2 °C (+72 ± 4 °F)• Warm-up period: 30 minInstallation• Inlet run > 10 ´ DN• Outlet run > 5 ´ DN• Sensor and transmitter grounded.• The sensor is centered in the pipe.To calculate the measuring range, use the Applicator sizing tool (® ä 55)Maximum measured error Accuracy of outputso.r. = of reading; o.f.s. = of full scale valueThe outputs have the following base accuracy specifications.Current outputAccuracy Max. ±0.05 % o.f.s. or ±5 µA
  20. 20. Proline Promag L 40020 Endress+HauserPulse/frequency outputAccuracy Max. ±50 ppm o.r.Error limits under reference operating conditionso.r. = of readingPulse output• ±0.5 % o.r.± 1 mm/s (0.04 in/s)• Optional: ±0.2 % o.r. ± 2 mm/s (0.08 in/s)Fluctuations in the supply voltage do not have any effect within the specified range.2.5[%]2.01.51.00.500.2 %0.5 %0 1 2 4 6 8 10 [m/s]v5 10 15 20 25 30 32 [ft/s]0A0005531å 13 Maximum measured error in % o.r.Repeatability o.r. = of readingmax. ±0.1 % o.r. ± 0.5 mm/s (0.02 in/s)Influence of ambienttemperatureo.r. = of reading; o.f.s. = of full scale valueCurrent outputTemperature coefficient Typically ±50 ppm/°C o.r. or ±1 µA/°CPulse/frequency outputTemperature coefficient Max.±50 ppm/°CInstallationNo special measures such as supports are necessary. External forces are absorbed by the construction of thedevice.Mounting location Preferably install the sensor in an ascending pipe, and ensure a sufficient distance to the next pipe elbow: h =³ 2 × DN
  21. 21. Proline Promag L 400Endress+Hauser 21hA0017061To prevent measuring errors arising from accumulation of gas bubbles in the measuring tube, avoid the followingmounting locations in the pipe:• Highest point of a pipeline.• Directly upstream of a free pipe outlet in a down pipe.Installation in down pipesInstall a siphon or a vent valve downstream of the sensor in down pipes whose length h ³ 5 m (16.4 ft).Thisprecaution is to avoid low pressure and the consequent risk of damage to the measuring tube. This measurealso prevents the system losing prime, which could cause air pockets.For information on the liners resistance to partial vacuum (® ä 29)h21A0017064å 14 Installation in a down pipe1 Vent valve2 Pipe siphonh Length of down pipeInstallation in partially filled pipesA partially filled pipe with a gradient necessitates a drain-type configuration. The empty pipe detection (EPD)function offers additional protection by detecting empty or partially filled pipes.• Do not install the sensor at the lowest point in the drain: risk of solids accumulating.• It is advisable to install a cleaning valve.³ 5 × DN³ 2 × DNA0017063For very heavy sensorsIf the nominal diameter DN ³ 350 (14 in), mount the sensor on a foundation of adequate load-bearing strength.
  22. 22. Proline Promag L 40022 Endress+HauserDo not support the weight of the sensor on the metal casing as this could damage the metal casing and theinternal magnetic coils.A0016276Orientation The direction of the arrow on the sensor nameplate helps you to install the sensor according to the flow direction(direction of medium flow through the piping).An optimum orientation position helps avoid gas and air accumulations and deposits in the measuring tube.The measuring device also offers the empty pipe detection function to detect:• Partially empty measuring pipes• Outgassing fluids• Applications with variable process pressuresVerticalThis is the optimum orientation for self-emptying piping systems and for use in conjunction with empty pipedetection.A0015591HorizontalThe measuring electrode plane must be horizontal. This prevents brief insulation of the two measuringelectrodes by entrained air bubbles.With horizontal orientation, empty pipe detection only works if the transmitter housing is pointingupwards as otherwise there is no guarantee that the empty pipe detection function will actually respondto a partially filled or empty measuring tube.AA1232A0016260å 15 Horizontal orientation1 EPD electrode for empty pipe detection2 Measuring electrodes for signal detection3 Reference electrode for potential equalizationInlet and outlet runs If possible, install the sensor upstream from fittings such as valves, T-pieces or elbows.
  23. 23. Proline Promag L 400Endress+Hauser 23Observe the following inlet and outlet runs to comply with accuracy specifications:• Inlet run ³ 5 ´ DN• Outlet run ³ 2 ´ DN5 × DN³ ³ 2 × DNA0016275Adapters Suitable adapters to DIN EN 545 (double-flange reducers) can be used to install the sensor in larger-diameterpipes. The resultant increase in the rate of flow improves measuring accuracy with very slow-moving fluids.The nomogram shown here can be used to calculate the pressure loss caused by reducers and expanders:• Calculate the ratio of the diameters d/D.• From the nomogram read off the pressure loss as a function of flow velocity (downstream from the reduction)and the d/D ratio.The nomogram only applies to liquids with a viscosity similar to that of water.100100.5d / D[mbar]0.6 0.7 0.8 0.91 m/s2 m/s3 m/s4 m/s5 m/s6 m/s7 m/s8 m/s1Ddmax. 8°A0016359Length of connecting cable To ensure correct measuring results when using the remote version,observe the maximum permitted cable length Lmax. This length is determined by the conductivity of the fluid.– If measuring liquids in general: 5 mS/cm– If measuring demineralized water: 20 mS/cm
  24. 24. Proline Promag L 40024 Endress+Hauser200100510 100 200L max[m][m][µS/cm]L max[ft]200 6000 400A0016539å 16 Permitted length of connecting cable for remote versionArea shaded gray = permitted rangeLmax = length of connecting cable in [m] ([ft])[mS/cm] = fluid conductivityInstalling the wall-mounthousingWall mounting149 (5.85)210.5(8.29)=5.8 (0.23)mm (in)18 (0.71) =14 (0.55)5.8 (0.23)A0017138
  25. 25. Proline Promag L 400Endress+Hauser 25Pipe mountingø 20…70( 0.79…2.75)ø~~ 4.0)102 (mm (in)4 x4 x SW 8PH 2A0017135A separate mounting kit can be ordered from Endress+Hauser for pipe mounting (® ä 54).Special mounting instructions Weather protection coverTo ensure that the optional weather protection cover can be easily opened, maintain the following minimumhead clearance: 350 mm (13.8 in)EnvironmentAmbient temperature range Transmitter –20 to +50 °C (–4 to +122 °F)Local display The readability of the display may be impaired at ambient temperatures below–20 °C (–4 °F).Sensor • Flange material carbon steel: –10 to +60 °C (+14 to +140 °F)• Flange material stainless steel: –40 to +60 °C (–40 to +140 °F)Mount the transmitter separately from the sensor if both the ambient and fluidtemperatures are high.Liner Do not exceed or fall below the permitted temperature range of the liner(® ä 26).If operating outdoors:• Install the measuring device in a shady location.• Avoid direct sunlight, particularly in warm climatic regions.• Avoid direct exposure to weather conditions.Weather protection covers can be ordered from Endress+Hauser: see "Accessories" section (® ä 54)Storage temperature The storage temperature corresponds to the operating temperature range of the measuring transmitter and theappropriate measuring sensors.
  26. 26. Proline Promag L 40026 Endress+Hauser• Protect the measuring device against direct sunlight during storage in order to avoid unacceptably high surfacetemperatures.• Select a storage location where moisture cannot collect in the measuring device as fungus or bacteriainfestation can damage the liner.• If protection caps or protective covers are mounted these should never be removed before installing themeasuring device.Atmosphere If a plastic transmitter housing is permanently exposed to certain steam and air mixtures, this can damage thehousing.If you are unsure, please contact your Endress+Hauser Sales Center for clarification.Degree of protection Transmitter• As standard: IP66/67, type 4X enclosure• When housing is open: IP20, type 1 enclosureSensor• As standard: IP66/67, type 4X enclosure• Optionally available for remote version:IP68, type 6P enclosure (for DN £ 300 (12") only possible in conjunction with stainless steel flanges)Not suitable for use in corrosive atmospheres/liquids or in buried applications if special precautions are nottaken.Shock resistance Acceleration up to 2 g following IEC 600 68-2-6Vibration resistance Acceleration up to 2 g following IEC 600 68-2-6Mechanical load • Protect the transmitter housing against mechanical effects, such as shock or impact; the use of the remoteversion is sometimes preferable.• Never use the transmitter housing as a ladder or climbing aid.Electromagnetic compatibility(EMC)• As per IEC/EN 61326 and NAMUR Recommendation 21 (NE 21)• Complies with emission limits for industry as per EN 55011 (Class A)Details are provided in the Declaration of Conformity.ProcessMedium temperature range • 0 to +80 °C (+32 to +176 °F) for hard rubber, DN 350 to 2400 (14 to 90")• –20 to +50 °C (–4 to +122 °F) for polyurethane, DN 50 to 1200 (2 to 48")• –20 to +90 °C (–4 to +194 °F) for PTFE, DN 50 to 300 (2 to 12")Conductivity • ³ 5 mS/cm for liquids in general• ³ 20 mS/cm for demineralized waterNote that in the case of the remote version, the requisite minimum conductivity also depends on the cablelength (® ä 23).Pressure-temperature ratings The following material load diagrams refer to the entire device and not just the process connection.
  27. 27. Proline Promag L 400Endress+Hauser 27Flange connection according to EN 1092-1 (DIN 2501)PN1605101520[bar][psi]-60 -40 -20 0 20 40 60 80 100 120 140 [°C][°F]0-40 100 200 3002001003000PN10A0011571å 17 Lap joint flange PN 10/16, materials 1.4306/304L and 1.4307/304L, DN 50 to 300 (2 to 12")Flange connection according to EN 1092-1 (DIN 2501)PN1005101520[bar][psi]-60 -40 -20 0 20 40 60 80 100 120 140 [°C][°F]0-40 100 200 3002001003000A0011573å 18 Lap joint flange PN 10, material 1.4301/304, DN 50 to 300 (2 to 12")Flange connection according to EN 1092-1 (DIN 2501)PN1605101520[bar][psi]-40 -20 0 20 40 60 80 100 120 140 [°C]0-40 100 200 3002001003000[°F]PN10PN6A0011568å 19 Fixed flange PN 6/10, materials 1.0038 (S235JRG2) and A105, DN 350 to 1000 (14 to 40") / DN 1200 to 2400(48 to 90"); lap joint flange PN 16, material 1.0038 (S235JRG2), DN 50 to 150 (2 to 6")
  28. 28. Proline Promag L 40028 Endress+HauserFlange connection according to ASME B16.5-40 [°C]-20 0 20 40 60 80 100 120 1400-40 100 200 300 [°F]0102030[bar][psi]2001004003005000DN50…250(2"…10")DN300(12")A0011572å 20 Lap joint flange Class 150, material A105, DN 50 to 300 (2 to 12")Flange connection according to ASME B16.50102030[bar][psi]-40 -20 0 20 40 60 80 100 120 140 [°C][°F]0-40 100 200 3002001004003000DN50…250(2"…10")DN300(12")A0011580å 21 Lap joint flange Class 150, material 316L, DN 50 to 300 (2 to 12")Flange connection according to ASME B16.5Class150010203040[bar][psi]-40 -20 0 20 40 60 80 100 120 140 [°C][°F]0-40 100 200 3002001004003005000A0017264å 22 Fixed flange Class 150, material A105, DN 50 to 300 (2 to 12")
  29. 29. Proline Promag L 400Endress+Hauser 29Flange connection according to AWWA C2070891011[bar][psi]-40 -20 0 20 40 60 80 100 120 140 [°C]0-40 100 200 3001451161600[°F]130A0017115å 23 Fixed flange Class D, materials 1.0044 (S275JR), 1.0425/316L (P265GH), A105 and A181, DN 1200 to 2200(48 to 90")Flange connection according to AS 2129 and AS 4087-40 -20 0 20 40 60 80 100 120 140[°C]01052015[bar]0-40 100 200 300[psi]2001003000TableEPN16[°F]A0017265å 24 Fixed flange Table E, materials 1.0038 (S235JRG2), 1.0345 (P235GH), 1.0425/316L (P265GH), A105 and FE410 WB, DN 350 to 1200 (14 to 48"); fixed flange PN 16, materials 1.0044 (S275JR), 1.0425/316L (P265GH)and A105, DN 350 to 1200 (14 to 48")Pressure tightness Liner: hard rubber, polyurethaneNominal diameter Liner Limit values for absolute pressure in [mbar] ([psi]) for fluidtemperatures:[mm] [in] +25 °C (+77 °F) +50 °C (+122 °F) +80 °C (+176 °F)350…2400 14…90 Hard rubber 0 (0) 0 (0) 0 (0)50…1200 2…48 Polyurethane 0 (0) 0 (0) -Liner: PTFENominal diameter Limit values for absolute pressure in [mbar] ([psi]) for fluid temperatures:[mm] [in] +25 °C (+77 °F) +90 °C (+194 °F)50 2 0 (0) 0 (0)65 2 ½ 0 (0) 40 (0.58)80 3 0 (0) 40 (0.58)100 4 0 (0) 135 (2.0)125 5 135 (2.0) 240 (3.5)150 6 135 (2.0) 240 (3.5)200 8 200 (2.9) 290 (4.2)
  30. 30. Proline Promag L 40030 Endress+HauserNominal diameter Limit values for absolute pressure in [mbar] ([psi]) for fluid temperatures:[mm] [in] +25 °C (+77 °F) +90 °C (+194 °F)250 10 330 (4.8) 400 (5.8)300 12 400 (5.8) 500 (7.3)Flow limit The diameter of the pipe and the flow rate determine the nominal diameter of the sensor. The optimum velocityof flow is between 2 to 3 m/s (6.56 to 9.84 ft/s). Also match the velocity of flow (v) to the physical propertiesof the fluid:• v < 2 m/s (6.56 ft/s): for abrasive fluids (e.g. potters clay, lime milk, ore slurry)• v > 2 m/s (6.56 ft/s): for fluids producing buildup (e.g. wastewater sludges)For an overview of the measuring range full scale values, see the "Measuring range" section (® ä 6)Pressure loss • No pressure loss occurs if the sensor is installed in a pipe with the same nominal diameter.• Pressure losses for configurations incorporating adapters according to DIN EN 545 (® ä 23)System pressure • Never install the sensor on the pump suction side in order to avoid the risk of low pressure, and thus damageto the liner.• Furthermore, install pulse dampers if reciprocating, diaphragm or peristaltic pumps are used.• For information on the liners resistance to partial vacuum (® ä 29)• For information on the measuring systems resistance to vibration and shock (® ä 26), (® ä 26)A0015594Vibrations In the event of very strong vibrations, the pipe and sensor must be supported and fixed. It is also advisable tomount the sensor and transmitter separately.For information on the permitted resistance to vibration and shock (® ä 26), (® ä 26)LA0016266å 25 Measures to prevent vibration of the deviceL > 10 m (33 ft)
  31. 31. Proline Promag L 400Endress+Hauser 31Mechanical constructionDesign, dimensions Compact versionOrder code for "Housing", option M "Compact, polycarbonate" with DN 50 to 300 (2 to 12")ACBLHJFGEDKA0017150Dimensions in SI unitsDN L 1)A B C D E F G H J K[mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm]50 200 216 189 120 165 84 196 280 84 94 10965 200 216 189 180 165 84 221 330 109 94 10980 200 216 189 180 165 84 221 330 109 94 109100 250 216 189 180 165 84 221 330 109 94 109125 250 216 189 260 165 84 261 411 150 140 109150 300 216 189 260 165 84 261 411 150 140 109200 350 216 189 324 165 84 286 466 180 156 109250 450 216 189 400 165 84 311 516 205 166 109300 500 216 189 460 165 84 336 566 230 166 1091) The length is independent of the selected pressure rating. Length in accordance with DVGW/ISO.Dimensions in US unitsDN L 1)A B C D E F G H J K[in] [in] [in] [in] [in] [in] [in] [in] [in] [in] [in] [in]2 7.87 8.50 7.44 4.72 6.50 3.31 7.72 11.02 3.31 3.70 4.293 7.87 8.50 7.44 7.09 6.50 3.31 8.70 12.99 4.29 3.70 4.294 9.84 8.50 7.44 7.09 6.50 3.31 8.70 12.99 4.29 3.70 4.296 11.8 8.50 7.44 10.2 6.50 3.31 10.28 16.18 5.91 5.51 4.298 13.8 8.50 7.44 12.8 6.50 3.31 11.26 18.35 7.09 6.14 4.29
  32. 32. Proline Promag L 40032 Endress+HauserDN L 1)A B C D E F G H J K[in] [in] [in] [in] [in] [in] [in] [in] [in] [in] [in] [in]10 17.7 8.50 7.44 15.8 6.50 3.31 12.24 20.31 8.07 6.14 4.2912 19.7 8.50 7.44 18.1 6.50 3.31 13.23 22.28 9.06 6.54 4.291) The length is independent of the selected pressure rating. Length in accordance with DVGW/ISO.Order code for "Housing", option M "Compact, polycarbonate" with DN 350 to 2400 (14 to 90")CABLEFHGDKA0017153Dimensions in SI unitsDN L A B D E F H K[mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm]350 550 216 189 165 84 360 290 119375 600 216 189 165 84 386 290 119400 600 216 189 165 84 386 290 119450 600 216 189 165 84 414 290 119500 600 216 189 165 84 439 290 119600 600 216 189 165 84 480 290 119700 700 216 189 165 84 545 424 119750 750 216 189 165 84 582 454 119800 800 216 189 165 84 601 500 119900 900 216 189 165 84 651 580 1191000 1000 216 189 165 84 701 660 1191050 1050 216 189 165 84 737 755 1191200 1200 216 189 165 84 815 828 1191350 1350 216 189 165 84 927 1008 1191400 1400 216 189 165 84 927 1008 1191500 1500 216 189 165 84 1027 1147 1191600 1600 216 189 165 84 1027 1147 1191650 1650 216 189 165 84 1078 1284 119
  33. 33. Proline Promag L 400Endress+Hauser 33DN L A B D E F H K[mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm]1800 1800 216 189 165 84 1135 1379 1192000 2000 216 189 165 84 1246 1569 1192150 2150 216 189 165 84 1346 1711 1192200 2200 216 189 165 84 1346 1711 1192300 2300 216 189 165 84 1451 1859 1192400 2400 216 189 165 84 1451 1859 119Dimension C Dimension GEN (DIN) ASME AS EN (DIN) ASME ASDN PN 6 PN 10 AWWA PN 6 PN 10 AWWA[mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm]350 490 505 533 525 605 612 627 622375 – – – 550 – – – 661400 540 565 597 580 656 668 684 676450 595 615 635 640 711 721 731 734500 645 670 699 705 761 774 788 791600 755 780 813 825 857 870 886 892700 860 895 927 910 975 992 1008 1000750 – – 984 995 – – 1074 1080800 975 1015 1060 1060 1089 1109 1131 1131900 1075 1115 1168 1175 1189 1209 1235 12391000 1175 1230 1289 1255 1289 1316 1346 13291050 – – 1346 – – – 1410 –1200 1405 1455 1511 1490 1518 1543 1571 15601350 – – 1683 – – – 1769 –1400 1630 1675 – – 1742 1765 – –1500 – – 1854 – – – 1954 –1600 1830 1915 – – 1942 1985 – –1650 – – 2032 – – – 2094 –1800 2045 2115 2197 – 2157 2192 2233 –2000 2265 2325 2362 – 2378 2408 2427 –2150 – – 2534 – – – 2613 –2200 2475 2550 2705 – 2583 2621 – –2300 – – – – – – 2803 –2400 2685 2760 – – 2793 2831 – –Dimensions in US unitsDN L A B D E F H K[in] [in] [in] [in] [in] [in] [in] [in] [in]14 21.6 8.50 7.44 6.50 3.31 14.2 11.4 4.6915 23.6 8.50 7.44 6.50 3.31 15.2 11.4 4.69
  34. 34. Proline Promag L 40034 Endress+HauserDN L A B D E F H K[in] [in] [in] [in] [in] [in] [in] [in] [in]16 23.6 8.50 7.44 6.50 3.31 15.2 11.4 4.6918 23.6 8.50 7.44 6.50 3.31 16.3 11.4 4.6920 23.6 8.50 7.44 6.50 3.31 17.3 11.4 4.6924 23.6 8.50 7.44 6.50 3.31 18.9 11.4 4.6928 27.6 8.50 7.44 6.50 3.31 21.5 16.7 4.6930 29.5 8.50 7.44 6.50 3.31 22.9 17.9 4.6932 31.5 8.50 7.44 6.50 3.31 23.7 19.7 4.6936 35.4 8.50 7.44 6.50 3.31 25.6 22.8 4.6940 39.4 8.50 7.44 6.50 3.31 27.6 26.0 4.6942 41.3 8.50 7.44 6.50 3.31 29.0 29.7 4.6948 47.2 8.50 7.44 6.50 3.31 32.1 32.6 4.6954 53.1 8.50 7.44 6.50 3.31 36.5 39.7 4.6960 59.0 8.50 7.44 6.50 3.31 40.4 45.2 4.6966 64.9 8.50 7.44 6.50 3.31 42.4 50.6 4.6972 70.8 8.50 7.44 6.50 3.31 44.7 54.3 4.6978 78.7 8.50 7.44 6.50 3.31 49.1 61.8 4.6984 84.6 8.50 7.44 6.50 3.31 53.0 67.4 4.6990 90.5 8.50 7.44 6.50 3.31 57.1 73.2 4.69Dimension C Dimension GEN (DIN) ASME AS EN (DIN) ASME ASDN PN 6 PN 10 AWWA PN 6 PN 10 AWWA[in] [in] [in] [in] [in] [in] [in] [in] [in]14 19.3 19.9 21.0 20.7 23.8 24.1 24.7 24.515 – – – 21.7 – – – 2616 21.3 22.2 23.5 22.8 25.8 26.3 26.9 26.618 23.4 24.2 25.0 25.2 28 28.4 28.8 28.920 25.4 26.4 27.5 27.8 30 30.5 31 31.124 29.7 30.7 32.0 32.5 33.7 34.3 34.9 35.128 33.9 35.2 36.5 35.8 38.4 39.1 39.7 39.430 – – 38.7 39.2 – – 42.3 42.532 38.4 40.0 41.7 41.7 42.9 43.7 44.5 44.536 42.3 43.9 46.0 46.3 46.8 47.6 48.6 48.840 46.3 48.4 50.7 49.4 50.7 51.8 53 52.342 – – 53.0 – – – 55.5 –48 55.3 57.3 59.5 58.7 59.8 60.7 61.9 61.454 – – 66.3 – – – 69.6 –60 – – 73.0 – – – 76.9 –66 – – 80.0 – – – 82.4 –72 80.5 83.3 86.5 – 84.9 86.3 87.9 –78 89.2 91.5 93.0 – 93.6 94.8 95.6 –
  35. 35. Proline Promag L 400Endress+Hauser 35Dimension C Dimension GEN (DIN) ASME AS EN (DIN) ASME ASDN PN 6 PN 10 AWWA PN 6 PN 10 AWWA[in] [in] [in] [in] [in] [in] [in] [in] [in]84 – – 99.8 – – – 103 –90 – – – – – – 110 –Transmitter remote versionOrder code for "Housing", option N "Remote, polycarbonate"BAKGFHLJJCDEA0017148Dimensions in SI unitsA B C D E F[mm] [mm] [mm] [mm] [mm] [mm]165 185 15 25 225 80G H J K L[mm] [mm] [mm] [mm] [mm]50 53 56 88.5 53
  36. 36. Proline Promag L 40036 Endress+HauserDimensions in US unitsA B C D E F[in] [in] [in] [in] [in] [in]6.5 7.28 0.59 0.98 8.86 3.15G H J K L[in] [in] [in] [in] [in]1.97 2.09 2.2 3.48 2.09Sensor remote versionDN 50 to 300 (2 to 12")JLGHFBCDAEA0017282Dimensions in SI unitsDN L 1)A B C D E F G H J[mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm]50 200 129 120 163 143 102 202 286 84 9465 200 129 180 163 143 102 227 336 109 9480 200 129 180 163 143 102 227 336 109 94100 250 129 180 163 143 102 227 336 109 94125 250 129 260 163 143 102 267 417 150 140150 300 129 260 163 143 102 267 417 150 140200 350 129 324 163 143 102 292 472 180 156250 450 129 400 163 143 102 317 522 205 156300 500 129 460 163 143 102 342 572 230 1661) The length is independent of the selected pressure rating. Length in accordance with DVGW/ISO.
  37. 37. Proline Promag L 400Endress+Hauser 37Dimensions in US unitsDN L 1)A B C D E F G H J[in] [in] [in] [in] [in] [in] [in] [in] [in] [in] [in]2 7.87 5.08 4.72 6.42 5.63 4.02 7.95 11.3 3.32 3.703 7.87 5.08 7.10 6.42 5.63 4.02 8.94 13.2 4.30 3.704 9.84 5.08 7.10 6.42 5.63 4.02 8.94 13.2 4.30 3.706 11.8 5.08 10.2 6.42 5.63 4.02 10.5 16.4 5.91 5.518 13.8 5.08 12.8 6.42 5.63 4.02 11.5 18.6 7.10 6.1410 17.7 5.08 15.8 6.42 5.63 4.02 12.5 20.6 8.08 6.1412 19.7 5.08 18.1 6.42 5.63 4.02 13.5 22.5 9.06 6.541) The length is independent of the selected pressure rating. Length in accordance with DVGW/ISO.DN 350 to 2400 (14 to 90")LGFEBA CDA0017284Dimensions in SI unitsDN L A C D E G[mm] [mm] [mm] [mm] [mm] [mm] [mm]350 550 129 163 102 353 290375 600 129 163 102 379 290400 600 129 163 102 379 290450 600 129 163 102 407 290500 600 129 163 102 432 290600 600 129 163 102 473 290700 700 129 163 102 538 424750 750 129 163 102 575 454800 800 129 163 102 594 500900 900 129 163 102 644 5801000 1000 129 163 102 694 6601050 1050 129 163 102 730 7551200 1200 129 163 102 808 828
  38. 38. Proline Promag L 40038 Endress+HauserDN L A C D E G[mm] [mm] [mm] [mm] [mm] [mm] [mm]1350 1350 129 163 102 920 10081400 1400 129 163 102 920 10081500 1500 129 163 102 1020 11471600 1600 129 163 102 1020 11471650 1650 129 163 102 1071 12841800 1800 129 163 102 1128 13792000 2000 129 163 102 1239 15692150 2150 129 163 102 1339 17112200 2200 129 163 102 1339 17112300 2300 129 163 102 1444 18592400 2400 129 163 102 1444 1859Dimension B Dimension FEN (DIN) ASME AS EN (DIN) ASME ASDN PN 6 PN 10 AWWA PN 6 PN 10 AWWA[mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm]350 490 505 533 525 598 605 620 615375 – – – 550 – – – 654400 540 565 597 580 649 661 677 669450 595 615 635 640 704 714 724 727500 645 670 699 705 754 767 781 784600 755 780 813 825 850 863 879 885700 860 895 927 910 968 985 1001 993750 – – 984 995 – – 1067 1073800 975 1015 1060 1060 1082 1102 1124 1124900 1075 1115 1168 1175 1182 1202 1228 12321000 1175 1230 1289 1255 1282 1309 1339 13221050 – – 1346 – – – 1403 –1200 1405 1455 1511 1490 1511 1536 1564 15531350 – – 1683 – – – 1762 –1400 1630 1675 – – 1735 1758 – –1500 – – 1854 – – – 1947 –1600 1830 1915 – – 1935 1978 – –1650 – – 2032 – – – 2087 –1800 2045 2115 2197 – 2150 2185 2226 –2000 2265 2325 2362 – 2371 2401 2420 –2150 – – 2534 – – – 2606 –2200 2475 2550 – – 2576 2614 – –2300 – – 2705 – – – 2796 –2400 2685 2760 – – 2786 2824 – –
  39. 39. Proline Promag L 400Endress+Hauser 39Dimensions in US unitsDN L A C D E G[in] [in] [in] [in] [in] [in] [in]14 21.6 5.08 6.42 4.02 13.9 11.415 23.6 5.08 6.42 4.02 14.9 11.416 23.6 5.08 6.42 4.02 14.9 11.418 23.6 5.08 6.42 4.02 16.0 11.420 23.6 5.08 6.42 4.02 17.0 11.424 23.6 5.08 6.42 4.02 18.6 11.428 27.6 5.08 6.42 4.02 21.4 16.730 29.5 5.08 6.42 4.02 22.8 17.932 31.5 5.08 6.42 4.02 23.6 19.736 35.4 5.08 6.42 4.02 25.6 22.840 39.4 5.08 6.42 4.02 27.5 26.042 41.3 5.08 6.42 4.02 28.9 29.748 47.2 5.08 6.42 4.02 32.0 32.654 53.1 5.08 6.42 4.02 36.0 39.660 59.0 5.08 6.42 4.02 40.2 45.266 64.9 5.08 6.42 4.02 42.1 50.672 70.8 5.08 6.42 4.02 44.4 54.278 78.7 5.08 6.42 4.02 48.8 61.884 84.6 5.08 6.42 4.02 52.7 67.490 90.5 5.08 6.42 4.02 56.9 73.2Dimension B Dimension FEN (DIN) ASME AS EN (DIN) ASME ASDN PN 6 PN 10 AWWA PN 6 PN 10 AWWA[in] [in] [in] [in] [in] [in] [in] [in] [in]14 19.3 19.9 21.0 20.7 23.5 23.8 24.4 24.215 – – – 21.7 – – – 25.716 21.3 22.2 23.5 22.8 25.6 26.0 26.7 26.318 23.4 24.2 25.0 25.2 27.7 28.1 28.5 28.620 25.4 26.4 27.5 27.8 29.7 30.2 30.7 30.924 29.7 30.7 32.0 32.5 33.5 34.0 34.6 34.828 33.9 35.2 36.5 35.8 38.3 39.0 39.6 39.330 – – 38.7 39.2 – – 42.2 42.232 38.4 40.0 41.7 41.7 42.8 43.6 44.4 44.436 42.3 43.9 46.0 46.3 46.7 47.5 48.5 48.740 46.3 48.4 50.7 49.4 50.6 51.7 52.9 52.242 – – 53.0 – – – 55.4 –48 55.3 57.3 59.5 58.7 59.7 60.6 61.8 61.354 – – 66.3 – – – 69.4 –60 – – 73.0 – – – 76.7 –
  40. 40. Proline Promag L 40040 Endress+HauserDimension B Dimension FEN (DIN) ASME AS EN (DIN) ASME ASDN PN 6 PN 10 AWWA PN 6 PN 10 AWWA[in] [in] [in] [in] [in] [in] [in] [in] [in]66 – – 80.0 – – – 82.2 –72 80.5 83.3 86.5 – 84.7 86.05 87.7 –78 89.2 91.5 93.0 – 93.4 94.5 95.3 –84 – – 99.8 – – – 102.6 –90 – – 106.5 – – – 110.1 –AccessoriesGround disks for flange connectionsHHHDN £ 300 (12")mm (in)DN 300…600 (14…24")t = 2 (0.08)t = 2 (0.08) t = 2 (0.08)9 (0.35)ƳDN 700 (28")6.5 (0.26)ÆBÆAÆ DÆAÆBÆDÆ DÆAÆ6.4 (0.25)ÆA0015442Dimensions in SI and US unitsDN Pressure rating A B D H[mm] [in] [mm] [in] [mm] [in] [mm] [in] [mm] [in]50 2 1)52 2.05 101 3.98 115.5 4.55 108 4.2565 2 ½ 1)68 2.68 121 4.76 131.5 5.18 118 4.6580 3 1)80 3.15 131 5.16 154.5 6.08 135 5.31100 4 1)104 4.09 156 6.14 186.5 7.34 153 6.02125 5 1)130 5.12 187 7.36 206.5 8.13 160 6.30150 6 1)158 6.22 217 8.54 256 10.08 184 7.24200 8 1)206 8.11 267 10.51 288 11.34 205 8.07250 10 1)260 10.24 328 12.91 359 14.13 240 9.45300 12 1)312 12.28 375 14.76 413 16.26 273 10.75350 14 DIN, PN 6 343 13.50 433 16.54 479 18.86 365 14.37350 14 DIN, PN 10 343 13.50 420 17.05 479 18.86 365 14.37350 14 ASME, Class 150 343 13.50 420 17.05 479 18.86 365 14.37400 16 DIN, PN 6 393 15.47 470 18.50 542 21.34 395 15.55400 16 DIN, PN 10 393 15.47 480 18.90 542 21.34 395 15.55400 16 ASME, Class 150 393 15.47 480 18.90 542 21.34 395 15.55450 18 DIN, PN 6 439 17.28 525 20.67 583 22.95 417 16.42
  41. 41. Proline Promag L 400Endress+Hauser 41DN Pressure rating A B D H[mm] [in] [mm] [in] [mm] [in] [mm] [in] [mm] [in]450 18 DIN, PN 10 439 17.28 538 21.18 583 22.95 417 16.42450 18 ASME, Class 150 439 17.28 538 21.18 583 22.95 417 16.42500 20 DIN, PN 6 493 19.41 575 23.31 650 25.59 460 18.11500 20 DIN, PN 10 493 19.41 592 22.64 650 25.59 460 18.11500 20 ASME, Class 150 493 19.41 592 22.64 650 25.59 460 18.11600 24 DIN, PN 6 593 23.35 676 27.28 766 30.16 522 20.55600 24 DIN, PN 10 593 23.35 693 26.61 766 30.16 522 20.55600 24 ASME, Class 150 593 23.35 693 26.61 766 30.16 522 20.55700 28 DIN, PN 6 697 27.44 – – 786 30.94 460 18.11700 28 DIN, PN 10 693 27.28 – – 813 32.01 480 18.9700 28 AS, PN 16 687 27.05 – – 807 31.77 490 19.29700 28 AWWA, Class D 693 27.28 – – 832 32.76 494 19.45750 30 AWWA, Class D 743 29.25 – – 833 32.8 523 20.59800 32 DIN, PN 6 799 31.46 – – 893 35.16 520 20.47800 32 DIN, PN 10 795 31.30 – – 920 36.22 540 21.26800 32 AS, PN 16 789 31.06 – – 914 35.98 550 21.65800 32 AWWA, Class D 795 31.30 – – 940 37.01 561 22.09900 36 DIN, PN 6 897 35.31 – – 993 39.09 570 22.44900 36 DIN, PN 10 893 35.16 – – 1020 40.16 590 23.23900 36 AS, PN 16 886 34.88 – – 1014 39.92 595 23.43900 36 AWWA, Class D 893 35.16 – – 1048 41.26 615 24.211000 40 DIN, PN 6 999 39.33 – – 1093 43.03 620 24.411000 40 DIN, PN 10 995 39.17 – – 1127 44.37 650 25.591000 40 AS, PN 16 988 38.90 – – 1131 44.53 660 25.981000 40 AWWA, Class D 995 39.17 – – 1163 45.79 675 26.571050 42 AWWA, Class D 1044 41.10 – – 1220 48.03 704 27.721200 48 DIN, PN 6 1203 47.36 – – 1310 51.57 733 28.861) Ground disks can be used for all the flange standards/pressure ratings which can be supplied in the standard version.Weight Compact versionWeight data:• Including the transmitter (1.15 kg (2.54 lbs))• Excluding packaging materialWeight in SI unitsLap joint flange; fixed flange DN ³ 350 (14")NominaldiameterEN (DIN) ASME, AWWA AS[mm] [in] Pressurerating[kg] Pressurerating[kg] Pressurerating[kg] Pressurerating[kg] Pressurerating[kg]50 2 PN 16 10.6 PN 6 – Class 150 10.6 PN 16 – Table E –65 2 ½ PN 16 12 PN 6 – Class 150 – PN 16 – Table E –80 3 PN 16 14 PN 6 – Class 150 14 PN 16 – Table E –
  42. 42. Proline Promag L 40042 Endress+HauserNominaldiameterEN (DIN) ASME, AWWA AS[mm] [in] Pressurerating[kg] Pressurerating[kg] Pressurerating[kg] Pressurerating[kg] Pressurerating[kg]100 4 PN 16 16 PN 6 – Class 150 16 PN 16 – Table E –125 5 PN 16 21.5 PN 6 – Class 150 – PN 16 – Table E –150 6 PN 16 25.5 PN 6 – Class 150 25.5 PN 16 – Table E –200 8 PN 10 45 PN 6 – Class 150 45 PN 16 – Table E –250 10 PN 10 65 PN 6 – Class 150 65 PN 16 – Table E –300 12 PN 10 70 PN 6 – Class 150 70 PN 16 – Table E –350 14 PN 10 90 PN 6 79 Class 150 139 PN 16 101 Table E 101375 15 PN 10 – PN 6 – Class 150 – PN 16 107 Table E –400 16 PN 10 106 PN 6 91 Class 150 170 PN 16 122 Table E 122450 18 PN 10 114 PN 6 101 Class 150 193 PN 16 135 Table E 145500 20 PN 10 134 PN 6 116 Class 150 230 PN 16 184 Table E 184600 24 PN 10 164 PN 6 157 Class 150 304 PN 16 262 Table E 262700 28 PN 10 242 PN 6 192 Class D 268 PN 16 369 Table E 348750 30 PN 10 – PN 6 – Class D 320 PN 16 447 Table E 435800 32 PN 10 317 PN 6 242 Class D 385 PN 16 505 Table E 495900 36 PN 10 395 PN 6 310 Class D 472 PN 16 704 Table E 6921000 40 PN 10 470 PN 6 361 Class D 589 PN 16 761 Table E 7631050 42 PN 10 – PN 6 – Class D 672 PN 16 – Table E –1200 48 PN 10 719 PN 6 531 Class D 903 PN 16 1221 Table E 1239– 54 PN 10 – PN 6 – Class D 1275 PN 16 – Table E –1400 – PN 10 1116 PN 6 786 Class D – PN 16 – Table E –– 60 PN 10 – PN 6 – Class D 1596 PN 16 – Table E –1600 – PN 10 1626 PN 6 1060 Class D – PN 16 – Table E –1650 66 PN 10 – PN 6 – Class D 2133 PN 16 – Table E –1800 72 PN 10 2109 PN 6 1420 Class D 2570 PN 16 – Table E –2000 78 PN 10 2632 PN 6 1879 Class D 3115 PN 16 – Table E –– 84 PN 10 – PN 6 – Class D 3757 PN 16 – Table E –2200 – PN 10 3424 PN 6 2514 Class D – PN 16 – Table E –– 90 PN 10 – PN 6 – Class D 4799 PN 16 – Table E –2400 – PN 10 4096 PN 6 2998 Class D – PN 16 – Table E –Lap joint flange, stamped plateNominal diameter EN (DIN)[mm] [in] Pressure rating [kg]50 2 PN 10 7.265 2 ½ PN 10 880 3 PN 10 9100 4 PN 10 11.5125 5 PN 10 15150 6 PN 10 19
  43. 43. Proline Promag L 400Endress+Hauser 43Nominal diameter EN (DIN)[mm] [in] Pressure rating [kg]200 8 PN 10 37.5250 10 PN 10 56300 12 PN 10 57Weight in US unitsLap joint flange; fixed flange DN ³ 350 (14")Nominal diameter ASME, AWWA[mm] [in] Pressure rating [lbs]50 2 Class 150 2365 2 ½ Class 150 –80 3 Class 150 31100 4 Class 150 35125 5 Class 150 –150 6 Class 150 56200 8 Class 150 99250 10 Class 150 143300 12 Class 150 243350 14 Class 150 306375 15 Class 150 –400 16 Class 150 375450 18 Class 150 425500 20 Class 150 507600 24 Class 150 670700 28 Class D 591750 30 Class D 706800 32 Class D 849900 36 Class D 10411000 40 Class D 1299– 42 Class D 14821200 48 Class D 1991– 54 Class D 2811– 60 Class D 35191650 66 Class D 47031800 72 Class D 56672000 78 Class D 6869– 84 Class D 8284– 90 Class D 10582Transmitter remote version1.15 kg (2.54 lb)
  44. 44. Proline Promag L 40044 Endress+HauserSensor remote versionWeight data:• Including the transmitter housing• Excluding the connecting cable• Excluding packaging materialWeight in SI unitsLap joint flange; fixed flange DN ³ 350 (14")NominaldiameterEN (DIN) ASME, AWWA AS[mm] [in] Pressurerating[kg] Pressurerating[kg] Pressurerating[kg] Pressurerating[kg] Pressurerating[kg]50 2 PN 16 8.6 PN 6 – Class 150 8.6 PN 16 – Table E –65 2 ½ PN 16 10 PN 6 – Class 150 – PN 16 – Table E –80 3 PN 16 12 PN 6 – Class 150 12 PN 16 – Table E –100 4 PN 16 14 PN 6 – Class 150 14 PN 16 – Table E –125 5 PN 16 19.5 PN 6 – Class 150 – PN 16 – Table E –150 6 PN 16 23.5 PN 6 – Class 150 23.5 PN 16 – Table E –200 8 PN 10 43 PN 6 – Class 150 43 PN 16 – Table E –250 10 PN 10 63 PN 6 – Class 150 63 PN 16 – Table E –300 12 PN 10 68 PN 6 – Class 150 108 PN 16 – Table E –350 14 PN 10 87 PN 6 76 Class 150 136 PN 16 98 Table E 98375 15 PN 10 – PN 6 – Class 150 – PN 16 104 Table E –400 16 PN 10 103 PN 6 88 Class 150 167 PN 16 119 Table E 119450 18 PN 10 111 PN 6 98 Class 150 190 PN 16 132 Table E 142500 20 PN 10 131 PN 6 113 Class 150 227 PN 16 181 Table E 181600 24 PN 10 161 PN 6 154 Class 150 301 PN 16 259 Table E 259700 28 PN 10 240 PN 6 190 Class D 266 PN 16 367 Table E 346750 30 PN 10 – PN 6 – Class D 318 PN 16 445 Table E 433800 32 PN 10 315 PN 6 240 Class D 383 PN 16 503 Table E 493900 36 PN 10 393 PN 6 308 Class D 470 PN 16 702 Table E 6901000 40 PN 10 468 PN 6 359 Class D 587 PN 16 759 Table E 7611050 42 PN 10 – PN 6 – Class D 670 PN 16 – Table E –1200 48 PN 10 717 PN 6 529 Class D 901 PN 16 1219 Table E 1237– 54 PN 10 – PN 6 – Class D 1273 PN 16 – Table E –1400 – PN 10 1114 PN 6 784 Class D – PN 16 – Table E –– 60 PN 10 – PN 6 – Class D 1594 PN 16 – Table E –1600 – PN 10 1624 PN 6 1058 Class D – PN 16 – Table E –1650 66 PN 10 – PN 6 – Class D 2131 PN 16 – Table E –1800 72 PN 10 2107 PN 6 1418 Class D 2568 PN 16 – Table E –2000 78 PN 10 2630 PN 6 1877 Class D 3113 PN 16 – Table E –– 84 PN 10 – PN 6 – Class D 3755 PN 16 – Table E –2200 – PN 10 3422 PN 6 2512 Class D – PN 16 – Table E –– 90 PN 10 – PN 6 – Class D 4797 PN 16 – Table E –2400 – PN 10 4094 PN 6 2996 Class D – PN 16 – Table E –
  45. 45. Proline Promag L 400Endress+Hauser 45Lap joint flange, stamped plateNominal diameter EN (DIN)[mm] [in] Pressure rating [kg]50 2 PN 10 5.265 2 ½ PN 10 680 3 PN 10 7100 4 PN 10 9.5125 5 PN 10 13150 6 PN 10 17200 8 PN 10 35.5250 10 PN 10 54300 12 PN 10 55Weight in US unitsLap joint flange; fixed flange DN ³ 350 (14")Nominal diameter ASME, AWWA[mm] [in] Pressure rating [lbs]50 2 Class 150 1965 2 ½ Class 150 –80 3 Class 150 26100 4 Class 150 31125 5 Class 150 –150 6 Class 150 52200 8 Class 150 95250 10 Class 150 139300 12 Class 150 238350 14 Class 150 301375 15 Class 150 –400 16 Class 150 369450 18 Class 150 420500 20 Class 150 501600 24 Class 150 664700 28 Class D 587750 30 Class D 701800 32 Class D 845900 36 Class D 10361000 40 Class D 1294– 42 Class D 14771200 48 Class D 1987– 54 Class D 1273– 60 Class D 35151650 66 Class D 46991800 72 Class D 5662
  46. 46. Proline Promag L 40046 Endress+HauserNominal diameter ASME, AWWA[mm] [in] Pressure rating [lbs]2000 78 Class D 6864– 84 Class D 8280– 90 Class D 10577Measuring tube specification Nominal diameter Pressure rating Measuring tube internal diameterEN (DIN) ASME AS 2129 Hard rubber Polyurethane PTFEAWWA AS 4087[mm] [in] [mm] [in] [mm] [in] [mm] [in]50 2 PN 10/16 Class 150 – – – 50.3 2.0 51.7 2.065 1)2 ½ PN 10/16 Class 150 – – – 66.1 2.6 67.7 2.780 3 PN 10/16 Class 150 – – – 78.9 3.1 79.9 3.1100 4 PN 10/16 Class 150 – – – 104.3 4.1 103.8 4.1125 5 PN 10/16 Class 150 – – – 129.7 5.1 129.1 5.1150 6 PN 10/16 Class 150 – – – 158.3 6.2 156.3 6.2200 8 PN 10/16 Class 150 – – – 206.7 8.1 202.1 8.0250 10 PN 10/16 Class 150 – – – 260.6 10.3 256.2 10.1300 12 PN 10/16 – – – – 311.5 12.3 305.5 12.0300 12 – Class 150 – – – 309.9 12.2 303.9 12.0350 14 PN 6 – – 341 13.4 344 13.5 – –350 14 PN 10 – – 341 13.4 344 13.5 – –350 14 – – Table E, PN 16 339 13.3 342 13.4 – –350 14 – Class 150 – 339 13.3 342 13.4 – –375 15 PN 10 – – 391 15.4 – – – –375 15 – – PN 16 389 15.3 392 15.4 – –400 16 PN 6 – – 391 15.4 394 13.5 – –400 16 PN 10 – – 442 17.4 394 13.5 – –400 16 – – Table E, PN 16 389 15.3 392 13.4 – –400 16 – Class 150 – 389 15.3 392 13.4 – –450 18 PN 6 – – 442 17.4 445 17.5 – –450 18 PN 10 – – 493 19.4 445 17.5 – –450 18 – – Table E, PN 16 440 17.3 443 17.4 – –450 18 – Class 150 – 438 17.2 441 17.3 – –500 20 PN 6 – – 493 19.4 496 19.5 – –500 20 PN 10 – – 595 23.4 496 19.5 – –500 20 – – Table E, PN 16 489 19.2 492 19.3 – –500 20 – Class 150 – 489 19.2 492 19.3 – –600 24 PN 6 – – 595 23.4 598 23.5 – –600 24 PN 10 – – 590 23.2 598 23.5 – –600 24 – – Table E, PN 16 591 23.2 594 23.4 – –600 24 – Class 150 – 589 23.1 592 23.3 – –700 28 PN 6 – – 696 27.4 699 27.5 – –700 28 PN 10 – – 694 27.3 697 27.4 – –
  47. 47. Proline Promag L 400Endress+Hauser 47Nominal diameter Pressure rating Measuring tube internal diameterEN (DIN) ASME AS 2129 Hard rubber Polyurethane PTFEAWWA AS 4087[mm] [in] [mm] [in] [mm] [in] [mm] [in]700 28 – – Table E, PN 16 690 27.2 693 27.3 – –700 28 – Class D – 694 27.3 697 27.4 – –750 30 PN 6 – – – – 699 27.5 – –750 30 PN 10 – – – – 697 27.4 – –750 30 – – Table E, PN 16 741 29.2 744 29.3 – –750 30 – Class D – 743 29.3 746 29.4 – –800 32 PN 6 – – 798 31.4 801 31.5 – –800 32 PN 10 – – 796 31.3 799 31.5 – –800 32 – – Table E, PN 16 792 31.2 795 31.3 – –800 32 – Class D – 794 31.3 797 31.4 – –900 36 PN 6 – – 897 35.3 900 35.4 – –900 36 PN 10 – – 895 35.2 898 35.4 – –900 36 – – Table E, PN 16 889 35.0 892 35.1 – –900 36 – Class D – 895 35.2 898 35.4 – –1000 40 PN 6 – – 999 39.3 1002 39.4 – –1000 40 PN 10 – – 997 39.3 1000 39.4 – –1000 40 – – Table E, PN 16 991 39.0 994 39.1 – –1000 40 – Class D – 995 39.1 998 39.3 – –1050 42 PN 6 – – – – – – – –1050 42 PN 10 – – – – – – – –1050 42 – – Table E, PN 16 – – – – – –1050 42 – Class D – 1046 41.2 1049 41.3 – –1200 48 PN 6 – – 1203 47.4 1206 47.5 – –1200 48 PN 10 – – 1199 47.2 1202 47.3 – –1200 48 – – Table E, PN 16 1191 46.9 1194 47.0 – –1200 48 – Class D – 1195 47.0 1198 47.2 – –– 54 – Class D – 1345 53.8 – – – –1400 – PN 6 – – 1402 56.1 – – – –1400 – PN 10 – – 1394 5578 – – – –– 60 – Class D – 1498 59.9 – – – –1600 – PN 6 – – 1600 64.0 – – – –1600 – PN 10 – – 1590 63.6 – – – –– 66 – Class D – 1646 65.8 1198 47.2 – –1800 72 PN 6 – – 1800 72.0 1206 47.5 – –1800 72 PN 10 – – 1790 71.6 1202 47.3 – –1800 72 – Class D – 1790 71.6 1198 47.2 – –2000 78 PN 6 – – 1998 79.9 – – – –2000 78 PN 10 – – 1990 79.6 – – – –2000 78 – Class D – 1986 79.4 – – – –– 84 – Class D – 2099 84.0 – – – –
  48. 48. Proline Promag L 40048 Endress+HauserNominal diameter Pressure rating Measuring tube internal diameterEN (DIN) ASME AS 2129 Hard rubber Polyurethane PTFEAWWA AS 4087[mm] [in] [mm] [in] [mm] [in] [mm] [in]2200 – PN 6 – – 2194 87.8 – – – –2200 – PN 10 – – 2186 87.4 – – – –– 90 – Class D – 2246 89.8 – – – –2400 – PN 6 – – 2394 95.8 – – – –2400 – PN 10 – – 2386 95.4 – – – –1) Designed acc. to EN 1092-1 (not to DIN 2501)Materials Transmitter housing• Order code for "Housing", option M, N: polycarbonate plastic• Window material: polycarbonate plasticTransmitter cable entriesOrder code for "Housing", option M "Compact, polycarbonate"; option N "Remote, polycarbonate"The various cable entries are suitable for hazardous and non-hazardous areas.Electrical connection MaterialCable gland M20 × 1.5 PlasticThread G ½"via adapterNickel-plated brassThread NPT ½"via adapterNickel-plated brassConnecting cable for remote version• Electrode cable: PVC cable with copper shield• Coil current cable: PVC cable with copper shieldSensor housing• DN 50 to 300 (2 to 12"): aluminum coated AlSi10Mg• DN 350 to 2400 (14 to 90"): carbon steel with protective varnishSensor connection housingAluminum coating AlSi10MgSensor cable entriesOrder code for "Housing", option N "Remote, polycarbonate"The various cable entries are suitable for hazardous and non-hazardous areas.Electrical connection MaterialCable gland M20 × 1.5 Nickel-plated brassThread G ½"via adapterNickel-plated brassThread NPT ½"via adapterNickel-plated brass
  49. 49. Proline Promag L 400Endress+Hauser 49Measuring tubes• DN 50 to 300 (2 to 12"): stainless steel 1.4301/304, 1.4306/304L• DN 350 to 2400 (14 to 90"): stainless steel 202, 304Liner• DN 50 to 300 (2 to 12"): PTFE• DN 50 to 1200 (2 to 48"): polyurethane• DN 350 to 2400 (14 to 90"): hard rubberElectrodes1.4435/304L, Alloy C-22Process connectionsEN 1092-1 (DIN 2501)• DN £ 300 (12"): 1.0038 (S235JRG2), 1.4301/304, 1.4306/304L, 1.4307/304L• DN ³ 350 (14"): 1.0038 (S235JRG2), A105ASME B16.5• DN £ 300 (12"): A105, 316L• DN ³ 350 (14"): A105AWWA C207DN 1200 to 2200 (48 to 90"): 1.0044 (S275JR), 1.0425/316L (P265GH), A105, A181AS 2129DN 350 to 1200 (14 to 48"): 1.0038 (S235JRG2), 1.0345 (P235GH), 1.0425/316L (P265GH), A105, FE 410WBAS 4087DN 350 to 1200 (14 to 48"): 1.0044 (S275JR), 1.0425/316L (P265GH), A105SealsIn accordance with DIN EN 1514-1AccessoriesWeather protection coverStainless steel 1.4301Ground disks1.4435/316L, Alloy C-22Fitted electrodes Measurement, reference and empty pipe detection electrodes available as standard with:• 1.4435/304L• Alloy C-22
  50. 50. Proline Promag L 40050 Endress+HauserProcess connections Flange connections:• EN 1092-1 (DIN 2501)– DN £ 300 (12"), lap joint flange (PN 10/16); lap joint flange, stamped plate (PN 10) = form A– DN ³ 350 (14"), fixed flange (PN 6/10) = flat face• ASME B16.5– DN £ 300 (12"), lap joint flange (Class 150)– DN ³ 350 (14"), fixed flange (Class 150)• AWWA C207DN 1200 to 2200 (48 to 90"), fixed flange (Class D)• AS 2129DN 350 to 1200 (14 to 48"), fixed flange (Table E)• AS 4087DN 350 to 1200 (14 to 48"), fixed flange (PN 16)For information on the materials of the process connections (® ä 49)Surface roughness Electrodes with 1.4435/304L, Alloy C-22:£ 0.3 to 0.5 mm (11.8 to 19.7 min)(All data relate to parts in contact with fluid)OperabilityOperating concept Operator-oriented menu structure for user-specific tasks• Commissioning• Operation• Diagnostics• Expert levelRapid and safe commissioning• Guided menus ("Make-it-run" wizards) for applications• Menu guidance with brief explanations of the individual parameter functionsReliable operation• Operation in the following languages:– Via local display:English, German, French, Spanish, Italian, Dutch, Portuguese, Polish, Russian, Turkish, Chinese, Japanese,Bahasa (Indonesian), Vietnamese, Czech– Via "FieldCare" operating tool:English, German, French, Spanish, Italian, Dutch, Japanese– Via Web browser:English, German, French, Spanish, Italian, Dutch, Portuguese, Polish, Russian, Turkish, Chinese, Japanese,Bahasa (Indonesian), Vietnamese, Czech• Uniform operating philosophy applied to device, operating tools and Web browser• If replacing the electronic module, transfer the device configuration via the plug-in memory (HistoROM DAT)which contains the process and measuring device data and the event logbook. No need to reconfigure.Efficient diagnostics increase measurement availability• Troubleshooting measures can be called up via the device, operating tools and Web browser• Diverse simulation options and optional line recorder functions
  51. 51. Proline Promag L 400Endress+Hauser 51Local operationESC1A00171201 Operation with touch controlDisplay elements• 4-line display• White background lighting; switches to red in event of device errors• Format for displaying measured variables and status variables can be individually configured• Permitted ambient temperature for the display: –20 to +50 °C (–4 to +122 °F)The readability of the display may be impaired at temperatures outside the temperature range.Operating elementsExternal operation via touch control; 3 optical keys: , , ERemote operation Via HART protocol1 2 3 574 6A0017124å 26 Options for remote operation via HART protocol1 Control system (e.g. PLC)2 Field Communicator 4753 Computer with operating tool (e.g. FieldCare, AMS Device Manager, SIMATIC PDM)4 Commubox FXA195 (USB)5 Field Xpert SFX1006 VIATOR Bluetooth modem with connecting cable7 Transmitter
  52. 52. Proline Promag L 40052 Endress+HauserVia service interface (CDI-RJ45)213A00171191 Service interface (CDI -RJ45) of the measuring device with access to the integrated Web server2 Computer with Web browser (e.g. Internet Explorer) for accessing the integrated device Web server or with"FieldCare" operating tool with COM DTM "CDI Communication TCP/IP"3 Standard Ethernet connecting cable with RJ45 connectorCertificates and approvalsCE mark The measuring system is in conformity with the statutory requirements of the applicable EC Directives. Theseare listed in the corresponding EC Declaration of Conformity along with the standards applied.Endress+Hauser confirms successful testing of the device by affixing to it the CE mark.C-Tick symbol The measuring system meets the EMC requirements of the "Australian Communications and Media Authority(ACMA)".Ex approval cCSAusCurrently, the following versions for use in hazardous areas are available:NI (Ex nA, Ex nL)Class I Division 2 Groups ABCD; NIFW**= NIFW parameter as per Control DrawingsDrinking water approval • ACS• KTW/W270• NSF 61• WRAS BS 6920Other standards andguidelines• EN 60529Degrees of protection provided by enclosures (IP code)• EN 61010-1Protection Measures for Electrical Equipment for Measurement, Control, Regulation and LaboratoryProcedures.• IEC/EN 61326Emission in accordance with Class A requirements. Electromagnetic compatibility (EMC requirements)• ANSI/ISA-S82.01Safety Standard for Electrical and Electronic Test, Measuring, Controlling and related Equipment - GeneralRequirements. Pollution degree 2, Installation Category II• CAN/CSA-C22.2 No. 1010.1-92Safety requirements for Electrical Equipment for Measurement and Control and Laboratory Use. Pollutiondegree 2, Installation Category II
  53. 53. Proline Promag L 400Endress+Hauser 53• NAMUR NE 21Electromagnetic compatibility (EMC) of industrial process and laboratory control equipment.• NAMUR NE 32Data retention in the event of a power failure in field and control instruments with microprocessors• NAMUR NE 43Standardization of the signal level for the breakdown information of digital transmitters with analog outputsignal.• NAMUR NE 53Software of field devices and signal-processing devices with digital electronics• NAMUR NE 105Specifications for integrating fieldbus devices in engineering tools for field devices• NAMUR NE 107Status classification as per NE107Ordering informationDetailed ordering information is available from the following sources:• In the Product Configurator on the Endress+Hauser website: www.endress.com ® Select country ®Instruments ® Select device ® Product page function: Configure this product• From your Endress+Hauser Sales Center: www.endress.com/worldwideProduct Configurator - the tool for individual product configuration• Up-to-the-minute configuration data• Depending on the device: Direct input of measuring point-specific information such as measuring rangeor operating language• Automatic verification of exclusion criteria• Automatic creation of the order code and its breakdown in PDF or Excel output format• Ability to order directly in the Endress+Hauser Online ShopApplication packagesMany different application packages are available to enhance the functionality of the device. Such packagesmight be needed to address safety aspects or specific application requirements.The application packages can be ordered from Endress+Hauser either directly with the device or subsequently.Detailed information on the order code in question is available from your local Endress+Hauser sales center oron the product page of the Endress+Hauser website: www.endress.com.Cleaning Package DescriptionElectrode cleaning circuit(ECC)The electrode cleaning circuit (ECC) function has been developed to have a solution forapplications where magnetite (Fe3O4) deposits frequently occur (e.g. hot water). Sincemagnetite is highly conductive this build up leads to measuring errors and ultimately to theloss of signal. The option is designed to AVOID build up of highly conductive matter andthin layers (typically magnetite).Diagnostics functions Package DescriptionHistoROM extendedfunctionComprises extended functions concerning the event log and the activation of the measuredvalue memory (data logger).Event log:Memory volume is extended from 20 message entries (basic version) to up to 100 entries.Data logging (line recorder):• Memory capacity for up to 1000 measured values is activated.• 250 measured values can be output via each of the 4 memory channels. The recordinginterval can be defined and configured by the user.• Data logging is visualized via the local display or FieldCare.
  54. 54. Proline Promag L 40054 Endress+HauserAccessoriesVarious accessories, which can be ordered with the device or subsequently from Endress+Hauser, are availablefor the device. Detailed information on the order code in question is available from your local Endress+Hausersales center or on the product page of the Endress+Hauser website: www.endress.com.Device-specific accessories For the transmitterAccessories DescriptionPromag 400 transmitter Transmitter for replacement or storage. Use the order code to define the followingspecifications:• Approvals• Output / input• Display / operation• Housing• SoftwareWeather protection cover Is used to protect the measuring device from the effects of the weather: e.g. rainwater, excessheating from direct sunlight or extreme cold in winter.Connecting cable for remoteversionCoil current and electrode cables, various lengths, reinforced cables available on request.Ground cable Set, consisting of two ground cables for potential equalization.Pipe mounting set Pipe mounting set for transmitter.Compact ® remoteconversion kitFor converting a compact device version to a remote device version.Promag 50/53 ® Promag400 conversion kitFor converting a Promag with transmitter 50/53 to a Promag 400.For the sensorAccessories DescriptionGround disks for flangeconnectionsAre used to ground the fluid in lined measuring tubes to ensure proper measurement.For details, see Installation Instructions EA00070DCommunication-specificaccessoriesAccessories DescriptionCommubox FXA191 HART For intrinsically safe HART communication with FieldCare via the RS232C interface.For details, see "Technical Information" TI00237FCommubox FXA195 HART For intrinsically safe HART communication with FieldCare via the USB interface.For details, see "Technical Information" TI00404FCommubox FXA291 Connects Endress+Hauser field devices with a CDI interface (= Endress+Hauser CommonData Interface) and the USB port of a computer or laptop.For details, see "Technical Information" TI00405CHART Loop ConverterHMX50Is used to evaluate and convert dynamic HART process variables to analog current signalsor limit values.For details, see "Technical Information" TI00429F and Operating InstructionsBA00371F
  55. 55. Proline Promag L 400Endress+Hauser 55WirelessHART adapter Is used for the wireless connection of field devices.The WirelessHART adapter can be easily integrated into field devices and existinginfrastructures, offers data protection and transmission safety and can be operated in parallelwith other wireless networks with minimum cabling complexity.For details, see Operating Instructions BA00061SFieldgate FXA320 Gateway for the remote monitoring of connected 4-20 mA measuring devices via a Webbrowser.For details, see "Technical Information" TI00025S and Operating InstructionsBA00053SFieldgate FXA520 Gateway for the remote diagnostics and remote configuration of connected HART measuringdevices via a Web browser.For details, see "Technical Information" TI00025S and Operating InstructionsBA00051SField Xpert SFX100 Compact, flexible and robust industry handheld terminal for remote configuration and forobtaining measured values via the HART current output (4-20 mA).For details, see Operating Instructions BA00060SService-specific accessories Accessories DescriptionApplicator Software for selecting and sizing Endress+Hauser measuring devices:• Calculation of all the necessary data for identifying the optimum flowmeter: e.g. nominaldiameter, pressure loss, accuracy or process connections.• Graphic illustration of the calculation resultsAdministration, documentation and access to all project-related data and parameters overthe entire life cycle of a project.Applicator is available:• Via the Internet: https://wapps.endress.com/applicator• On CD-ROM for local PC installation.W@M Life cycle management for your plantW@M supports you with a wide range of software applications over the entire process: fromplanning and procurement, to the installation, commissioning and operation of themeasuring devices. All the relevant device information, such as the device status, spare partsand device-specific documentation, is available for every device over the entire life cycle.The application already contains the data of your Endress+Hauser device. Endress+Hauseralso takes care of maintaining and updating the data records.W@M is available:• Via the Internet: www.endress.com/lifecyclemanagement• On CD-ROM for local PC installation.FieldCare FDT-based plant asset management tool from Endress+Hauser.It can configure all smart field units in your system and helps you manage them. By usingthe status information, it is also a simple but effective way of checking their status andcondition.For details, see Operating Instructions BA00027S and BA00059SSystem components Accessories DescriptionMemograph M graphicdisplay recorderThe Memograph M graphic data manager provides information on all the relevant measuredvariables. Measured values are recorded correctly, limit values are monitored and measuringpoints analyzed. The data are stored in the 256 MB internal memory and also on a SD cardor USB stick.For details, see "Technical Information" TI00133R and Operating InstructionsBA00247R
  56. 56. DocumentationThe following document types are available:• On the CD-ROM supplied with the device• In the Download Area of the Endress+Hauser Internet site: www.endress.com ® DownloadStandard documentation Device type Communication Document type Documentation code5L4B**- - - - - Brief Operating Instructions KA01113DHART Operating Instructions BA01062DSupplementary device-dependent documentationDevice type Document type Approval Documentation code5L4B**- Installation Instructions Specified for eachindividual accessoryRegistered trademarksHARTÒRegistered trademark of the HART Communication Foundation, Austin, USAMicrosoftÒRegistered trademark of the Microsoft Corporation, Redmond, Washington, USAApplicatorÒ, FieldCareÒ, Field XpertTM, HistoROMÒRegistered or registration-pending trademarks of the Endress+Hauser GroupInstruments InternationalEndress+HauserInstruments International AGKaegenstrasse 24153 ReinachSwitzerlandTel.+41 61 715 81 00Fax+41 61 715 25 00www.endress.cominfo@ii.endress.comTI01045D/06/EN/01.1271183813EH-COSIMA ProMoDo

×