Your SlideShare is downloading. ×
Picarro Combustion Module Cavity Ring Down Spectroscopy Stable Isotope Analysis
Upcoming SlideShare
Loading in...5
×

Thanks for flagging this SlideShare!

Oops! An error has occurred.

×

Saving this for later?

Get the SlideShare app to save on your phone or tablet. Read anywhere, anytime - even offline.

Text the download link to your phone

Standard text messaging rates apply

Picarro Combustion Module Cavity Ring Down Spectroscopy Stable Isotope Analysis

902
views

Published on

Picarro's Combustion Module - Cavity Ring-Down Spectroscopy (CRDS) platform can run automated stable isotope sampling for Carbon 13 for as long as five hours without interruption. This is dramatic …

Picarro's Combustion Module - Cavity Ring-Down Spectroscopy (CRDS) platform can run automated stable isotope sampling for Carbon 13 for as long as five hours without interruption. This is dramatic improvement in cost and ease of use of traditional IRMS. Read about the technology at www.picarro.com

Published in: Technology, Business

0 Comments
0 Likes
Statistics
Notes
  • Be the first to comment

  • Be the first to like this

No Downloads
Views
Total Views
902
On Slideshare
0
From Embeds
0
Number of Embeds
0
Actions
Shares
0
Downloads
11
Comments
0
Likes
0
Embeds 0
No embeds

Report content
Flagged as inappropriate Flag as inappropriate
Flag as inappropriate

Select your reason for flagging this presentation as inappropriate.

Cancel
No notes for slide

Transcript

  • 1. Combustion Module-CRDS A Unique Solution for Bulk 13 C Analysis
  • 2. What if you could run bulk 13 C analysis…
    • With high-precision
    • With unprecedented ease-of-use
    • With a compact system
    • For far cheaper than IRMS ?
  • 3. From Food Apps to Packaging Material
    • Food adulteration
    • CID in Plant breeding
    • Soils / Sediments
    • Carbonates
    • Green Packaging QC
  • 4. Picarro CM-CRDS Enables…. High-precision δ 13 C analysis of solids & liquids Combustion Module (CM) Liaison ™ Interface CRDS
  • 5. Liaison Universal Interface 1. Admits gas sample from front-end 2. Collects current processed sample 3. Delivers previous sample to CRDS 4. Purges gas bag to collect next sample In Parallel: Fully Automated
  • 6. Excellent precision
    • 5 hr analysis of 29 solid samples of acetanilide std:
    • => Excellent precision: SD < 0.1 ‰
    0 1 2 3 4 5 Hrs
  • 7. Honey Samples Analysis Sample Delta ( ‰ ) S.D. ( ‰ ) Std Certified value USGS 40 First -26.370 0.090 -26.389 ‰ Honey A -22.687 0.165 Honey B -22.324 0.120 Honey C -25.690 0.184 Honey D -25.555 0.047 Honey E -27.670 0.082 USGS 40 Mid -26.398 0.054 -26.389 ‰ Honey F -24.946 0.089 Honey G -24.245 0.032 Honey H -24.121 0.065 Honey I -24.128 0.046 Honey J -24.217 0.107 USGS 40 Last -26.381 0.083 -26.389 ‰
  • 8. Honey Analysis Compared with IRMS Sample Delta ( ‰ ) S.D. ( ‰ ) IRMS ( ‰ ) Difference ( ‰ ) Honey A -22.22 0.165 -22.2 -0.02 Honey B -21.85 0.120 -22.0 0.15 Honey C -25.22 0.184 -25.6 0.38 Honey D -25.08 0.047 -25.1 0.02 Honey E -27.20 0.082 -26.8 -0.40 Honey F -24.47 0.089 -24.4 -0.07 Honey G -23.77 0.032 -23.9 0.13 Honey H -23.65 0.065 -23.9 0.15 Honey I -23.66 0.046 -23.8 0.14 Honey J -23.74 0.107 -24.2 0.46
  • 9. Food Fraud Detection
    • Honey adulteration with corn syrup: 5%, 10% & 15% w/w
    Exceptional correlation of calculated & measured δ 13 C values Δδ values are < 0.1 ‰ , even at the 5% adulteration level 100% HFCS 100% Honey