• Full Name Full Name Comment goes here.
    Are you sure you want to
    Your message goes here
  • Please let me download.. dausxray@yahoo.com
    Are you sure you want to
    Your message goes here
  • very well put together . please allow me to download this sb555@hotmail.com
    Are you sure you want to
    Your message goes here
  • great one.. please let me download this.. dushy.patel0@gmail.com
    Are you sure you want to
    Your message goes here
  • Please let me download too. patrickyip0 at gmail dot com Thanks
    Are you sure you want to
    Your message goes here
  • please share sir it is very helpful,
    'k_rampersad03@yahoo.com
    Are you sure you want to
    Your message goes here
No Downloads

Views

Total Views
35,637
On Slideshare
0
From Embeds
0
Number of Embeds
2

Actions

Shares
Downloads
0
Comments
15
Likes
39

Embeds 0

No embeds

Report content

Flagged as inappropriate Flag as inappropriate
Flag as inappropriate

Select your reason for flagging this presentation as inappropriate.

Cancel
    No notes for slide

Transcript

  • 1. Basic physics of ultrasound
    SPEAKER: Dr. JAYAPRABHA NARAYAN
    1st YEAR PG
    MODERATOR: Dr. ANAND S H
    ASSISTANT PROFESSOR
    DEPARTMENT OF RADIODIAGNOSIS
  • 2. The basis of all diagnostic ultrasound applications are the detection and display of acoustic energy reflected from within the body.
    In Medical Ultrasound, images representing human organs are formed by transmitting sound waves into the body and receiving back and processing the resultant echoes from the tissues.
    To accomplish this, medical ultrasound uses a process very similar to an ocean-going vessels “depth sounding” equipment or oceanic survey equipment. All of these systems make use of sound waves and their reflections.
    Sea
  • 3. BASIC ACOUSTICS
    Sound is the result of mechanical energy travelling through matter as a wave
    Sound waves are transmitted as a series of alternating pressure waves with high pressure and low pressure pulses.
    The high pressure areas (compression) are where the particles have been squeezed together; the low pressure areas (rarefaction) are where the particles have been spread apart.
    Sound waves cannot travel
    in vaccum
  • 4. Audible sound waves range in frequency from 20Hz to 20,000Hz.
    Ultrasound waves – frequency more than 20 KHz.
    For medical imaging – 2 to 20 MHz
    Sound waves – Transverse waves and longitudinal waves
    20 MHz
    Diagnostic
    Imaging
    1 MHz
    Ultrasonic
    20 KHz
    Audible sound
    20 Hz
    Infrasonic
    0 Hz
  • 5. Ultrasound waves – propagating through body tissues and fluids - longitudinal waves – producing compression and rarefaction of conducting medium.
    A zone of compression and an adjacent zone of rarefaction constitute one cycle of an ultrasound wave.
  • 6. Wavelength :The distance covered by one cycle is the wavelength of the ultrasound wave, λ
    The time for the ultrasound wave to pass a given point is Period, T
    Frequency: The number of cycles per unit time (cps, or just sec−1) introduced into the medium each second is referred to as the frequency of the wave, expressed in units of Hertz, Kilohertz or Megahertz, where 1Hz equals 1 cps.
    The maximum height of the wave cycle is the amplitude of the ultrasound wave.
  • 7. Propagation: in diagnostic ultrasound pressure waves are longitudinal waves and they propagate along the direction of particle movement.
    The product of the frequency (ν) and the wavelength (λ) is the propagationvelocity of the wave; that is, c = νλ.
    Propagation depends on the resistance of medium, resistance depends on density and its stiffness or elasticity of the medium .
  • 8. More stiff/less compressible is the molecules or less dense, more is the velocity in that medium.
    i.e. if density is more, velocity will be decresed and vice versa.
  • 9. On average the propagation velocity of sound in general body tissue is approximately 1540 m/sec
  • 10. Reflection
    Current diagnostic ultrasound scanners rely on the detection and display of reflected sound or echoes from interfaces between different tissues in the body.
    The fraction of the impinging energy reflected from an interface depends on the difference in acoustic impedance of the media on opposite sides of the interface.
    The acoustic impedance Z of a medium is the product of the density ρ of the medium and the velocity of ultrasound (c) in the medium:
    Z = ρc
    With a large impedance mismatch at an interface, much of the energy of an ultrasound wave is reflected, and only a small amount is transmitted across the interface.
  • 11. Incident wave
    Transmitted wave
    Medium 1
    Medium 2
    Reflected wave
  • 12. Specular Reflections, which occur at large change
    in impedance producing a large reflection, and also
    reducing the continuing wave amplitude.
    Medium Reflections, which occur with dense tissues
    such as muscle.
    Diffuse Reflections, which occur with soft tissues such
    as liver.
  • 13. Specular reflectors :
    Diaphragm
    Wall of urine filled urinary bladder
    Endometrial stripe
    Specular reflector Diffuse reflector
  • 14. Different types of reflections
    Scattering
  • 15. Distance measurement
  • 16. Refraction
    When a propagating ultrasound wave
    encounters a Specularinterface at an
    oblique angle, it is Refracted in the
    same way that light is refracted through a lens.
    The portion of the wave that is not reflected continues into the second medium, with a change in direction.
    It is dependent on the velocities of the
    two medium. If the velocities are equal,
    There would be no refraction occurred
    and the beam goes straight into the
    second medium.
    For the velocities of the different tissues in the human body
    are quite close, refraction's can be ignored
  • 17. Absorption
    Cause
    (a) Removal of energy from the ultrasound beam and
    (b) Eventual dissipation of this energy primarily as heat.
    Ultrasound is propagated by displacement of molecules of a medium into regions of compression and rarefaction
    Therefore, the energy of the ultrasound beam is gradually reduced as it passes through the medium.
  • 18. Attenuation
    • Loss of intensity of sound during propagation.
    • 19. Due to mainly reflection, refraction and absorptionAttenuation of ultrasound wave occurs when it is propagating through the medium.
    • 20. Loss of propagating energy will be in the form of heat absorbed by the tissue, approximately 1 dB/cm/MHz, or caused by wavefront dispersion or wave scattering
    .
  • 21. Scattering
    • Occurs when small particles absorb part of the ultrasound energy and re-radiate it in all directions as a spherical field. This means that the transducer can be positioned at any angle to the ultrasound beam and still receive echoes back. Scattering allows
    reflections from objects even
    smaller than the wavelength.
    Many biological interfaces
    have irregular surfaces,
    tending to give scatter-like
    reflection, which is quite
    Useful, as it will give at
    least some echoes even
    though the beam is not
    directly perpendicular
    to the reflecting interface.
    Spherical Scatter-wave
  • 22. Backscatter
    Backscatter or Rayleigh scattering occurs with structures smaller than the transmitted wavelength. Reflected energy is very low,
    but contributes to the texture of the image.
  • 23. INSTRUMENTATION
  • 24. BASIC COMPONENTS TO PERFORM KEY FUNCTIONS
    A transmitter/pulser – energize the transducer
    The Transducer
    A receiver
    Processor – to detect and amplify the backscattered energy and to manipulate the reflected signals for display
    A display – that presents ultrasound image or data
    A method to record or store data or images.
  • 25. Transducer
    A device – generates ultrasound waves for medical imaging.
    Two functions – Coverts electrical energy provided by transmitter into acoustic pulses directed into the patient.
    also serves as receiver of reflected echoes.
    Works on Piezoelectric Effect/piezoelectricity.
    it’s the unique property of certain natural or artificial crystals , Piezoelectric crystals – when subjected to electric voltage – crystal changes shape and vibrates with its resonant frequency(realignment and change in dimension) , produces sound beam which propagates into tissues( electrical energy converted to mechanical energy) and vice versa.
  • 26. Piezoelectric Crystals
    Some naturally piezoelectric occurring materials include Berlinite (structurally identical to quartz), cane sugar, quartz, Rochelle salt, topaz, tourmaline, and dry bone
    An example of man-made piezoelectric materials includes barium titanate and lead zirconate titanate (PZT).
    Most widely used
  • 27. TRANSDUCER DESIGN
    A crystal exhibits its greatest response at the resonance frequency
    The resonance frequency is determined by the thickness of the crystal and propagation velocity in its material
    Most efficient operation is achieved for a crystal with a thickness equal to half the wavelength of the desired ultrasound
    In most of the pulsed mode of operating ultrasound the output sound has frequency both above and below the resonance frequency
    The range of frequencies are called bandwidth. Generally shorter the pulse length of the transducer more is the bandwidth.
    Broad bandwidth helps to reduce speckle by frequency compounding
  • 28. TRANSDUCER DESIGN contd
    Spatial pulse length or Pulse length is the number of cycles per pulse
    In diagnostic ultrasound we need very small pulse length for better axial resolution.
    So suitable damping materials (backing) are kept behind the transducer material to shorten the pulse length and spacial pulse length by stopping reverberating pulses coming from the interface behind the transducer, thus gives improved axial resolution.
    also reduces ultrasound amplitude and decreases efficiency and sensitivity of the system.
    Damping materials used – mixture of metal powder(tungsten or aluminum) and a plastic or epoxy
  • 29. A 1/4λ (quarter-wavelength) thickness matching layer or coating in front of the transducer (generally made up of aluminum powder and epoxy resin) maximizes energy transfer from the transducer to the patient( i.e. increases sound transmission)
    The filler material enables the transducer assembly to be flat.
    The elements with its associated case, damping and matching material is called Transducer Assembly Or Probe
  • 30. Transducer schematic diagram
  • 31. PULSE – ECHO PRINCIPLE
    When the crystal is electrically pulsed, it changes shape and vibrates produces sound beam that propagates through tissues and reflected echoes reach the transducer it vibrates and produces electric voltage comparable to returning echo.
  • 32. The ultrasound pulses produced by the transducer results in a series of wave fronts that form a three dimensional beam of ultrasound.
    the features of beam are influenced by constructive and destructive interferences of the pressure waves.
  • 33. ULTRASOUND BEAMS
    Ultrasound from a point source creates spherical wave fronts. Ultrasound from a two-dimensional extended source creates planar wave fronts.
    These sources can be considered to be a collection of point sources, each radiating spherical wave fronts (termed wavelets) into the medium.
    In regions where compression zones for one wavelet intersect those of another, a condition of constructive interference is established & the wavelets reinforce each other. Reverse that is cancellation occurs in destructive interference
  • 34. Interference of the waves
  • 35. In this figure, the reinforcement and cancellation of individual wavelets are most noticeable in the region near the source of ultrasound.
    They are progressively less dramatic with increasing distance from the ultrasound source
    The near field where pressure amplitude change is maximum that zone is called “Fresnel zone”
    Beyond the Fresnel zone, some of the energy escapes along the periphery of the beam to produce a gradual divergence of the ultrasound beam – called Fraunhofer ( or far) zone
  • 36.
  • 37. For medical applications of ultrasound, beams with little lateral dispersion of energy (i.e., long Fresnel zones) are preferred.
    Rules for Transducer Design
    For a given transducer diameter,
    the near-field length increases with increasing frequency;
    beam divergence in the far field decreases with increasing frequency
    For a given transducer frequency
    the near-field length increases with increasing transducer diameter;
    beam divergence in the far field decreases with increasing transducer diameter.
  • 38. Focused Transducers
    A focused ultrasound transducer produces a beam that is narrower at some distance from the transducer face than its dimension at the face of the transducer.
    In the region where the beam narrows (termed the focal zone of the transducer), the ultrasound intensity may be heightened by 100 times or more compared with the intensity outside of the focal zone.
    Because of this increased intensity, a much larger signal will be induced in a transducer from a reflector positioned in the focal zone.
    The distance between the location for maximum echo in the focal zone and the element responsible for focusing the ultrasound beam is termed the focal length of the transducer
    An ultrasound beam also may be focused with mirrors and refracting lenses or phasing the linear array electronically
  • 39. Focused Transducers
  • 40. Receiver
    Returning echoes strike the transducer face (which is on quiescent mode electronically) and they produce minute voltages in the piezoelectric crystals.
    The receiver detects those signals and amplifies those
    They also provide compensatory amplification of the weaker signals those are coming from the deeper tissues.
    This is calledtime gain compensation
  • 41. Time gain compensation
  • 42. Time gain compensation
    We can also suppress increased amplitudes of the near field for better comparable image
  • 43. Receiver
    Another function of the receiver is compression of the wide range of amplitudes returning to the transducer into a range that can be displayed to the user. The range is called dynamic rangeof the transducer.
    Compression and remapping of the data is required to adapt the dynamic range of backscattered signal intensity to the dynamic range of display
  • 44. Receiver – dynamic range
    The widest dynamic range (60 dB) shown permits the best differentiation of subtle differences in echo intensity and preferred in most imaging application
    The narrower ranges increase conspicuity of larger echo differences.
  • 45. ARRAYS(TRANSDUCER TYPES)
  • 46. Linear Array
    The linear array scanners produce sound waves parallel
    to each other and produces a rectangular image.
    The width of the image and number of scan lines are the same at all tissue levels.
    This has the advantage of good near field resolution.
    Often used with high frequencies i.E. 7mhz.
    Can be used for viewing surface texture of the liver.
    There disadvantage is artifacts when applied to a
    curved part of the body creating air gaps between
    skin and transducer.
  • 47. Sector/Phased array
     ( freq 1 to 3MHz in adult, in pediatric sector probes upto 8 MHz) Produces a fan like image that is narrow near the transducer and increase in width with deeper penetration.
    It is useful when scanning between the ribs as it fits in the intercostals space.
    The disadvantage is poor near field resolution
  • 48. Curved Array
    Often with frequencies of 2 - 5 MHz (to allow for a range of patients from obese to slender).
    It is a compromise of the Linear and Sector scanners. The density of the scan lines decreases with increasing distance from the transducer.
    Can be difficult to use in curved regions of the body eg. the spleen behind the left costal margin.
    For abdominal ultrasound curved type scanners are used as the best compromise of two other standard type probes the linear and the sector scanner. 
  • 49. Annular Array
    A series of concentric elements nested within one another in a circular piece of piezoelectric crystal to produce an annular array
    Use of multiple concentric elements enables precise focussing.
  • 50. Linear & Phased array
    Linear array-individual elements or groups are fired in sequence resulting parallel US beams
    Phasic array: produce a sector field of view by firing multiple transducer elements in precise sequence to generate interference of acoustic wave fronts
  • 51.
  • 52. Transducer Selection
    For general purpose – convex with 3.5 MHz
    For obstetric purpose – convex or linear with 3.5 MHz
    For superficial structures – linear with 5MHz
    For pediatric or in thin people – 5MHz
  • 53. COUPLING AGENTS
    Commonly known as “GEL”
    Fluid medium needed to provide a link between the transducer and the patient
    Composition :
    Carbomer – 10.0 gm
    Propylene glycol – 75.0gm(72.4ml)
    Trolamine – 12.5gm(11.2ml)
    EDTA – 0.25gm
    Distilled water – upto 500gm or 500ml
  • 54. IMAGE DISPLAY
  • 55. A-Mode (amplitude mode)
    A mode display consists of a horizontal baseline. This baseline represents time and or distance with upward (vertical) deflections (spikes depicting the acoustic interfaces)
    In the A-mode presentation of ultrasound images, echoes returning from the body are displayed as signals on an oscilloscope.
  • 56. A-Mode (amplitude mode)
    A-mode reveals the location of echo-producing structures only in the direction of the ultrasound beam.
    A-mode displays are not found on most imaging systems used today.
    The concept of A-mode is however, useful in explaining how pixels are obtained from scan lines in B-mode imaging.
  • 57. B-mode (Brightness Mode)
    Real-time gray scale B-Mode display
    A two-dimensional display of ultrasound.
    To generate a two dimensional image multiple ultrasound pulses are sent down as a series of successive scan lines, building a 2D representation of echoes arising from the area being scanned.
    When an ultrasound image js displayed on a black background, signals of greatest intensity appear as white, absence of signal is shown as black(anechoic) and signals of intermediate intensity appear as shades of grey.
  • 58.
  • 59. B-mode, or brightness mode, ultrasound display.
  • 60. B-mode images may be displayed as either “static” or “real-time” images.
    In static imaging the image is compiled as the sound beam is scanned across the patient, and the image presents a “snapshot” averaged over the time required to sweep the sound beam.
  • 61. B mode real time imaging
    In real-time imaging, the image is also built up as the sound beam scans across the patient, but the scanning is performed automatically and quickly, and one image follows another in quick succession.( as many as 30-60 complete images per second)
    Real-time B-mode images are useful in the display of moving structures such as heart valves.
  • 62. M-mode Ultrasound (motion mode)
    Is the motion mode displaying moving structures along a single line in the ultrasound beam
    A single beam in an ultrasound scan can be used to produce an M-mode picture where movement of a structure such as a heart valve can be depicted in a wavelike manner.
    Because of its high sampling frequency (up to 1000 pulses per second) This is useful in assessing rates and motion and is still used extensively in cardiac and fetal cardiac imaging
  • 63. IMAGE QUALITY
    Spacial Resolution
    Axial : higher frequency – low wavelength  higher axial resolution
    Lateral resolution: excessive beam width and thickness limit the ability to delineate small features such as tiny cystic changes in atherosclerotic plaque
    Lateral resolution is controlled by focusing the beam by electronic phasing
    Elevation or Azimuthal resolution is determined by construction of the transducer and can not be controlled by the user
  • 64. IMAGE ARTIFACTS
  • 65. Mirror Image Artifact
    • This is where a strong reflector at an angle to the probe causes structures that lie in front and to the side of it to appear as if they lie behind it, just as something viewed through a mirror appears to lie behind it.• e.g. diaphragm
  • 66. Reverberation artifact
    When the ultrasound signal reflects repeatedly between highly reflective interfaces that are usually near the transducer
    Reduced by changing the scan angle and placement of the transducer
  • 67. Comet tail artifacts
    • This is the same process as reverberation, but occurs within a very small structure, with smooth highly reflective borders, e.g. metal fragment.
    • Tiny bright reverberations are seen deep to the structure slowly diminishing is size as if it had a tail
  • 68. Refraction Artifact
    This can lead to subtle miss placement of structures and some degradation of image quality when the angle of incidence is particularly acute.
    Rectus abdominis acting as refracting lens
  • 69. Range artifact
  • 70. Side Lobe Artifacts
    The probe cannot produce a pulse that travels purely in one direction.
    Pulses also travel off at specific angles.
    Side lobes may interact with strong reflectors that lie outside the scan plane.
    These side lobes are relatively weak and so normally do little to degrade the image.
    Can be reduced by repositioning of the transducer ,or its focal zone or using a different transducer.
    GB
  • 71. Acoustic shadowing
    Marked reduction in the intensity of ultrasound deep to a strong reflector or attenuator.
    ( other cause of loss of image information are improper system gain and TGC setting, poor scanning angles and poor resolution)
  • 72. Acoustic enhancement
    Due to high penetration of sound through a structure and then being reflected from the tissue behind it.
  • 73. Multipath artifact
    Echoes reflected (e.g. from diaphragm and wall of ovarian cyst) may create complex echo path that delay return of the reflected signal thus showing false deeper/abnormal position of the tissues
    Ovarian cyst --- false mural nodule
  • 74. Speckle Artifact
    Usual grainy appearance of the USG image
    Due to random interference of the scatterer .
    Irregular interference pattern. This is generated by more scatterers somewhat randomly distributed. The speckle pattern is thus random too. 
    Interference pattern. Here is simulated two wave sources or scatterers at the far field (white points).
  • 75. Speckle Artifact
  • 76. Spacial imaging modes
    Harmonic imaging: to improve spatial resolution and reduction of side lobes
    Spatial compounding: to reduce speckle using multiple scan angle
    Spatial compounding
  • 77. “HARDWORK , SINCERITY AND PERSEVERENCE
    PAVES THE WAY TO SUCCESS”
    THANK YOU