  • 1. Developing Key Optometric Skills Workshops ManualDeveloping Key Optometric Skills
  • 2. Commissioned for for NHS Education for Scotland by: Replay Learning Townhead Street Thornhill DG3 5NW 01848 332020 info@replaylearning.com www.replaylearning.com Page 3
  • 3. ContentsForeword ......................................................................................................................................... PAGE 7Timetables ...................................................................................................................................... PAGE 9Abstracts of Workshops ............................................................................................... PAGE 11Chapter 1: Indirect Ophthalmoscopy .................................................................. PAGE 13Chapter 2: Contact Applanation Tonometry .............................................. PAGE 33Chapter 3: Gonioscopy and Anterior Angle Assessment ........... PAGE 49Chapter 4: Slit Lamp Biomicroscopy .................................................................... PAGE 69Chapter 5: Visual Field Analysis ................................................................................... PAGE 89Useful References ................................................................................................................. PAGE 104Notes ................................................................................................................................................... PAGE 105 Page 5
  • 4. Page 6 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 5. Foreword“Following the successful establishment of the new GOS EyeExam Contract in 2006, the Scottish Government’s HealthDirectorate has recognised the increasing role Optometryis playing in primary health care by supporting the ongoingeducation and training for optometrists with a £1m annual fundto be administered through NHS Education Scotland (NES).The initial priority is to encourage every optometrist to becomeskilled in the use of Volk lenses, advanced slit lamp use, contacttonometry and anterior angle assessment - including gonioscopy.These techniques should no longer be considered “advancedinvestigative techniques” but part of the skills which the publicshould expect of all optometrists. In order to give as manypractitioners as possible the chance to improve and extendtheir practical skills, NES has commissioned a series of Scotland-wide practical workshops covering these areas. The workshopswill give practitioners the opportunity to practise these skillsin small groups under the supervision of expert tutors whohave experience of using the techniques in a day-to-day clinicalsetting.”Regards.Donald CameronProgramme Director (Optometry)NHS Education for Scotland Page 7
  • 6. Timetable09.00 to 09.30 Registration & Coffee09.30 to 10.00 Introduction10.00 to 11.15 Workshop 1 Indirect Ophthalmoscopy11.15 to 12.30 Workshop 2 Contact Applanation Tonometry12.30 to 13.30 Lunch13.30 to 14.45 Workshop 3 Gonioscopy and Anterior Angle Assessment14.45 to 16.00 Workshop 4 Slit Lamp Biomicroscopy16.00 to 16.30 Tea16.30 to 17.30 Tutorial Visual Field Assessment17.30 to 17.45 Close Page 9
  • 7. Tutorial Visual Fields Seminar Lecture: 1 hour Target Audience: Optometrists, Dispensing Opticians Abstract In small groups, delegates will be asked to interpret a series of visual field plots from the most common visual field analysers. The field plots are presented as case histories covering the most important areas of Visual Field Assessment. Each case will be discussed with the whole group in an interactive manner.Page 10 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 8. Workshop AbstractsIndirect OphthalmoscopyWorkshop: 1 hourCET points: 2Target Audience: Optometrists, Dispensing OpticiansA hands-on workshop using condensing lenses and a slit lamp and/or a headband binocular indirect ophthalmoscope to view the fundusof dilated and non dilated eyes of patients with pathology. Delegateswill be guided in the examination of the central and peripheral fundusby an experienced optometrist. Different powers of lenses will beavailable to give delegates an appreciation of the differences in the viewobtained. Using video capture, the relative advantages of each lens willbe reinforced.Contact Applanation TonometryWorkshop: 1 hourCET points: 2Target Audience: OptometristsA hands-on workshop where delegates will measure IOP on each otherusing both a Goldmann and Perkins tonometer under the supervisionof an experienced optometrist. The advantages and disadvantages ofboth techniques will be discussed as will the use of various topicalanaesthetics and fluorescein. Page 11
  • 9. Workshop Abstracts Gonioscopy and Anterior Angle Assessment Workshop: 1 hour CET points: 2 Target Audience: Optometrists A hands-on workshop where delegates will use a gonioscope to view the anterior angle of prepared subjects. Delegates will also have an opportunity to assess the same angles using Van Herrick’s or Smith’s method. Video capture will be used to demonstrate interpretation of results and the results of the different techniques will be compared. Slit Lamp Biomicroscopy Workshop: 1 hour CET points: 2 Target Audience: Optometrists,Dispensing Opticians A hands-on workshop where delegates will work in small groups to examine patients with pathology. The lecture will focus on slit lamp techniques such as angle assessment, anterior chamber assessment and anterior vitreous assessment. Other slit lamp techniques will be performed according to the requirements of the group.Page 12 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 10. Chapter 1 Indirect OphthalmoscopyA thorough examination of the ocular fundus is mandatory in any eyeexamination. The routine employed by the practitioner will largelydepend on the specific history and symptoms a patient presents with.Factors such as age, systemic medication and race may also play animportant role and may warrant adaptation to the practitioner’s normalroutine.To dilate or not?There can be no argument that fundus examination is greatly facilitatedby dilating the pupil. However, the decision of whether or not todilate a patient (or indeed every patient) must be based on reasonedclinical decision-making, particularly as an adequate fundus examinationcan be performed with most instruments without pupil dilation. Theadvantages of an enhanced fundus view must generally outweigh anynegative effects of pupillary dilation.It is still a concern of some practitioners that pupil dilation may induceangle-closure and lead to an attack of glaucoma. Although the rootsof this statement are solid, the incidence, which is often quoted aroundless than 1:100,000, is very small. Indeed, there are experiencedpractitioners who readily dilate patients, who have never witnessedsuch an event in their lengthy career. It is important however, notto be frugal with dilation and it is essential that all the risk factors areeliminated before attempting to dilate an individual.Normal history and symptoms may pinpoint individuals who might havea predisposition to angle closure. Slit lamp examination may also be Page 13
  • 11. useful in identifying patients with iris/angle anomalies who may also be at risk. An assessment of the anterior chamber depth can be obtained using a Van Herrick technique. However, it is important to bear in mind that a patient with a shallow anterior chamber depth by this technique is not necessarily at risk of closure. The cautious practitioner may also want to determine intraocular pressure (IOP) pre-dilation and repeat this measurement approximately 30minutes after maximum dilation has been achieved. Any significant rise in IOP (>5mmHg) is indicative of angle-closure and would warrant immediate referral to a casualty department. In addition to those with shallow-angles or risk factors that may precipitate angle closure, it is accepted that there are a number of patients for who dilation is contraindicated. These are: • Patients with history of penetrating injury • Patients with iris-clip intra ocular lenses Driving following pupil dilation is generally contraindicated, however, for patients who need to drive home following an eye examination a supplementary appointment can be made for fundus examination, on a day that they do not have to drive or when they can be accompanied. An exception to this last point is when the patient presents with symptoms indicative of a serious retinal pathology, such as a detachment. Here the patient must be dilated immediately as a thorough examination is required. There is some debate as to which pharmacological agent should be used to dilate a pupil. The choice of drug may vary with each patient and the nature of the examination to be performed. For example if fundus examination and a cycloplegic refraction in a young patient are required there is very little point in using a mild miotic drug. Any drop used should be fast acting, able to be effective for the full duration required, minimally toxic, have low risk of any side effects (local or systemic) and have a quick recovery time. Most authorities advise using either of the following combinations: • One drop of 2.5% phenylephrine (stimulating the dilatator pupillae of the iris) together with one drop of 0.5% tropicamide (restricts the action of the sphincter pupillae). The synergisticPage 14 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 12. action of these 2 drugs produces maximal pupil dilation and improves patient comfort as it restricts pupil constriction during examination. • Two drops of 0.5% tropicamide instilled a minute apart. The choices above are not prescriptive and the practitioner may modify their choice based on factors such as age of patient, race or patients general health. It is important to bear in mind that dilation can be facilitated up by instilling one drop of topical anaesthetic prior to instillation of the dilating drop(s), as this increase the permeability across the cornea.Instillation of dropsA number of simple precautions are essential when instilling drops.Most practitioners use single dose units, Minims, for ophthalmic drugs,but where multidose drops are used it is essential to assess the patencyof the bottle and not use any where there appears to be any damage.In all cases, when instilling drops; • Check the date of expiry as old drugs are less effective and may even cause ocular surface reaction. • Check the drug and dose is the one wanted. • Check that the patient has had no previous adverse response to a drop –though unlikely, this may be useful in cases where there has been IOP changes in the past. • Wash hands. • Getting the patient to look upward, the lower lid may be pulled down gently and a single drop instilled in the lower fornix. This avoids any risk of corneal contact were the patient to flinch suddenly or flick their eyes up in a Bell’s movement. • Never let the delivery unit touch the patient. • The same unit is usable for both patient eyes. • Issue a leaflet summarising the exact nature of the drop action and any symptoms to report, to whom, in the unlikely event of an adverse response. Examples of these are available from the College of Optometrists website. Page 15
  • 13. The theory behind indirect ophthalmoscopy Direct ophthalmoscopy, rather unsurprisingly, allows direct viewing of the patient fundus. This is achieved by the focusing of light from a lamp via an array of condensing lenses onto a mirror which reflects light to a focus on the patient retina. This is then viewed via a sight hole by the observer, usually through a hole in the mirror. A battery of lenses ensures adequate focus of the patient retina irrespective of patient and observer ametropia, and filters and stops allow a variety of fundus views. The positioning of the hole usually above the light source at the mirror leads to an uneven retinal illumination with inferior views often darker. This is exacerbated by the increasingly elliptical pupil as one looks further down into the inferior fundus leading to the familiar ‘shadow’ seen in many patients. Indirect ophthalmoscopy requires the observer to view an image of the fundus that has been projected to a point in space. This is usually by means of a high powered positive lens which therefore forms an inverted aerial image immediately in from of the lens held before the patient eye. The observer therefore requires some form of viewing instrument. This is usually a slit-lamp, but also head set viewing lenses are used. In this case the light source is nearer to the patient eye but laterally displaced such that the incident and reflected paths are different. This has the advantage of both increasing the fundus view through media changes (such as cataract or corneal transparency loss) and improving the contrast view of various lesions such as naevi which may not be visible under direct view. As will further be discussed, the technique also offers the advantages inherent in the viewing system, such as binocularity, filter and rheostat adjustment, easy image capture, and a more comfortable viewing distance (for both parties!). Direct ophthalmoscopy The direct ophthalmoscope still remains the favoured instrument of many practitioners. It is probably because of its ease of use and magnified view that this instrument has so long been at the vanguard of fundus examination. As this was the instrument that most practitioners used routinely during perhaps this is why it has become an inherent part of most clinicians’ armoury. The direct ophthalmoscope has manyPage 16 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 14. positive features and used in conjunction with an indirect ophthalmoscope,fundus examination can be optimised. Table 1 below lists some of themajor benefits and pitfalls of the direct ophthalmoscope.Table 1. Features of the direct ophthalmoscope Advantages Disadvantages High magnification (x15) Magnification varies with the patient’s refractive error. Higher in myopes. Relatively good image obtained Very low field of view (around 10º even through a small pupil in an emmetrope). Portable Monocular field of view Relatively inexpensive Can be difficult to use on uncooperative patients. View of fundus is the correct Image degrades with media opacities way up and non-reversed Poor at showing colour change or elevation. Very close proximity to patient requiredThe low field of view makes the direct ophthalmoscope difficult to usewhen scanning across the fundus, with only small areas being visibleat any one point in time (see figure 1). This becomes particularlyrelevant when an overall view of the posterior pole is required suchas in a diabetic. A small pupil and media opacities further heightenthis problem. The difficulties are minimised by dilation. Therefore, fora patient with small pupils and/or media opacities, pupillary dilation isrecommended to perform a thorough examination. Page 17
  • 15. Fig 1a. View of fundus as might be seen using a slit lamp with fundus viewing lens. Fig 1b. Shows a typical highlighted area as may be seen with a direct view of a high myope. Note the very limited field of view related to the ametropia. The monocular view of the direct ophthalmoscope renders judgement of depth difficult and the observer has to rely on other cues to form an opinion. A useful addition to most modern direct ophthalmoscopes is the inclusion of a slit graticule. This can be projected onto the fundus and any change in its regularity would indicate an area of elevation or depression.Page 18 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 16. Wide field direct ophthalmoscopyRecently launched instruments claim to overcome some of the shortfalls of the direct ophthalmoscope. An example is the PanOptic fromWelch Allyn. Using a patented optical system such instruments have 3main advantages of the conventional direct ophthalmoscope.These are: • An enhanced field of view in the undilated eye, typically quoted to be in the order of 25º. • A 26% increase in magnification. This results in the image being a quarter bigger than with the standard ophthalmoscope, thus improving resolution of the retina, hence allowing fine detail such as vascular changes to be more easily viewed. • An increase in the working distance between patient and clinician, allowing improved comfort for both.The optic of such systems differ from the conventional directophthalmoscope in that they allow light to converge to a point at thecornea and then diverge towards the retina, allowing for a wider fieldof view. In a standard ophthalmoscope, light is projected directly ontothe observer’s retina. The image produced is erect and not inverted.This point convergence makes the distance from the patient a crucialfactor in maintaining a good field of view and a common error amonginexperienced users of such instruments is to fail to adopt a closeenough viewing position and therefore never achieve the describedbenefits.Fundus contact lensesAlthough there are a number of lenses available on the market, probablythe best-known fundus contact lens is the Goldmann three-mirror lens(figure 2). Fundus contact lenses provide the best degree of stereopsisof all of the methods of examination. Additionally, they allow theobserver to view the extreme periphery of the fundus, as well as thecentral region. It is perhaps fair to say that these lenses haven’t really PAGE 19 Page 19
  • 17. taken off in popularity in general optometric practice, as they are perceived difficult to use. Moreover, the need to contact the surface of the eye (albeit with a coupling fluid in place) is off putting to the majority of practitioners. Fig 2. 3-mirror fundus viewing contact lens (seen in side profile in inset) But possibly the most significant factor as to why these lenses have now become relatively obsolete is because optical and government agencies have advised against the use of procedures that involve direct contact with the cornea when the device has already been in contact with the cornea of another person, unless required, because of a theoretical risk of transmitting the variant Creutzfeldt-Jacob disease (vCJD) prion. This was based on advice given by SEAC, the Spongiform Encephalopathy Advisory Committee to the Department of Health (DOH) in 1999. The College of Optometrists and Association of British Dispensing Opticians have recommended guidelines for sterilisation in response to the warning. However, the recommended decontamination procedure (one hour in 2% Sodium Hypochlorite solution followed with rinsing in sterile saline) is only for RGP contact lenses. The efficacy of the procedure for sterilisation of other instruments is still under research. Therefore, while this advice is still in place and alternative non-contact methods are available, these lenses are generally confined to the back of the drawer.Page 20 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 18. Slit lamp binocular indirect ophthalmoscopy (BIO)This technique has gained huge popularity in recent times, such thatmany practitioners have incorporated the use of hand-held indirectlenses to view the fundus into their general routine. It is also arequired General Optical Council core competency skill for qualifyingoptometrists. Indirect lenses are high-plus lenses that are used inconjunction with a slit lamp to produce a virtual image of the fundusthat is laterally reversed and inverted. Therefore the clinician that isnew to this technique has to adjust their observation to account for anupside down, mirror image.There are a number of lenses available on the market. They are availablewith a number of modifications. These include: • Yellow filters – that are either fixed or detachable. This reduces the amount of blue light impacting on the fundus, which is important in prolonged examinations; • Lid adapters – that help separate the lids and set the lens at the correct working distance; • Graticules – that are useful for measuring; or • Mounts – which steady the lens on the slit lamp.The best field of view is obtained by the higher dioptric poweredlenses, but at a compromise to working distance. Each lens producesa stereoscopic fundus view that is only marginally magnified, with thegreatest magnification being achieved with the lower powers. The slitlamp’s observation system is utilised to increase the magnification ofthe image. Table 2 below lists some of the properties of lenses that arecommercially available. The information has been collated from themanufacturer’s data. Note how, generally speaking, the higher the powerof the lens, the greater the field of view but the lower the magnification.This is important. It is tempting with a slit lamp to keep using onesingle viewing lens for all occasions and to adjust the magnificationusing the slit lamp. This, however, reduces the crispness of the image.It is therefore preferable to keep a lower power lens (say a 60D) foraccurate viewing of specific areas or lesions, such as disc or macular,and a higher powered lens (say a Superfield) for wider views, as wouldbe needed for a multifocal lesion disorder such as diabetic retinopathy.Most regular users would confess to owning at least two lenses. Page 21
  • 19. Table 2. A comparison of the different hand-held indirect lenses Power (D) Magnification Field of Working Indications view (º) distance for use from the cornera/mm 60 1.15x 76 11 Disc and macula 78 0.93 84 8 General screening 90 0.75x 94 6 General screening. Works well with small pupils SuperField 0.76x 95 7 General screening. Works well with small pupils Variable 0.77-0.94x 78-100 5 General screening zoom 132 SP 0.45 99 4 Wide field, small pupils Another simple point is that the lenses described with the ‘super’ prefix, such as the SuperPupil and the SuperField, are not equi-convex so need to be held with the tapered edge (figure 3) towards the patient. The other lenses may be held either way. Fig 3. A SuperField and Super 66 lens in side view. The tapered edge needs to point towards the patient.Page 22 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 20. Examination technique • For the best results, the pupil is dilated with an appropriate pharmaceutical agent(s). • Patient is instructed on the need for maximum compliance and asked to fixate the slit lamp fixation target. • The microscope is placed in the straight-ahead position with the lamp coaxial (allowing binocular viewing). The magnification is set to it lowest position and slit adjusted to a height around 5mm and width 3-4mm. • The slit is focussed on the patient’s cornea and centred. The lens is introduced close to the patient’s eye (figure 4). The lens needs to be orientated accordingly if it is not equi-convex as mentioned earlier.Fig 4. The lens is held before the patient’s eye. If not at the correct working distance for the lens, a full stereo view will be unlikely. Page 23
  • 21. • Each lens will have a working distance at which it should ideally be held from the patient eye. It is important to maintain this distance to offer a binocular view. The working distances vary from lens to lens but are generally between around 8 and 12mm, much less that most people might imagine. For this reason, many practitioners are happily using their lenses and achieving monocular views. Always check that the fundus view is visible to both eyes and if not, the lens needs to be positioned closer to the patient. • The slit lamp is moved away from the patient until the inverted image of the retina (usually a thin vertical strip at this stage) is seen (figure 5). At this stage, fine movements of the joystick allow finer focusing, and adjustment of the slit width and rheostat may be made optimal for the needs of viewing but also to minimise discomfort for the patient (the brightest widest image may not always be appreciated). Fig 5. The first view of the retina is usually a thin vertical strip. Accurate focusing and adjustment of brightness and slit width is essential at this stage.Page 24 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 22. • Focussing just before the retinal surface allows investigation of the vitreous, which can be particularly useful in posterior vitreous detachment (PVD) where a Weiss ring might be visible (figure 6).Fig 6. Weiss’ ring may be viewed by focusing anteriorly to the retinal plane • By bringing the lens closer to the eye the field of view can be increased. Any reflections from the slit lamp can be minimised or eliminated by slightly tilting the lens either vertically or horizontally. • The image can be optimised by adjusting the magnification and slit width/height on the slit lamp. • The eye is examined in various positions of gaze. The lens is repositioned each time to optimise the view. If the observer requires the image to move then the lens is moved in the same direction as the desired movement. Page 25
  • 23. A good systematic approach for full fundus examination might be; • Start at the disc and not all salient features – check that the view is stereoscopic as this will help to interpret disc topography. • Move the slit lamp nasally to the disc and cross the foveal region, noting all vascular, pigmentary and other features, and noting the foveal quality without dwelling there to avoid patient discomfort. This same method should allow scanning of most of the macular and paramacular and peripapillary region without patient eye of lens movement, bar any subtle tilting to minimise reflections. • Asking the patient to look up allows viewing of the superior retina. Once they look up their pupil will obviously move up too so the lens will accordingly have to be moved up until in front of the pupil at which point a retinal view will again be seen. Excessive reflections in this position may be minimised by maintaining coaxial illumination by tilting the lens top away from the patient slightly. The same should be repeated for all 8 positions of gaze. • Remember the view is laterally and vertically reversed. Some practitioners prefer to invert the record card when noting appearance. In the superior position of gaze, as described above, the view is of superior retina but the far periphery will be in the inferior field of the viewer and the mid-periphery in the superior field. When a patient looks nasally, nasal retina is viewed, the lens tilted so that the most nasal aspect of the lens is tilted away from the patient, and the far nasal periphery is in the temporal field of the viewer. Slit lamp BIO can be used to examine a patient with undilated pupils. The lenses, such as the SuperPupil which are specifically designed for use on a small pupil, work best in these instances. Nonetheless, a good view of the disc can be obtained in most patients. BIO has many advantages but also has some disadvantages. These are listed below in table 3.Page 26 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 24. Table 3. Benefits and pitfalls of slit lamp BIO Advantages Disadvantages Wide field of view making it Image is virtual, laterally reversed easy to scan fundus and inverted Image degrades significantly less Magnification and image quality with media opacities compared limited by slit lamp with direct methods Excellent in detecting stereoscopic Difficult to use on small pupils and colour changes Inexpensive (assuming slit lamp Difficult to use on non-compliant is available) patients Magnification does not vary with Unsuitable for domiciliary work, patient’s refractive error patient has to be upright Comfortable distance from Needs practice to become proficient patient maintained Can be uncomfortable for the patient Reflections reduce image quality Scanning required to build full pictureWhatever lens is used, it is essential to keep it completely clean at alltimes. A good quality lens cloth and cleaning fluid is advised. Alcoholbased cleaners are never a good idea (for example a Medi-Swab) asthey may dissolve the cement holding the lens in its case and result inthe all too familiar ‘rattle’ that many older lenses emit in use. Page 27
  • 25. Head-borne binocular indirect ophthalmoscopy The headset BIO is used in conjunction with a condensing lens and best results are obtained when the pupil is maximally dilated. The condensing lens, typically a lower powered lens such as 20D, collects the image of the retina from the patient’s eye. The optometrist views this image binocularly. Greatest magnification is achieved with a low powered condensing lens; however, this compromises the field of view. Higher-powered lenses have to be held closer to the patient. The field of view that is produced is approximately 10x wider than that achieved with direct ophthalmoscopy. The binocular viewing conditions ensure good stereopsis and there is negligible change in magnification with the patient’s refractive error. The image produced by a headset system is laterally reversed and upside down. The most peripheral part of the image is seen in the opposite direction to the patient’s gaze, so on looking up, the superior periphery is seen at the bottom of the image. The wide field of view produced by this technique makes it ideal for examining the peripheral fundus however; its low magnification can lead to subtle changes or small lesions being missed. The first time an instrument is used by a practitioner adjustment of the headband is required. This is done as follows: • Slacken the head size and height adjustment knobs and place the headband on the head like a hat. • Tighten the height adjustment control until the headband sits level on the head and is located about 2cms above the eyebrow. Tighten the head size adjustment knob to comfortably lock the unit into position. It should not feel excessively tight but be stable. • Slacken the browbar knob and the hinge adjustment control to locate the eyepieces level with the eyes and as close as possible to them. Tighten the knobs to maintain this position.Page 28 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 26. • Providing the unit is used by the same practitioner, the above should not need to be reset each time it is used. The oculars of the observation system then need to be set accurately for the examiner:• Look through the eyepieces at an object about 40cm away (eg one’s thumb).• Slide the two knobs for PD adjustment laterally while looking monocularly through each eyepiece so that view of the object is centred exactly in front of the eye.• Move the illumination beam so that it overlaps an object in the centre of the field of view through the eyepieces.• Select the appropriate beam size. Generally, the largest beam will be most suitable for peripheral examination. Smaller diameters are better for undilated pupils or high-power lenses. Put the diffuser (if available) into the system as it improves the peripheral view.• A filter can be selected if appropriate, although for peripheral assessment white light will generally be used. The observational technique requires practice and it will require several weeks of experience before the practitioner will be confident. The procedure is as follows:• Instruct the patient that their eyes are going to be examined using a bright light but that it will not be at all harmful. Both eyes should remain open throughout the exam.• Lie the patient down in a darkened room so that they are at waist height and their head is in front of the examiner.• Start by examining the superior periphery. This will entail the patient looking up and the examiner standing 180 degrees in the opposite direction ie towards the patient’s feet.• Once the red reflex is seen, insert the lens with the steeper curvature of the bi-aspheric lens facing the examiner. The side facing the patient has a silver or white band at the edge.• Hold the lens parallel to the patient’s iris plane and a short distance from the cornea with the fingers steadied on the patient’s forehead or cheek. Page 29
  • 27. • If necessary, hold the lids apart with the fingers of the other hand. • The anterior segment of the eye should be seen through the lens with a red reflex in the pupil. • Move the lens away from the eye until the fundus image fills the lens, but not so far that an inverted minified image of the anterior segment is seen. • Adjust the illumination, if necessary, to the minimum permitting clear visualisation of fundal details. • Progressively cover all peripheral quadrants by getting the patient to adjust the direction of gaze and moving to a position opposite to the gaze direction. The lens will need to be re-aligned each time. Tilting the lens can improve the image dramatically by moving the lens surface reflections out of the examiner’s way. • Finally, examine the posterior pole, limiting exposure time to reduce dazzle. • Release the lids before each re-fixation to allow the patient to blink and separate as necessary to improve your view. Scleral indentation This is usually essential should the examiner wish to examine to the extreme periphery. The compression of the sclera caused by pressure from a scleral indentor (figure 7) places the ora serrata within the field of view of the headset BIO. In addition, it permits a kinetic evaluation of the peripheral retina to be made. Fig 7. A scleral indentorPage 30 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 28. Any break in the retina located beyond the equator will move or bemore easily seen on indentation. However, it is not easy and shouldnot be attempted until the art of this type of BIO has been mastered.While scleral indentation does not make retinal lesions worse or causea detachment, it should not be attempted in any eye that has hadintraocular surgery in the past 8-10 weeks because the large increasein IOP it temporarily produces could cause damage. It should not beused if the eye has had a penetrating or orbital injury. The technique isas follows: • A thimble-type scleral indentor is placed on the first finger of the dominant hand. • The indentor should always be tangential to the eye and wherever possible the force should be applied through the lids, beyond the tarsal plates. • The patient looks initially in the opposite direction to the region to be viewed. The superior temporal region is easiest to examine. • Therefore, start with the patient looking down. Press the indentor on the upper lid behind the tarsal plate. Remember the ora begins at least 7mm behind the limbus. • With the indentor gently pressed against the eye, ask the patient to look up. At the same time advance the indentor parallel to the globe back into the anterior orbit. • While looking through the headset the indentor is pressed into the eye, applying force tangentially. The mound produced in the superior retinal periphery is seen like a mouse under a carpet and appears darker than surrounding tissue. • The indentor should be removed and the procedure repeated at each clock hour while working clockwise over the eye. The motion of the mound (or mouse) is observed, whilst keeping the examiner’s eyes, the condensing lens, the fundus image and the indentor all in a straight line. • At the three and nine o’clock positions indentation can be done by dragging the lids from the 10 or two o’clock position. If this is not possible then gentle pressure on the conjunctiva itself may be required, with or without anaesthesia. Page 31
  • 29. Modified (monocular) indirect ophthalmoscopy Examination of an uncooperative patient is difficult. Many of the techniques described above require patient compliance. The direct ophthalmoscope generally offers the best solution, but it too has difficulty – patients often fret as the clinician approaches and moreover, the patient is most likely to fixate the ophthalmoscope light. This generally results in good view of the fovea and macula but little else! The direct ophthalmoscope can be used in conjunction with a condensing lens. This provides the observer with a moderately magnified and wider-angle view of the posterior pole. An inverted, laterally reversed image is produced. A magnification of 4-5x is achieved which can be increased slightly by the practitioner moving closer It also allows the clinician to work at 20-50cms from the patient, thus allowing for the anxious patient. This technique is particularly useful for viewing the central fundus in children and with practice the practitioner is able to get a reasonable view of the disc and macula. The main disadvantage of this technique is the lack of stereopsis. However, lateral movement to induce parallax, can provide clues to depth perception.Page 32 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 30. Chapter 2 Contact Applanation TonometryWhen should IOP be measured? • There are many indications for tonometry to be carried out. Most practitioners, aware of the link with open-angle glaucoma, the major risk factor for which is age, measure IOP routinely on all patients over a certain age. 40 years of age is a cut off point where epidemiological studies suggest the incidence of glaucoma becomes significant. 1-2.1% of Caucasians and 4.7-8.8% of Afro-Caribbeans over the age of 40 years suffer the disease. The incidence increases dramatically to 3.5% Caucasians and 12% Afro-Caribbeans over the age of 70 years. These figures, and the great variability of the disease, are both reflected in the occurrence of open-angle glaucoma in the under 40’s, so this author measures IOP on all patients over 30 years of age. • It is also often useful to have a baseline measurement of IOP on all patients for interpretation of any future readings taken. • IOP is also routinely measured prior and subsequent to dilation. A rise of more than 5mmHg which fails to stabilise may be indicative of an induced angle closure requiring medical attention. One should also remember that cycloplegic drugs may interfere with aqueous flow and IOP measurement, even in the very young, may need to be considered. Page 33
  • 31. • Younger patients may present with signs that may indicate a risk of secondary open-angle or primary and secondary closed angle glaucoma, for example a very narrow angle or the presence of material on the corneal endothelium or in the anterior chamber. An IOP measurement would be useful in these instances. • Symptoms suggesting possible retinal damage (photopia, floaters) may prompt tonometry to see if there has been a significant fall in pressure. Techniques of measuring intraocular pressure There have been many methods used to measure IOP, some more successfully than others and those that are likely to be used or have been used widely in the past are mentioned below. Before discussing some mechanical methods available, some mention should be made of digital palpation. Despite this method being largely discounted as inaccurate by many authorities, the palpation of the superior sclera by the forefingers of both hands on the closed eyes is still used, often by practitioners who claim great accuracy. Certainly, a large difference between the eyes should be detectable to an experienced practitioner and there may well be situations, in an emergency for example, where it is the only available technique. Tonometry relies upon the application of external force causing deformation of the cornea (or the sclera in eyes where this is not possible such as when anaesthesia cannot be used and the cornea is grossly scarred) and relating the deformation to the eyes internal pressure. The tonometers may thus be classified according to the deformation produced.Page 34 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 32. Indentation or impression tonometryThese rely upon a plunger of variable weight to deform the corneaby a shape resembling a truncated cone. The level of indentation withdiffering weights is related to the IOP. The prototype and most famousindentation tonometer is the Schiotz. For many reasons, not leastpatient compromise and the effects of ocular massaging, this techniqueis now largely obsolete in the UK. There has been some renewed interest recently in indentationtonometry. The use of electronic measurement to assess recoil of amagnetically projected probe, so small that no anaesthetic is required,is employed in a new instrument (the i-Care tonometer) currentlyundergoing clinical evaluation at the time of writing. Another newproduct, the Diaton, uses indentation through the closed lid andassesses intraocular pressure without anaesthesia. Results vary with lidtension and this author believes the instrument to be best used if othertechniques are not available.Applanation tonometryAs suggested by the name, this technique relies upon the tonometerflattening or applanating an area of cornea. The weight applied may berelated to the pressure within and the area applanated by the applicationof the Imbert Fick Law (also originally known as the Maklakov-FickLaw).This law states that an external force (W) against a sphere equals thepressure in the sphere (P1) times the area applanated or flattened bythe external force (A). Page 35
  • 33. This physical law assumes that the sphere is perfectly spherical, dry, perfectly flexible and infinitely thin. The cornea fails to satisfy any of these criteria. It is aspheric, wet, and neither perfectly flexible or infinitely thin. The tear film creates a surface tension which has the effect of drawing the applied weight onto the eye (S), while the lack of corneal flexibility requires an extra force to deform the cornea. Furthermore, as the cornea has a central thickness of around 0.55mm, the area applanated is larger on the external surface (A) than the internal surface (A1). To overcome this, the Imbert-Fick law may be modified to; W + S = P1 A1 + B Where B represents the force needed to bend the cornea. When A1 is 7.35 mm_ then S balances out B and therefore; W = P1 An area of 7.35mm_ has a diameter of 3.06mm and the above cancelling out holds true for areas of diameter between 3 and 4mm. 3.06mm is useful because if that diameter is chosen then an applied force of 1g corresponds to a internal pressure of 10mmHg so making calibration of any applanating instrument easier. Furthermore, the volume displacement for this applanation is approximately 0.50mm_ such that ocular rigidity does not significantly affect the reading. Ocular massaging plays no part in applanation either. It is increasingly being recognised that the resistance to the deforming force offered by the cornea is strongly associated with the thickness of the cornea. The thicker the cornea, the higher the measured IOP irrespective of its actual value. For this reason, most practitioners requiring accuracy of tonometry also take a pachymetry reading. It is also useful to know by how much a patient’s IOP reading needs to be adjusted after they have undergone refractive surgery. Their thinner corneas will offer less resistance to the force of contact applanation and give an artificially reduced reading to that found prior to surgery.Page 36 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 34. The very first applanation tonometers employed a fixed weight appliedcause a variable area of applanation, such as the Maklakoff,Tonomat andGlaucotest instruments. These instruments are now rarely used as, witha lower IOP, a force can applanate a larger area so promoting aqueousdisplacement.Much more widely used and usually taken as the standard techniqueis that of Goldmann where there is a variable weight and a fixed(7.35mm_) area.Goldmann contact tonometerThis instrument is widely used and is generally accepted as theinternational standard by which other instruments are compared andwith which the vast majority of research in IOP measurement is carriedout.The applanation is caused by the probe that consists of a cone with aflat end containing two prisms mounted with their apices together. Oncontact with the cornea, the tear film forms a meniscus around the areaof contact and the ring so formed is seen by the practitioner throughthe probe. The split prism allows the ring to be seen as two semicircleswhich may be moved in position relative to one another by varying theweight of the probe applied to the cornea.This use of a vernier reading method adds to the accuracy of theinstrument and when the inner edges of the semicircles just touch thenthe diameter of the applanated is 3.06mm. The basic instrument itself,into which the cone is inserted, is basically a lever weight system with anadjustable scale, the scale calibrated in grams to allow varying force tobe applied to the cornea by the probe when the wheel is turned. Page 37
  • 35. • The basic procedure for Goldmann tonometer use should always involve thorough cleaning of the probe head. As well as the common infective agents one may come across in the tear film, there have been reported cases of hepatitis B and HIV viru isolated in the tears. The need for disposable contact apparatus in the aftermath of the variant Creutzfeld-Jacob Disease scare has led to most practitioners using a disposable tonometer head (the most common being the Tonosafe) which is placed within the housing and orientated appropriately. • Adequate sterilisation may be carried out with a weak solution of sodium hypochlorite. Though it is tempting to air dry the probe, dry residue of the cleaning agent has been found to occasionally aggravate the cornea so a rinse with saline is advised. • The probe should be inserted so that the white marking on the head aligns with that on the instrument to ensure that the split between the rings is horizontal. For astigmatic corneas (of greater than 3.00 DC), it has been shown that the probe should be aligned such that the interprism face is set at 43° to the meridian of the lowest power then the area applanated is still correct. A little red marker on the probe head allows this adjustment for ‘with the rule’ astigmats. • The cornea is anaesthetised (this author prefers 0.5% proxymetacaine which should cause less stinging and reactive tearing) and fluorescein instilled. Though difficult to quote an exact amount that would be meaningful in practice, too much fluorescein and therefore wide ring width will tend to give a high reading. If too little is instilled it is difficult to visualise the rings and vernier adjustment is less easy. It may be noted that some experienced clinicians carry out the procedure without any stain at all.Page 38 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 36. • The instrument is set on the plate on the slit lamp before the eye to be examined and this usually allows the probe to be directly slightly from the nasal aspect to allow the incidence to be on axis despite any slight convergence by the patient. To minimise the contact time with the cornea it is useful to set he instrument to a weight setting of 1g (10mmHg) or a previous reading if known. On low to medium magnification, the probe is viewed through the microscope with the cobalt blue light incident on the probe head, the lamp set at 60° temporally. • The probe is moved on to the cornea and the rings visualised and adjusted as mentioned. If the semicircles are not of equal size then small vertical adjustments may be needed. Always move towards the larger semicircle. If the semicircles are not equal, the reading will be too high. If they are apart, more force is needed, if overlapping, less is needed. See figure 8.Fig 8. Schematic representation of a variety of appearances of the meniscus rings; (a) the ideal alignment (b) more force needed (c) too much force used, reduce force (d) move cone upward (e) move cone downward (f) wait for fluorescein to drain Page 39
  • 37. • The probe should be removed from the cornea, the weight reading on the scale noted (in grams or X 10 for mmHg), and the cornea checked for staining. The time of the recording is also noted along with the instrument used. This is essential as very little may be discerned from an isolated reading bearing in mind the cyclical variation in IOP. • If staining is induced, most epithelial stain will disappear within a matter of several hours but caution by the optometrist should be exercised; if worried, monitor. For significant abrasions it is good practice to flush out residual fluorescein as this will reduce the risk of Pseudomonas infection. Simply hold tissues to the cheek of the patient and squirt in as much sterile saline as felt necessary (3 Minims is typically suitable). In the unlikely event of significant stromal fluorescence, an ophthalmologist might be contacted to ensure no further intervention is needed. Prolonged contact should be avoided to minimise corneal compromise, but it does allow the practitioner to visualise the ocular pulse, seen as small oscillations of the semicircles relative to each other. Prolonged contact also has the effect of reducing IOP by an ocular massaging process. A reduction in this effect or a marked difference between the ocular pulse in each eye may be indicative of vascular occlusive disease. The thickness of the cornea may also cause some error as a very thin cornea will produce low IOP readings. A very irregular cornea may make visualisation of the rings difficult but a contact method is probably more accurate in this case than a non-contact method. In cases of extreme corneal vulnerability or where anaesthesia is not advisable, the contact methods may not be first choice. As with any accurate measuring instrument, regular calibration is necessary and is easily carried out by the practitioner. For the Goldmann this requires the use of a calibration rod which is held in a small screw attachment which in turn inserts into the housing on the side of the instrument. The following needs to be carried out regularly, particularly if the instrument has not been used for a while;Page 40 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 38. • Set up the Goldmann unit complete with tonometer prism on the slit lamp as if to use it to take a measurement. • Position the rod and lock so that the central marking on it is in line with that on the holding unit. In this position the rod exerts no force on the prism holder. • The dial should be gently turned from less than ‘0’ until the head is seen to move forward. The force for this to occur, ideally around 0.1g, should be noted. The force should then be reduced and the force noted when the head moves backward again, ideally -0.1g. The middle of the two readings in this case is 0. • The same is done for the rod positioned with the second marker in position, the rod being pulled away from the direction of the prism such that 2.1g is needed to move the head forward and 1.9 for it to move back (a 2g averaged reading). The last line on the rod does the same for 6g of force. • If the rod moves at 0.3g, 2.3g, and 6.3g, then it is still a usable instrument but the operator needs to remember to reduce each reading by 0.3g (or 3mmHg). • If the rod moves at 0g, 2.3g and 6.6g, then there is clearly a stretched spring and the instrument should not be used.Perkins contact tonometerThis instrument was developed as a hand-held, and therefore portable,version of the Goldmann. Having its own light source and viewinglens negates the need for the slit lamp (see figure 9). The probe isheld on a counterbalanced mounting with a coiled spring that allowsthe instrument to be used accurately in either the horizontal orvertical position, useful for the supine patient in a domiciliary visit. Thismechanism originally required a slightly differently weighted probe tomaintain accuracy, denoted by a red ring marking as opposed to theblack ring marking on Goldmann probes. The advent of newer designsof both instrument and probe has done away with this discrepancy.The instrument has a headrest attachment that may be extended andheld in position on the patients’ forehead to minimise instrument shake,possibly one of the main problems with this instrument. The Mark IIversion now available has 2 light sources for ease of viewing and may Page 41
  • 39. Fig 9. be fitted with a magnifying device (the Perkins Examination Telescope) allowing viewing of the rings at arm’s length. The technique is similar to Goldmann but with a number of variations. • Patient and prism preparation is the same as for the Goldmann. The patient should be positioned with their head against a head rest to minimise movement and at a height that the practitioner is comfortable with. Too awkward a position when bending and there is a tendency towards shaking. • It is essential to set the force at the level expected in the eye, based on a previous reading or on a non-contact reading if already available. This minimises instrument adjustment on eye which is more likely to cause staining with the Perkins. • The forehead rest (if used; some practitioners prefer to remove this and use their hand as fulcrum) needs to be set at a length individually for each patient such that, when pivoted about the forehead, the prism will meet the cornea flat on. For deep set eyes, the arm needs to be shorter and vice versa for more protruding eyes. • It is as important to be as close to the viewing position for ring interpretation as possible before applanating. Large movements towards the eyepiece while on the cornea invariably leads to the prism drifting off the cornea or at least moving about more than desired. • Once the rings are viewed, unless very close to their final alignment, it is usually best practice to withdraw from the cornea, make the necessary adjustment, and then return. Avoid undue time on the cornea with the Perkins.Page 42 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 40. As already stated, Tonosafe probes may be used with this instrument.Calibration should be carried out regularly and may be done by thepractitioner. It is similar to the Goldmann technique in that it relies onthe use of 3 artificial forces, in this case supplied by weights. • The instrument should be fitted with a prism and laid on its back, prism upwards. O ensure that the force is vertically applied, some authorities suggest placing the supplied black disc beneath the head of the instrument (or the metal cylinder for the older Mark 1 instrument) having first removed the battery pack from the handle. • The prism should first pop up just after 0g is set and back down just under 0g. • A 2g and a 5g weight should in turn be balanced on the prism (obviously rendering it non-sterile and unusable) and the force needed to pop up and down the prism should be just over and under 2 and 5g respectively. Constant error may be adjusted in the final reading, whereas a non-linear error means the instrumentis unusable. • As with the Goldmann, a good examination of the cornea is important subsequent to the technique but in situations where this is not possible (Perkins is often used in domiciliary settings for instance) then a gross view should be used and acuity taken. Advice should be given about reporting any symptoms of discomfort or blur.Not surprisingly, the results gained are comparable to the Goldmann.Tonopen XL mentorThis is a small, portable electronic contact tonometer which has astainless steel probe which, on contact with the cornea, measures IOPby an electronic signal from a solid-state strain gauge held within. Theshort contact time requires several readings to be taken and averagedbut this, combined with its small appearance, appears to make it patientfriendly.Results appear to be reliable in comparison with the Goldmann, albeitwith some greater spread as may be expected in any instrument takinga series of instant readings compared with one prolonged reading. Page 43
  • 41. Non-contact tonometers (NCT’s) The NCT was first introduced in 1972 and has the obvious advantage of not requiring contact with the cornea. This minimises the potential for corneal compromise, negates the need for anaesthesia, and in many cases is preferred by the patient. Automation of the mechanism means that the instruments are often used by non-professional staff as a screening instrument. There are a wide variety of instruments on the market, and rarely a year goes by without some new upgrade of an instrument. The basic operation of the instrument relies upon the applanation of an area of cornea by a jet of air which is usually switched on and off at a point dictated by the quality of image of reflected light from a source on the tonometer. The instrument then electronically converts either the time taken to flatten the cornea or the pressure of air needed into a reading of intraocular pressure. Because the instruments rely upon the flattened cornea acting to reflect incident light into a receiver, anything compromising the reflection (such as heavy scarring or distortion) will impact upon the accuracy of the reading. However, most designs include an ‘override’ feature allowing a reading to be made even if the received light beam is not maximal. This is also useful if the patient has poor fixation or rapid blinking. As all the NCT’s take an instantaneous reading of IOP, there is some variation between readings. If the reading is taken on the peak of the ocular pulse, respiratory cycle and immediately after a blink then a disproportionately high reading may be found. It is standard practice, therefore, for the NCT to take 3 or 4 consistent readings before arriving at a useful value. Most instruments automatically calculate an average. A recent innovation has been to incorporate into the patient forehead rest a sensor which monitors the patient pulse. The thinking behind this is to only allow a reading to be taken at a specified point on the cardiac cycle so overcoming any variation due to the ocular pulse. It does not, however, have influence over the other physiological variations mentioned earlier so the days of a ‘single puff ’ NCT are not here yet.Page 44 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 42. The overall perception of NCT’s giving higher readings is likely to bebased on practitioners using an inappropriate average or not takingenough readings. In fact any significant studies have suggested if there isany difference it is the NCT being lower than the contact method.There is now a substantial body of literature comparing results takenfrom the various NCT’s with the Goldmann standard, certainly withinthe non-hypertensive population. At higher readings, say above28mmHg, the correlation breaks down and NCT readings seem lessconsistent, despite most incorporating adjustments to deal with highervalues. It is for this reason, which is becoming increasingly debatable, thatmany ophthalmologists recommend referral of consistently measuredhigh IOP taken with a contact method, or that high NCT readings beconfirmed with a contact method before referral.There are many NCT’s available, a full description of each being outsidethe scope of this book. Most incorporate a calibration mechanism anda demonstration setting to allow the patient to know what to expectfrom the “puff of air”. This action usually also serves to clear any dustparticles from the air chamber which could otherwise be transferred tothe patients cornea. Most use a flashing non-accommodative target toreduce the variation due to accommodation and wandering fixation.Corneal hysteresisAs already stated, the deformation of the cornea due to an appliedforce is, to a certain extent, related to the corneal thickness as wellas the IOP within. The deformation is visco-elastic rather than trulyelastic, meaning that the recovery of the cornea to its original shapewhen the deforming force is reduced will occur at a different rateto the original deformation. This is measurable if the force to causeapplanation is measured and the cornea further deformed to a concaveshape before the force is reduced and that at which the applanatedposition is reached again also measured. The difference between thesetwo forces represents what is known as the corneal hysteresis and maybe used to indicate the resistance of the cornea to force and so toadapt a tonometer reading. An instrument, such as the Reichert OcularResponse Analyzer, attempts this and clinical trials will soon show thevalidity of such a measurement in predicting actual IOP. Page 45
  • 43. Ocular blood flow tonometry It has already been suggested that that is a close correlation between the IOP and the ocular blood flow. Over recent years a contact tonometer has been developed, the Ocular Blood Flow Tonometer, which measures IOP, and the ocular pulse in terms of pulse amplitude, pulse volume and pulsatile blood flow. In contact with the eye, the machine samples over 200 readings in a second over a period of 7-10 seconds so allowing for several beats of the heart. As well as providing accurate IOP information, the data regarding haemodynamics provides useful baseline data allowing for interpersonal comparison and changes over time to be analysed. At present, such measurements are rarely carried out outside specialist clinics. Analysis of results It would be a very easy, if possibly boring, world if it were possible to have a list of situations which would warrant referral, including a cut-off IOP measurement. It should now be clear that IOP alone as a predictor of eye disease is not reliable and that referral in the absence of any other risk factor is only agreed to be necessary at consistently very high values. A safe policy is for the optometrist to find out what local hospital policy is regarding referral on IOP alone. In the practices this author works in, local ophthalmologists are happy to see patients with repeatable IOP readings of over 25mmHg, and consider treatment if verified to be over 30mmHg. It is not unheard of for the patients in the 25-30mmHg group to be checked and discharged with advice to attend regularly at the optometric practice for IOP monitoring. Many surgeons now treat ocular hypertension of 25mmHg or more with hypotensive agents in the absence of any signs of glaucoma, particularly in high risk groups such as Afro-Caribbeans. This is in the light of the Ocular Hypertension Treatment Studies suggesting such action reduces significantly the risk of any progression in such patients to glaucoma. A very high IOP reading in an asymptomatic eye may occasionally occur in chronic secondary glaucoma conditions, such as Posner-Schlossman syndrome, where the history of inflammatory activity may be vague and the IOP rise only transient.Page 46 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 44. It is also a significant finding if there is a significant and repeatabledifference between the IOP’s of each eye (5mmHg or more), if thisfinding cannot be related to previous history, such as intraocular surgery,or a significant anisometropia.As already stated, an unusually low IOP (less than 8mmHg for example)may be clinically.The key points to remember are repeatability and that the IOP fluctuates.A repeatable reading of 26mmHg is significant. It may be that this isthe peak of a cycle averaged at around 23.5mmHg and this may beless significant. Interpretation needs to be set against repeated readingsas well as all the other risk factors known to influence glaucomatousprogression. Page 47
  • 45. Page 48 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 46. Chapter 3 Gonioscopy And Anterior Angle AssessmentAssessment of the anterior chamber depthPractitioners consistently dilate pupils in everyday practice and needto take some precaution before undertaking this procedure. As wellas a measure of IOP, some assessment of the calibre of the anteriorchamber is needed to predict the rare occurrence of angle shutdownupon dilation.Direct viewing of the anterior chamber angle itself, because ofthe refraction of light by the cornea, requires the use of a contactgonioscopy lens. With the current concerns regarding repeatable-usecontact lenses for clinical assessment, many practitioners who workoutside hospital departments prefer to use other methods of anteriorchamber assessment. Perhaps the most widely used and best-knowntechnique is that first described by van Herick, a brief revision of whichis outlined here.The ease with which van Herick’s method may be carried out as part ofa routine slit-lamp evaluation has led to its almost universal use.This has,to some extent, overshadowed various other ways to assess anteriorchamber depth, some descriptions of which are also given. The patencyof the anterior chamber is therefore assessable by several means ofdiffering accuracy and ease of use; • Depth assessment – Smith’s, pachymetry, ultrasound, Scheimpflug • Angle estimation – van Herick’s, eclipse method • Angle appearance – gonioscopy, ultrasound Page 49
  • 47. Each technique has its merits. In a general routine, van Herick technique is perhaps the easiest method of assessing the openness or otherwise of the anterior chamber angle. It is essential to do this prior to dilation, but also in patients where there may be assumed some risk of angle closure, such as high hypermetropes. If there is a suspicion of any angle narrowing, such as reported transient prodromal symptoms (haloes around lights, for example), or if there is evidence of the possibility of a secondary open angle episode, such as keratic precipitates, then gonioscopy is recommended. With both Smith’s and, particularly the more commonly used van Herick, it is important to remember the limitations of the techniques. As the most likely area of the angle to initially shut down is the superior region, all these techniques can reveal are whether there is an unduly shallow chamber or if the mid-peripheral iris is closer to the cornea than is typical. It is perhaps most useful if the following are identified; • Changes (that is reductions) in the grade or depth over time. A chamber graded with van Herick as 4 a year previous to being graded 3 would be more significant than an angle graded as 2 over many years. • Large differences in the grading between the two eyes which cannot be explained on high anisometropia. Van Herick’s technique It is important to remember that van Herick’s method is qualitative in that it does not directly measure anterior chamber depth, but gives a grading of the depth, and so allows some prediction of the risk of angle closure. A typical procedure is as follows: • The slit-lamp magnification should be set at ten to sixteen times to allow an adequate depth of focus. • With the patient positioned comfortably and staring straight ahead towards the microscope, the illumination system is set at 60° temporal to the patient’s eye. The angle is chosen so that the illuminating beam is approximately perpendicular to the limbus and, as the angle is constant whenever the technique is used, this enables consistency of interpretation every time the patient is assessed. • A section of the cornea as close to the limbus as possible is viewed.Page 50 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 48. A comparison is made between the thickness of the cornea and thegap between the back of the cornea and the front of the iris where thebeam first touches (figure 10).Fig 10.The van Herick appearance showing the corneal section, gap and lighton the iris. In this case, the corneal section is…. • The ratio of these two measurements may be graded and interpreted, as outlined in Table 4. An increasing number of practitioners use a decimal score system whereby 1:0.25 is written as 0.25. This allows a greater lexibility to distinguish between a chamber that is say 0.6 or 0.9, both scored as 3 under the traditional system.Table 4. Cornea/gap ratio Grade Angle 1:1 or more 4 Open 1:0.5-1.0 3 Open 1:0.25-0.5 2 Closure possible 1:less than 0.25 1 Closure likely No gap 0 Seek ophthalmologist advice Page 51
  • 49. • Many authorities recommend a measure at the nasal limbus also, and that if a large variation is found the narrower angle be considered. When the nasal measurement is carried out on many patients, to avoid the patient’s nose with the light beam is problematic. Thus, it is often useful to lock the angle between microscope and illumination system at 60°, swing the whole slit lamp temporally and ask the patient to maintain fixation on the microscope. As the nasal angle is expected to be wider, a narrow nasal angle might be considered as more significant. • Elderly patients, or those with corneal disease, are often difficult to assess because of the lack of transparency in the peripheral cornea. The temptation is to move the beam further onto clear cornea but, of course, if the assessment is not made at the very limbus, the anterior chamber gap is always wide and hence the grade will always be over-estimated. In such cases it is far better to make a judgement in the correct position and then to qualify the reading in the records with a note about the poor image see Smith’s method First described by Smith in 1979, this technique has been somewhat overlooked, but differs from van Herick’s in being quantitative, and so provides the practitioner with an actual measurement (in millimetres) of the anterior chamber depth. The procedure is carried out as follows: • The microscope is placed in the straight-ahead position in front of the patient, with the illumination placed at 60° temporally. To examine the patient’s right eye, the practitioner views through the right eyepiece, and for the left eye through the left eyepiece. • A beam of moderate thickness (1–2mm) is orientated horizontally and focused on the cornea. In this position, two horizontal streaks of light are seen, one on the anterior corneal surface and the other on the front surface of the crystalline lens. • Altering the slit-height adjustment on the instrument is seen as a lengthening or shortening of the two horizontal reflexes. • Beginning with a short slit, the length is slowly increased to a point at which the ends of the corneal and lenticular reflections appear to meet.Page 52 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 50. Fig 11. The appearance of the horizontal slit reflections from the cornea and the lens. The height is adjusted until ends meet. • The slit length at this point is then measured (it is assumed that the slit lamp is calibrated for slit length). • This length may be multiplied by a constant to yield a figure for the anterior chamber depth (Table 5).Table 4. Conversion of slit length to anterior chamber depth Slit length (mm) Anterior chamber depth (mm) 1.5 2.01 2.0 2.68 2.5 3.35 3.0 4.02 3.5 4.69For an angle of incidence of 60°, simple geometry allows the depth ofthe anterior chamber to be calculated by the formula:AC = AC/sin 60°Note that the diagram does not include the change in the incidentbeam through corneal refraction, and corneal curvature does introducea variable into the calculation. Studies that compared the anteriorchamber depth measured by the pachymeter with that by the slit-length method showed a constant ratio of 1.4, which approximates wellwith the reciprocal of sin 60°. Page 53
  • 51. When Smith’s method is compared with pachymetry the constant is around 1.31, a little different, but still predicting an accuracy of _0.33mm in 95 per cent of the samples assessed. Ultrasonography measurements predict an accuracy of +/-0.42mm in 95 per cent of cases.The constant given by the results using ultrasound, and hence considered to be the more accurate, is that used to calculate the values given in Table 2. From a clinical point of view, a chamber depth of 2mm should be treated with caution when considering pupil dilation. Adapted Smith’s method An adapted version of the above method has been proposed for use in those situations in which variable slit height is not possible. In this case, the slit length is noted prior to measurement at a point where it corresponds to 2mm. The instrument is then set up exactly as in the Smith method, but with the initial angle of incidence at 80°. By gradually closing the angle, the two reflected images on the cornea and lens appear to move closer and the angle at which they first touch is noted. A corresponding value for the chamber depth for differing angles is then calculated as given in Table 6. Table 6. Conversion of slit length to anterior chamber depth Slit lamp angle (degrees) Anterior chamber depth (mm) 35 4.22 40 3.87 45 3.52 50 3.17 55 2.83 60 2.48 Verification of the accuracy of this result with pachymetry results showed a good correlation between the methods.Page 54 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 52. Gonioscopy (adapted from Eperjesi F, Harvey W; InvestigativeTechniques and Ocular Examination. Ed; Doshi S & Harvey W. ElsevierScience Publications)Gonioscopy is a technique that allows practitioners to examine anarea of the eye otherwise hidden from view. The anterior chamber wallcannot be directly observed externally because of the limbal overhang,and as the observer moves parallel to the iris surface, the light rays fromthe chamber angle undergo internal reflection, preventing the light fromleaving the eye. The word gonioscopy, derived from Greek, means toview the angle. This procedure allows evaluation of the width of theanterior chamber angle, detailed inspection of the structures within theanterior segment of the eye prior to dilation and the investigation ofocular signs or symptoms. It has been described by some authoritiesas an integral part of any complete ophthalmic evaluation. However,gonioscopy can initially seem difficult.Critical angleA critical angle occurs when light from an object passes from a mediumof higher refractive index to another of lower refractive index. Theangle of refraction in the lower refractive index medium increases morerapidly than the angle of incidence. At certain angles of incidence, theangle of refraction will be 90 degrees and the refracted light will runparallel along the surface. This angle is known as the critical angle. Lighthitting the surface at an angle greater than the critical angle will notenter the second medium and will be reflected at the surface. In theeye, the light rays from the anterior chamber strike the anterior cornealsurface at an angle greater than the critical angle, estimated at 46-49degrees, and as such are totally internally reflected.When a goniolens is placed on the eye, the air interface with the anteriorcornea is replaced by the lens surface with an index of refraction similarto the index of refraction of the corneal tissue. The critical angle iseliminated because the light is refracted by the steep curved outersurface of the goniolens. Light from the anterior chamber can then bereflected by a mirror into the examiner’s eyes. Page 55
  • 53. This is the basic concept by which indirect goniolenses, used in conjunction with a slit lamp, allow visualisation of the anterior chamber angle. The image produced is indirect, virtual and unmagnified and not laterally reversed about the vertical. Anterior chamber gonio-anatomy The purpose of gonioscopy is to permit visualisation of the irido- corneal angle. It is better known as the anterior chamber angle and will be referred to as the angle in this article. This is the area in which the trabecular meshwork lies, and is therefore responsible for aqueous outflow. Before examination of the angle is discussed, it is important to review the anatomy and function of its structures. The appearance of the anterior chamber angle varies according to congenital individual differences and with acquired changes due to age, injury or disease. >From anterior to posterior the gonio-anatomy consists of 10 normally visible structures: the cornea; Schwalbe’s line; Schlemm’s canal; trabecular meshwork; scleral spur; ciliary body; iris processes (not always present); iris root; iris surface; and the pupil border. The most important are described in detail below. Fig 12. The anatomy of the Anterior AnglePage 56 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 54. Schwalbe’s lineSchwalbe’s line is a condensation of connective tissue fibres. It forms thetermination of Descemet’s membrane and marks the transition fromthe transparent cornea to the opaque scleral tissue. It also forms theanterior boundary of the trabecular meshwork. It is a difficult structureto find and may be barely distinguishable. It appears as a thin, whiteglistening line of variable thickness. Its position can be identified usingan optic section, which creates focal lines representing the anterior andposterior cornea. These focal lines merge at the end of the internalcornea and both denote the internal limbus as well as mark the locationof Schwalbe’s line.Trabecular meshworkThe trabecular meshwork runs from Schwalbe’s line to the anteriorlimit of the cornea, where it inserts into the scleral spur posteriorly.In cross-section, the trabeculum forms an obtuse angle, with the basein contact with the scleral spur, the anterior face of the ciliary body,and the iris root. The apex terminates at the corneo-scleral boundary– Schwalbe’s line.The trabecular meshwork can be arbitrarily divided into anterior andposterior portions. The former runs from Descemet’s membrane tothe area anterior to Schlemm’s canal. This is considered to be thenon-filtering portion of the meshwork because it has no contact withSchlemm’s canal. The posterior portion is involved in filtering andoverlies Schlemm’s canal. The filtering portion has a grey, translucentappearance. In older or in heavily pigmented eyes, the filtering portionof the trabecular meshwork is often pigmented and darker than theremainder of the meshwork. This helps to identify the portion ofthe meshwork involved in filtration and in localisation of the canal ofSchlemm, which lies internal to the meshwork at this point. In eyeswithout pigmentation, the filtering portion can be identified by itsknown location from the scleral spur and Schwalbe’s line, as well as byits grey, granular appearance in comparison to the white scleral spur. Page 57
  • 55. Schlemm’s canal Schlemm’s canal lies within the trabecular meshwork and is approximately 0.2 mm to 0.3 mm wide. It cannot be detected by gonioscopy unless it is filled with blood, which may appear as a pinkish band anterior to the scleral spur. Normally there is no visible blood in the canal because the pressure within the canal is lower than the pressure in the anterior chamber. The canal of Schlemm connects with the venous system of collector channels anastomosing to form a deep scleral venous plexus. This plexus drains the aqueous into the episcleral veins and anterior ciliary veins. Scleral spur The scleral spur is a fibrous ring that is attached anteriorly to the trabecular meshwork and posteriorly to the sclera and longitudinal portion of the ciliary muscle. It is paler than any of the other structures in the angle, making it the most easily recognised landmark. Its identification and visualisation confirm that the filtering meshwork is not obstructed in the area of observation. Ciliary body The ciliary body lies posterior to the scleral spur and varies in colour from grey-white in light eyes to darker grey in pigmented eyes. It may also appear reddish brown, chocolate brown, or reddish-grey. Its many functions include: manufacture of aqueous humor; regulation of aqueous outflow; control of accommodation; secretion of hyaluronate into the vitreous and the maintenance of a portion of the blood- aqueous barrier. Although most outflow of aqueous occurs through the trabecular meshwork, approximately 10 to 30 per cent is by a non-conventional route, primarily through the ciliary body face into the suprachoroidal space. Uveo-scleral outflow is pressure-independent, unlike the trabecular outflow. Because it is pressure independent, it appears to have more significance with a lower intraocular, probably because the trabecular system can autoregulate and reduce flow dependent intraocular pressure.Page 58 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 56. Iris processesIris processes can appear as: thin, fine lacy fibres; tree-branch likestructures; a coarse dense, interlacing network; or as a membraneattaching the iris to the trabecular meshwork. The reason for theiroccurrence is unclear but they are known to be physiological. Sometimesiris processes can be confused with peripheral anterior synechiae, whichare forward adhesions of the iris tissue to the angle wall, and tend tobe full-thickness attachments of actual iris tissue that rarely bridge theangle. Compression gonioscopy can be useful in differentiating betweenthe two: a synechiae will continue to hold against the wall, whereas thenetwork of iris processes will become lacier in appearance, exposingthe wall behind them.Synechiae result from the iris being placed abnormally against theangle wall. There may be pigment scattering when there is evidence ofsynechiae.This is characteristic in cases of post inflammatory conditionssuch as uveitis or acute angle-closure glaucoma. According to somespecialists, synechiae are never found in primary open-angle glaucomaor pure exfoliative or pure pigmentary glaucoma. Their presence isindicative of another abnormality. Peripheral anterior synechiae can bean attachment to the ciliary body, or it can be as high as Schwalbe’s lineor any structure between the two. The low, broad-based attachmentsmay be difficult to recognise but are critical for identifying early chronicangle-closure glaucoma. Synechiae interfere with aqueous outflow,whereas iris processes do not.IrisThe iris generally inserts at a variable level into the face of the ciliarybody, posterior to the scleral spur. Less commonly, the iris will inserton, or anterior, to the scleral spur. The iris thins at the periphery nearits insertion. Page 59
  • 57. Types of lenses Gonioscopy is carried out using goniolenses of which there are two main types - direct and indirect. Indirect mirrored lenses are more commonly used and the rest of this article will focus on this type. Indirect mirrored goniolenses are plastic cone-shaped contact lenses that are available in various sizes and designs and contain one, two three or four mirrors plus a plano-concave lens in the centre of the cone apex, which effectively eliminates the cornea as a refracting surface. Lenses may be obtained with or without a flange (lip). Some authorities claim that flanged lenses are easier to insert and keep in the eye. For some indirect lenses the small difference in curvature between the lens and the cornea is minimised by interposing a viscous coupling gel between the two surfaces (for example, Viscotears by Ciba Vision, proprietary artificial tears such as methylcellulose, or a contact lens wetting solution). This solution needs to be optically homogeneous and have a refractive index similar to that of the cornea.There may, however, be a transient superficial punctate corneal epithelial stain, which resolves spontaneously after 30 to 60 minutes, probably due to a combination effect of the preservative in the gel and pressure from the lens being held on the cornea. The single-mirror goniolens is useful for teaching purposes and for those new to the procedure. The multi-mirrors of the two-, three- and four- mirror versions can prove confusing for the novice. The single-mirror lens is considered a good choice for children, and adults with small palpebral apertures. In its three-mirror form, the indirect goniolens (often known as the Goldmann 3-mirror, see figure 2) permits angle examination using the smallest arc-shaped internal mirror. A good view of the mid- and far-periphery of the fundus using the additional mirrors can be obtained through a dilated pupil. As both the single- and Goldmann three-mirror lenses only have one mirror positioned to view the angle, rotation through 270 degrees with an appropriate adjustment to the illumination is required. The slit beam should always be approximately perpendicular to the base of the arc-shaped mirror.Page 60 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 58. The Thorpe four-mirror goniolens is designed to provide a view of theangle in each mirror.The mirrors are angled at 62 degrees and show theanterior chamber angle in normal size. This is useful as the lens only hasto be rotated slightly for the complete angle to be observed. This lensis the preferred choice of many practitioners for assessing the integrityof the angle. The mirrors are housed in a large cone, which is easier tomanipulate by people with large fingers and hands who might find thesmaller lens types difficult to control once on the eye. It does however,require an optical coupling gel.Sussmann and Volk four-mirror lenses have a small contact surface so acoupling gel is not required: the lens is coupled to the eye by the tearfilm. The author has found this type of lens difficult to keep on the eye,especially with patients who are eye squeezers. All goniolenses enableexamination of the posterior pole through a dilated pupil using thecentrally mounted plano-concave lens.Fig 13a. The Volk 4-mirror gonioscope requires no contact fluid. The Ocular Instruments 2-mirror gonioscope has a larger contact area and therefore requires aFig 13b. coupling fluid. Page 61
  • 59. Clinical procedure As with all invasive ocular surface procedures, corneal health should be assessed with fluorescein before examination.The procedure should not be carried out on a compromised cornea or on an eye that has suffered recent trauma, especially if the globe has been perforated. Tonometry should be performed before gonioscopy, as excessive pressure on the eye can artificially lower intraocular pressure. There are two psychological barriers that need to be overcome before obtaining good results with the goniolens. With practice the practitioner will appreciate that the technique is not as difficult as first perceived, that the lens will not damage the eye and that the patient will not feel any discomfort when it is on the eye. It is good practice before commencing gonioscopy to explain what is going to happen and why it is being done since the size of the lens can disconcert the patient. The patient should be advised that the lens may feel strange but will not be uncomfortable. This will usually reduce patient anxiety and improve co-operation. • The cornea should be anaesthetised with one drop of topical anaesthetic (for example, benoxinate 0.4 per cent or proxymetacaine 0.5 per cent). If gonioscopy is performed immediately after contact tonometry then no further anaesthetic is required. • The patient should be positioned comfortably at the slit lamp, with the lateral canthus of the patient’s eye aligned with the marker on the slit lamp. This will permit sufficient vertical excursion of the slit lamp to enable the superior and inferior mirrors of a multi-mirrored lens to be viewed. The magnification should be set at approximately X10 -15 and the illuminating slit beam to 2mm width and 2-4 mm in length, with the rheostat at a medium setting. If excess light enters the pupil, pupillary constriction may make the angle appear more open than it is under general lighting conditions. The position of the microscope needs to be straight ahead and the light source directly in front of the microscope so that neither eye is occluded.Page 62 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 60. • The angle is observed by reflected light and the mirror is placed opposite to the section being observed. When the mirror is superior, the angle being observed is the inferior angle. The angle under observation is inverted but not reversed laterally. In the case of narrow angles, the angle recess can be easily observed by tilting the lens to the angle of observation and having the patient rotate the eye toward the gonioscopy mirror.• To avoid or reduce glare from the front surface of the lens, room lights should be dimmed, the illumination arm moved slightly to one side by 5-10° or the lens tilted along the axis of base of mirror. For example, when the mirror is superior the lens is tilted to the left or right, and when the mirror is temporal is tilted up or down.• Deliberate compression (dynamic gonioscopy) gives the observer a certain amount of control over the iris configuration. In an eye with a relatively narrow angle, deeper structures can be visualised by flattening the periphery of the iris gonioscopically. The lens can be shifted or slightly rocked on the corneal surface in any direction for a millimetre or so. Manipulation of the lens in this way will allow the practitioner to distinguish between true peripheral anterior synechiae and simple apposition of the iris to the cornea.• The goniolens can be removed by parting the lids so that they clear the lip of the lens. In most cases the lens will fall from the eye. If the lens remains attached to the cornea by capillary attraction, the lateral sclera adjacent to the rim of the lens should be pressed firmly with the tip of one finger to break the capillary attraction holding the lens in place. Page 63
  • 61. Fig 14. The appearance of a normal, open angle in gonioscopy. Cleaning and disinfection As soon as the lens is removed from the patient’s eye, it should be thoroughly rinsed in cool or tepid water and cleaned by hand using a contact lens surfactant. The lens should then be rinsed with saline and dried with a non-linting tissue. Household bleach can be used to disinfect the lens. The lens is soaked for a minimum of 20 minutes in a mixture of one part bleach to nine parts cool/tepid water, rinsed thoroughly with cool or tepid water, dried and placed in a dry storage case.These lenses should never be steam autoclaved or boiled. Acetone, alcohol and hydrogen peroxide will damage the lens. Indications for gonioscopy Some optometrists are hesitant to dilate pupils for fear of precipitating an attack of closed-angle glaucoma. Van Herrick’s slit lamp technique (see later in this chapter) allows an approximation of the openness of the angle, however the use of a goniolens will provide a more accurate evaluation of angle width as well as assessment of the structures discussed above, for anatomic anomalies and disease or the effects of trauma.Page 64 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 62. If the angle appears narrow with the van Herrick procedure (typicallygrade 2) gonioscopy is recommended. If only half or less than half ofthe trabeculae is visible in all quadrants with the goniolens, the eyeshould be considered at risk for closure during dilation. By taking theseprecautions the risk of angle-closure by pupillary dilation is very low.Diabetic retinopathy, particularly severe background, early or establishedproliferative in which the retina is in a state of hypoxia, are particularlyprone to neovascularisation of the angle (rubeosis irides). As with newvessels on the retina, those in the angle tend to be fragile. They mayrupture, with resultant haemorrhaging and consequential formationof fibrotic tissue closing the angle. A similar scenario may occur withbranch or central vein occlusion. Neovascularisation at the angle mayfollow approximately three months after vein closure and hence thepossibility of “90-day glaucoma”. Gonioscopic evaluation of these caseswill allow the optometrist to make a referral at the appropriate levelof urgency.Gonioscopy can give useful information in those cases where thereis historical evidence of angle-closure. This includes symptoms ofintermittent blur, frontal pain or headaches or haloes around lights andhistory of a previous attack of angle closure documented increased intra-ocular pressure from previous base or IOP greater than 21mmHg.The differential diagnosis between primary open-angle and chronicclosed-angle glaucoma can only be made by gonioscopic evaluation ofthe anterior chamber angle.Slit-lamp examination may show evidence of active or past inflammationin the anterior chamber, keratic precipitates, flare or cells in theaqueous, synechiae attachments of the iris to the lens or cornea, ciliaryflush, irregular or poorly responsive pupil. These anomalies can all beinvestigated further with a goniolens. Page 65
  • 63. The plano-convex contact lens mounted in the centre of most goniolenses is considered by many to be the gold standard for examining the posterior pole, as the clarity of view and degree of stereopsis are unsurpassed. The image is very steady because of the damping effect of the contact lens on ocular movement as well as its prevention of lid closure. However, corneal anaesthesia and the application of contact gel make it more time consuming than other methods and in addition the contact gel may interfere with the clarity of view. Clinical classification There are several grading systems which depend upon visibility of anatomical landmarks. The most widely used classification is that of Kolker and Hetherington, in which the widest angles are designated grade 4, through narrow angles to the occluded angle which is graded as zero. It may be more practical in the clinical setting to describe in words the extent of the visibility of the structures seen, as with the Becker-Schaffer grading system. When the angle is 35-45 degrees it can be described as being wide open, incapable of closure with the ciliary body easily visible (grade 4). At 30 degrees the angle is incapable of closure and the scleral spur should be easily identified (grade 3). A 20 degrees angle is described as moderately narrow and the trabeculum is the deepest structure which can be identified (grade 2). Angle closure is possible but unlikely. At 10 degrees, the angle is narrow in which only Schwalbe’s line and perhaps the anterior aspect of the trabeculum can be identified (grade 1). The risk of angle closure is high. If the angle is zero with the iris in contact with the chamber wall, occluding the angle, it is described as closed (grade 0). The van Herrick technique of angle assessment using a slit lamp optic section corresponds well with this classification.Page 66 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 64. Fig 15. A schematic diagram showing Schafer Grading of the angle (grade 4 to Grade 0).Spaeth’s classification is more accurate but also more complicated.Three characteristics need to be noted: • Location of the iris along the angle wall. • The configuration or slope of the peripheral iris, and • The angular approach to the iris recess or the width of the recess.Although this is considered to be the best system, it is not widely used.It seems too cumbersome for most practitioners. Page 67
  • 65. Page 68 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 66. Chapter 4 Slit-lamp biomicroscopyThe modern slit lamp biomicroscope consists of two majorcomponents: the observation (a compound binocular microscope)and the illumination systems. The observation systems of mostmodern instruments have a magnification range between 5 and 40x.Whereas a high magnification would seem desirable, its use oftencompromises the overall image quality (resolution).The illumination system of a modern slit lamp projects a uniform brightimage of a slit onto the plane of focus. The ability of a system to projecta very sharp, very thin and non-distorted slit is crucial to obtaining atrue representation of the ocular structures.The normal white emission of the slit lamp can be modified with theintroduction of filters in the path of the beam. The normal filtersincorporated in an illumination system are: • Diffusing filter. These disperse light over a wide area allowing the observer to illuminate a wide width of the object under investigation. When used in conjunction with low magnification it is particularly useful in giving a good overview of the eye and its adnexa • Cobalt blue filter. This works by altering the emitted light so that the resultant beam is blue. It is most commonly used in conjunction with a fluorescein stain and particularly in contact lens work, where a yellow Wrattan filter may also be used to enhance areas of staining (figure 1). Page 69
  • 67. • Red-Free filter. This filter produces results in the emitted beam being green. The result is that red objects are enhanced, appearing a much darker colour. As a result these filters are frequently used to enhance blood vessels, particularly the limbal arcades (loops) or new vessels that are often very difficult to see particularly when these fine-calibre vessels are devoid of blood. • Heat reducing filters. These absorb light at the red end of the spectrum that are capable of causing thermal damage. These are particularly useful for long examinations as they enhance patient comfort and are strongly recommended for extended fundus assessment. • Neutral density filters. These have largely been replaced in modern instruments by a rheostat control. These filters are a very crude way of controlling the amount of ambient illumination. • Polarising filters. These can be used to reduce the amount of glare and can be useful in techniques such as specular reflection. Fig 1. Use of Wratten filter enhances stainingPage 70 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 68. Set upIt is essential that both the practitioner and patient are comfortablethroughout the examination procedure. The parts of the instrumentin contact with the patient should be sterile for each patient (thisauthor prefers isopropyl alcohol wipes of chin and forehead restsrather than replaceable paper contact sheets) and the patient chaircorrectly elevated such that the patient is not bending or stretchingto maintain forehead contact with the rest. The canthus mark on theheadrest should be in line with the patient outer canthus (though againthis author recommends checking the vertical run of the instrumentruns clear above and below the palpebral area).In normal usage the observation and illumination systems have acommon centre of rotation and it is this that allows sections oftransparent material to be viewed in detail at high magnification. Priorto using the slit lamp (on a daily basis) it is generally recommended thatthe slit lamp eyepieces are focussed. Failure to do so is the most likelyreason why high magnification images are not distinct. This is achievedby placing a focussing rod in the hole of the central pivot. A focussedslit will appear on the rod and the eyepieces are adjusted individuallyso that the grainy appearance of the rod is clearly visible through eacheyepiece. This procedure is performed firstly at low and then highmagnification in order to maximise accuracy.Where a focussing rod is unavailable, the best option then is tofocus eyepieces on as flat an area directly over the hole into whichthe rod is placed. The patient’s forehead may suffice and the correctpositioning of the illumination is known when there is no movementof a slit beam projected onto the area when the angle of theillumination system is altered. Focussing the eyepieces at this point willagain ensure a common centre of rotation.Examination techniquesEach examination technique is as individual as the practitionerutilising it. There can be no prescriptive guidelines on the use ofthis instrument. In the following section a general slit lamp routineis discussed highlighting five recognised techniques are: diffuseillumination, sclerotic scatter, direct illumination, specular reflectionand retro-illumination. There will then be some more specific clinicalexamples of technique specific to different patient presentations. Page 71
  • 69. Diffuse illumination This is generally utilised when an overview of the ocular surface and adnexa is required, but is also useful for photography of structures anterior to the iris when a point reflection wants to be avoided (figure 2). A simple way of performing this technique is to use a broad beam on mid-rheostat setting and scan across the eye in a zigzag pattern. The angle between the observation and illumination systems is kept such that incident light is kept perpendicular to the surface of regard to ensure a good yet even illumination. A mid palpebral sweep from outer to inner canthus would reveal bubbling in tears, obvious bulbar conjunctival and corneal lesions, punctal and inner canthal anomalies. Asking the patient to look upward, the lower lid may be pulled downward while the lower third may be viewed with particular regard for the lower palpebral conjunctiva and the lid margin and lashes. Asking the patient to look down, the upper third sweep may be done again with the lid lifted (and everted if required). Many instruments have an easily introduced diffuser filter which, if used in conjunction with a full rheostat setting, gives optimal diffuse illumination. Older instruments produce this on the reverse of the mirror though reversing the mirror in practice is often cumbersome (as is the sometimes quoted placement of a tissue in front of the mirror to act as a diffuser). Fig 2. Diffuse illuminationPage 72 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 70. Sclerotic scatterThe basic principle of this illumination technique is total internalreflection of light through the cornea. A narrow beam is focussed atthe temporal edge of the limbus at an angle of around 45 to 60º to thecorneal apex.The beam is kept 1 to 2mm and its height is kept between4 and 6mm to avoid producing glare off the sclera. As the light entersthrough the limbal junction total internal reflection occurs within thecornea and emerges at the opposite limbal junction.The resultant image produced by the total internal reflection resemblesa halo around the cornea. If there is a change in the local refractiveindex of the cornea light is scattered from this region. This scatter willcause the image to appear bright against a relatively dull background.Some authorities advocate viewing sclerotic scatter with the naked eyefrom the side, rather than using the microscope, as looking throughthe microscope could result in the viewer looking at the apex of thecornea, while its focus lies in the plane of the limbus. However, at lowmagnification it can be argued that the depth of focus should be lowenough to allow both areas to be viewed. If higher magnification isrequired in this technique then uncoupling of the instrument will berequired. In this case the central cornea should be focussed uponand then the light decoupled and moved to shine at the limbus. Thetechnique shows up areas of corneal compromise as with oedema,which will show as a general glow from the oedematous area, or cornealpenetration, showing as a pinprick of light at the impact point.Direct illuminationAs its name implies this technique allows the clinician to illuminatethe object of interest with the slit beam and observe it undermagnification with the observation system. A complimentary adjunctof this technique is indirect illumination. More often than not thepractitioner utilises these techniques in tandem without being awareof it. With the former high contrast objects are readily visible andwith the latter low contrast entities are often visible. Corneal lesions,for example, are readily visible with direct illumination, but scars maybe more easily viewed with indirect illumination, assuming that thefield of view is large enough. Page 73
  • 71. Corneal and lens sections are best viewed with direct focal illumination (figure 3 and 4). The slit needs to be as thin as possible and the rheostat on maximum to achieve best detail. For corneal sections, the wider the angle between microscope and illumination system, the wider the ‘slice’ of cornea viewed. For the lens, the angle is to some extent dictated by the pupil size. A dilated pupil should allow for a much wider angle to view a larger ‘slice’, while for a miosed pupil the angle of the slit lamp parts may have to be small if a decent section depth is to be achieved. Where as all the layers of the cornea may be visible in section at high magnification slight focussing movements may be required to view its respective layers. However, with the lens it is unlikely that all the layers will ever be visible in one position. In order to maximise the area viewed the separation angle needs to be kept small and low magnification would give a better field of view. The height of the beam is normally restricted to the height of the pupil to allow the maximum amount of light to enter through to the lens (avoiding pupil constriction) and fine focussing movements will allow the different layers of the lens section to be viewed. As the lens is viewed from its anterior to posterior surface the angle of separation is reduced accordingly. It is sometimes desirable to gain a 3D view of the cornea (or lens, although less useful). It is particularly useful when scanning the cornea or lens. This is achieved by widening the beam width once a good section has been obtained. Where as this results in a loss of clarity of the layers, the resultant parallelopiped allows the clinician to form an opinion about the depth of any interesting feature. By then narrowing the beam the observer can then make a more accurate assessment of the depth.Page 74 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 72. Fig 3. Corneal section Lens sectionFig 4.Conical beam and sectionThis useful adjunct of direct illumination is used to check for flareand cells that may be associated with an anterior uveitis. It is not toodissimilar to forming a section but the broader and shorter. Whenthe ambient light is as low as possible (ideally a pitch black room) thecone of light is directed into the anterior segment. A circular apertureis preferred to perform this technique. The beam is focussed in theanterior chamber, which means that in a coupled instrument themicroscope should also focus on the same plane. If the magnificationis kept low enough the entire anterior chamber should be in focus.A bright reflex is seen off the corneal surface and less bright one offthe lens surface. Normally the area between the two reflexes is emptyand so appears dark, but when inflammatory material is present lightis scattered giving this normally optically empty space an appearancesimilar to that seen in the beam of a cinema projector. The scattercaused by the particles is known as Tyndall’s phenomena. It is worthnoting the difference between cells and flare. Flare is dissolved proteinin the aqueous and will be seen as a continuous beam between thecorneal reflex and the lens reflex. Page 75
  • 73. Flare present or not should be noted. Cells are inflammatory cells that glitter as they pass through the beam. These, if present, can be counted. This author prefers to note how many cells are seen in 10 seconds. Also, don’t forget it is useful to look for cells and flare when assessing corneal lesions. Their presence may well be one of the first signs of microbial keratitis in an otherwise innocuous looking corneal lesion. Specular reflection When a corneal section is formed on the same side as the illuminating system a bright reflex is often visible through the observation system. This can often be a hindrance as it tends to be superimposed upon the corneal section. What the clinician is observing are in fact the first and second Purkinje images. The image on the epithelial side tends to be brighter and is a direct reflection of the bright white light of the bulb, where as the image on the endothelial side tends to be a less bright golden colour. Due to the positioning of the illumination system the images are only visible through one of the eyepieces of the observation system. Specular reflection is the preferred choice of illumination for observing the corneal endothelium. In its simplest definition, specular reflection is the bouncing of light of a flat surface to reveal details of the surface. In order to achieve a good view the slit beam is normally shortened and widened. To achieve the best view of the endothelium high magnification should be used, typically in the order of 40x (Figure 5). At such a high magnification slight adjustments to the focussing of the instrument are required. Additionally slight adjustment of the illuminating system may also be required to optimise the image. Fig 5. Specular reflection used to view the endothelium at 40x magnificationPage 76 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 74. Surfaces such as the corneal endothelium are visible by specularreflection as they are not perfectly smooth. Light is reflected at variousangles, which allows the ‘roughness’ of the surface to be assessed.Light that is not reflected back towards the microscope appears darkwhereas that which is appears light.Retro-illuminationViewing certain ocular structures with direct illumination sometimesposes difficulty, either as a result of their proximity or as a resultof their structure. Take for example the limbal vasculature, its closeproximity to the corneo-scleral junction often results in excessiveglare from the sclera when viewed in direct illumination (Figure 6). Indirect illumination is often more beneficial however the relative lackof illumination often make these fine vessels difficult to see. Anotherexample is corneal nerves, which are notoriously difficult to see asthey enter the cornea beyond the limbus. In both examples thesestructures are best viewed by illuminating them from behind ratherthan directly in front.As the name indicates retro-illumination involves illuminating theobject of interest from behind. The reflective nature of some of theocular structures is utilised to create a secondary diffuse sourceof illumination from within the eye. A good example of retroillumination that most practitioners have experiences of is the darkappearance of a cataract seen against the red-reflex generated by aretinoscope. In a slit lamp examination the diffuse reflector is oftenthe iris. It is a popular misconception that when performing retroillumination the slit lamp has to be decoupled. It is only necessarywhen the corneal apex is being viewed, the separation angle is toolarge, high magnification is used and the beam is narrow. For smallseparation angles and areas close to the limbus the cornea (orassociated structures) can be viewed without uncoupling (see figure6). If, however, one wanted to look for microcysts in the centralcorneal area then an area of cornea in front of the pupil margin maybe focussed on. At this point, the illuminated iris part will not bedirectly behind because of the larger distance between the corneaand iris here. The light on the iris may then be carefully moved untilbehind this position by decoupling and then the microcysts be viewed(figure 7). Page 77
  • 75. Fig 6. Retro illumination used to highlight limbal blood vessels Fig 7. Decoupling is only required at high magnification General examination of the eye In the absence of any specific symptoms the slit lamp examination of the eye can be thought of, in the broadest terms, as a two-step procedure. The routine begins with a generalised examination of the eye and then continues using advanced examination techniques, with a more detailed examination of the ocular structures. Although the slit lamp is the principle instrument for examination of the eye in a contact lens wearer for the purpose of this chapter we will assume the eye under observation belongs to a non-lens wearer, although where particularly relevant reference will be made to differences between the two types of eye.Page 78 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 76. In most general examinations it is customary to begin at the front ofthe eye and move deeper-in to examine the finer ocular structures.As such, it is normal to begin with an examination of the eyelids,tear-film and conjunctiva. Observing these structures is generallybest-performed using diffuse and direct illumination together with lowmagnification, but if an object of particular interest is seen, then highermagnification is frequently be utilised.Examination of the eyelidsA large angle between the illumination and observation systems anddiffuse illumination together with low magnification offer the best set-up for a general examination of the surface of the eyelid. This allowsthe clinician to assess the eyelid position, examine for ptosis and checkfor any lid lumps or bumps.Examination of the tear filmThe tear film is arguably the most under examined ocular structureduring a general examination of the ocular surface. Many symptomsof ocular discomfort or asthenopia can be attributed to deficienciesin the tear layer. The tear film is essentially a tri-laminar structureconsisting of: a mucin layer (closest to the cornea), an aqueous layerand a lipid layer (closest to the surface). Anomaly of any of theselayers can result n the patient experiencing symptoms. It is normalfor the tear film to be examined following the instillation of sodiumfluorescein in blue light (produced by the cobalt blue filter), however,with the reduction in the amount of ambient light produced by thistechnique some subtle changes to the tear film may be missed.Diffuse illumination and low magnification offer a good overview ofthe tear film. This allows the clinician to observe any debris in thelayer. Direct illumination or an optical section can then be utilised toinvestigate detail. Debris in the tear film of ten appears as particlesand can be indicative of blepharitisThe tear prism can be seen by direct illumination. This is typically0.2 to 0.5mm high in the centre and tapers off to approximately halfthis height in the periphery. A known height (or width) of slit may besuperimposed over the prism allowing an accurate measurement tobe made. Page 79
  • 77. Instilling fluorescein is an invasive method of assessing the tear film. Prism height measurement may be attempted without fluorescein first. Similarly, break up time before instillation may be attempted by assessing the time taken for a reflection from the tear film to become distorted. Excessive tearing is a hindrance to the patient and is of course clinically relevant. However, insufficient tears tend to give rise to more symptoms and signs. A depleted tear film may give an indication of a tear layer deficiency. Clinicians regularly determine the tear break-up time (TBUT) once fluorescein has been instilled, a procedure familiar to most readers. The eye is illuminated with a broad beam and the cobalt blue filter is in place. Low magnification is used. The patient is asked to blink normally a few times and then ask not to blink. The time taken for dark spots or streaks to appear is noted (figure 8). This indicates a break-up of the tear layer. Normally, this would take 15-20 seconds, any figure below 10 seconds is deemed to be abnormal and indicative of a dry eye. Where the same area consistently breaks-up rapidly, there is likely to be a surface irregularity rather than a dry eye. The clinician should investigate this further using direct illumination, optical section and high magnification. Fig 8. Tear break up timePage 80 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 78. Examination of the conjunctivaAs with the lids and tear film, the best way to obtain a general view ofthe conjunctiva is to use diffuse illumination and a low magnification.Dyes, filters and stains (lissamine green for dry eye states)all helpin the investigation of any anomaly seen in addition direct, focalillumination and higher magnification will allow more accuratediagnosis of the problem.For ease of description between fellow practitioners the conjunctivacan be sub-divided into zones. Although useful, a far simplertechnique is to indicate any areas of interest on a diagram, particularlyas there seems to be a lack of standardisation between the varioussystems that exist. The degree of redness can be determined bycomparing the clinical observations to a pictorial grading scale.Everting the eyelids to expose the conjunctiva in this zone is essentialif follicles or papillae are suspected (figure 9)Fig 9. Eyelid eversion Page 81
  • 79. Specific examination of the anterior segment structures Following a general examination of the ocular surface and adnexa it is desirable to utilise all of the varied illumination techniques to assess some of the finer structures of the anterior segment. These include the cornea, anterior chamber, iris and pupil, lens and the anterior vitreous body. This may be as a part of the routine examination or as a result of specific symptoms reported by the patient. Examination of the cornea For the initial examination of the cornea the illumination system should be set at an angle of 45 to 60º from the observation system and a relatively wide beam should be used in order to create a relatively thick section or parallelopiped. It is possible then to scan the cornea in 3 zones: the superior, middle and lower regions. Using focal, direct (and indirect) illumination allows the clinician to pick up any lesion, which can then be investigated in greater detail using other illumination techniques. Staining the cornea with fluorescein and observing any areas of pooling is also considered an integral part of any routine. If a lesion is found on the cornea there are a number of factors that need to be determined before the appropriate course of management can be initiated. These include its location, size, density, colour and depth. The size of a lesion can be determined using the height of the beam as a measuring device. The best way to observe the depth of a corneal lesion is to observe it in a thin optical section. The following three sections are particularly relevant to corneal examination of the contact lens wearer but are deemed essential to a thorough slit lamp investigation of the cornea. Corneal vascularisation New vessels are often present in the cornea of long-standing soft contact lens wearers. These are usually most in the upper cornea, under the top lid. To the inexperienced microscopist they can be difficult to distinguish from the normal conjunctival overlay of the cornea that exists in this region. New vessels are generally finer and unlike the normal limbal vascular arcades form tight hairpin loops. They are best visualised with an angled beam and a combination of direct, indirect and retro illumination. Using a red-free filter may also enhance the vessel appearance. Ghost vessels (or previously perfused new vessels) are best seen with indirect or retro-illumination.Page 82 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 80. MicrocystsMicrocysts are only visible at magnifications of 25x and greater andare only identifiable at magnifications around 40x. They are bestexamined using a modified technique known as marginal retro-illumination.Corneal endotheliumSetting the slit lamp for specular reflection and utilising the Purkinjeimages is the best way to view the endothelium.Examination of the anterior chamberThere are two principal reasons to examine the anterior chamber:• To assess for the risk of angle-closure glaucoma; and• To assess for any particlesPerhaps the best-known method of assessing the anterior chamberdepth is the Van Herick technique. This is discussed in detail inanother chapter. Other methods of assessment of the anteriorchamber depth do exist. If performed correctly this technique canbe very effective but is not completely without flaw, for example, ananterior chamber can appear shallow using the Van Herick technique,thus leading to suspicions of risk of angle-closure glaucoma, butgonioscopy could reveal that there is little or no risk at all.A typical procedure for performing a Van Herick technique is asfollows. • The patient is asked to look straight ahead towards the microscope and the illumination is set at 60º temporal to the eye under investigation. Setting the angle always at 60º ensures consistency whenever this technique is used and allows results to be compared between observers. • The microscope magnification is kept relatively low to around 10 to 16x thus giving a good depth of focus • The slit width is reduced to a minimum in order to create an optical section of the cornea as close to the limbus as possible. Reducing the slit width normally will involve having to increase the lamp intensity. Page 83
  • 81. • A comparison is made between the thickness of the cornea and the gap between the posterior corneal surface and the iris (figure 10). The ratio of these two may be graded and interpreted. This author prefers a decimal system now as more accurate. So if the black gap is about a half the corneal section thickness, note 0.5. If just under the thickness, note 0.9. Note how, under the original grading system both would have been noted as grade 3. • The measurement is repeated for the nasal limbal area. If there is a large difference between the two measurements the narrower measurement is considered. Fig 10. Van Herick technique Examination of the anterior chamber for inflammatory cells or flare should be made using the conical beam technique described earlier in this chapter. It is worth noting that low-grade flare can be a subtle sign which is easily missed (figure 11). The technique should be carried out in near darkness with all room lights off and the beam should be directed through the pupil to avoid light scatter from the iris. Obtaining a good angle between the illumination and viewing systems may mean moving the viewing system off to one sidePage 84 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 82. Fig 11. Flare and cells in the anterior chamberExamination of the iris and pupilFor the inexperienced clinician the iris is a fascinating area toobserve with a slit lamp biomicroscope. The crypts, pigment ruff andvasculature are all readily visible with direct, focal illumination, with theangle of illumination set between 20 and 30º and a moderate beamwidth and magnification.Examination of the lensThe lens is best observed using a combination of direct, focal andretro-illumination. With the former illumination technique a thinparallelopiped or section reveals the multiple layers of the lens, ratherlike the layers of an onion.Retro-illumination used to investigate the lens can also bedemonstrated with a retinoscope or an ophthalmoscope. The slitlamp technique involves the same principle – observing any defectilluminated from behind by light reflected from the retina. Themicroscope is placed in the straight-ahead position with the lampnearly coaxial (but allowing binocular viewing). The microscope isthen focussed on the object under consideration; the beam is thendecoupled and swung so that it enters the pupil near the margin. Thebeam should be around 4mm in width and the height is adjusted toavoid any unnecessary reflections from the iris. Retro-illumination isparticularly useful in observing cortical cataract. Page 85
  • 83. Examination of the anterior vitreous Examination of the anterior vitreous is difficult as the overlying structures cause scatter thus interfering with the final image produced. Direct focusing of a lens section towards the rear of the lens and then look carefully into the space immediately behind the lens. Asking the patient to move their eye up and then forward again may help in seeing any material here. White ‘wisps’ are generally vitreous material of no concern whereas particulate matter, possibly pigment or cells, needs explanation and may indicate a peripheral retinal tear so requiring referral (Shafer’s sign – figure 12). Fig 12 Shafer’s sign (tobacco dust) Further reading Franklin A (2002). Clinical use of the slit lamp biomicroscope. In Investigative Techniques and Ocular Examination. Pp 33-51 Eds Doshi S. and Harvey W. Butterworth-Heinemann, Optician. Oxford UK. Elliott DB (1997). Clinical procedures in primary eyecare. Pp-177-180. Butterworth-Heinemann,. Oxford UK. Doshi s, Harvey W (2005). Eye Essentials: Assessment & Investigative Techniques. Chapters 4 and 5. Elsevier Science.Page 86 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 84. Page 87
  • 85. Page 88 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 86. Chapter 5 Visual Field Assessment The investigation of visual fields is an essential component of any eye examination as it is able to detect early ocular and neurological disease processes which other investigations may miss. The extent of the absolute visual field is dependent upon the shape of the head of the patient, but it is important to remember that the extent of the temporal field is almost always greater than 90 degrees. Typical values are 100 degrees temporally, 75 degrees inferiorly, and 60 degrees nasally and superiorly.The last two are most patient dependent because of variable nose and brow size. Binocularly, the two monocular visual fields overlap, resulting in a stereoscopic zone which is approximately 120 degrees in the horizontal dimension. The extreme temporal periphery of the binocular field is seen monocularly. (figure 1) The retinal image of the visual field is upside down and back to front. Therefore, the projection of the visual field is such that the superior visual field corresponds to the inferior retina and vice versa. Similarly, the temporal component of the visual field corresponds to the nasal retina and vice versa.Fig 1. Binocular field Page 89
  • 87. Static or kinetic It is well understood that the retinal sensitivity to any target decreases the further away from the fovea the image falls. This was classically described by Traquair as representing a hill of vision whose peak corresponded to the fovea and whose slopes represented the gradual reduction in sensitivity towards the periphery of the retina until ‘the sea of blindness’ is reached; the point beyond which no target presented will be seen. Kinetic perimetry involves the movement of a constant sized target from beyond the field of vision to the point where it is just seen. The larger target should be seen closer to the extent of the absolute field, the smaller target further to the centre of the field. In this way, kinetic perimetry may define the ‘hill of vision’ by the use of variously sized targets moved to the point of identification. Generally, kinetic perimetry is primarily used in optometric practice as a gross assessment of the visual field when a quick test is needed to detect a major scotoma or the cause of a constricted field. Static perimetry employs the use of a target or targets (multiple stimulus or tachistoscopic presentation) of increasing intensity presented to the same point or points within a field until the intensity is reached at which the target is just seen. The majority of automated central visual field screeners employ multiple static stimulus presentation as it is more repeatable, often quicker, and more useful for defining a more subtle scotoma - allowing for the depth of a partial or relative scotoma to be measured. Less controlled, gross variations of static perimetry are also useful, such as some of the confrontation tests used in neurology or the Amsler grid. Confrontation Though the terms confrontation and gross perimetry are often used as synonyms, strictly speaking confrontation describes one of several “comparison” tests whereas gross perimetry is the use of a target to measure the extent of the visual field and to map any large scotomas within the field. One form of confrontation involves the use of a target moved along an imaginary flat plane between and perpendicular to the gaze of the patient and the practitioner.This obviously will not allow the temporal extent of field to be measured, but will allow the practitioner to confirm that any areas they see can also be seen by the patient. One could also describe other tests as confrontation tests, for example, thePage 90 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 88. presentation of two red targets to the hemifields of the patient to findout whether one of the targets is desaturated, or the use of a shiny coinin four quadrants of the field of one eye. Many confrontation tests areused by neurologists in investigating possible neurological lesions.Gross perimetryGross perimetry describes the use of a handheld target held at aconstant distance from the patient’s eye which, when brought in anarc from beyond their visual field boundary, allows them to announcewhen the target is first seen (the extent of field or boundary for thatparticular target size and colour). Continuing the movement of thetarget through the visual field to the central point of fixation allows anylarge defect to be detected.The choice of target dictates the extent of the isopter, just as with anykinetic assessment. A larger target will be seen at a greater eccentricity,while the isopter for a small target will be contracted. Similarly, a whitetarget brought from beyond the field will be seen before a red targetwhich will itself be seen before a green target, the reduction in sensitivitybeing related to the lower luminance of the coloured target comparedto a white target. In practice, although a small white target might bejustifiable in terms of sensitivity, a red target is generally chosen as itwill contrast better with typical wall coverings within the consultingroom.To maintain some degree of correlation with a 5mm white target(typical for gross perimetry), a 15mm diameter red target may beused instead, its extra size counteracting the reduced sensitivity to redcompared to white.The Amsler chartThe Amsler chart represents a simple method of assessing the qualityof the central field of vision. The test includes several charts, the basicand most commonly used being a square white grid printed on a matt-black background comprising twenty rows and twenty columns ofsmaller squares (figure 2). Page 91
  • 89. Fig 2. Amsler recording chart When viewed at 28cm, each square subtends one degree at the retina. The grid therefore is used to assess the field ten degrees either side of the fixation point when viewed monocularly (as should always be the case). It is possible to map out the physiological blind spot by directing the patient to view the nasal edge of the grid such that the blind spot 15 degrees temporal to this point falls onto the grid. Though rarely done, it is (at least in theory) possible to detect blind spot extension in, for example, myelinated nerve fibres around the disc or in early papilloedema. Typically, however, the grid is used to assess macular function. Indications for its use include the following; • Evidence of macular disturbance seen on ophthalmoscopy in either eye (if seen in just one eye, both eyes require investigation because of the bilateral nature of most macular diseases) • Unexplainable loss of central visual acuity • Reduction of acuity through a pinhole • Symptoms of central visual disturbance, such as distortion • History of systemic disease or more commonly drugs which may predispose to a maculopathy (such as tamoxifen or chloroquine). For the assessment of possible early toxic maculopathy, the red Amsler grid may be slightly more sensitivePage 92 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 90. • For the mapping of a central scotoma already detected. This is useful for monitoring any progression of a scotoma. For this assessment, the grid with diagonal cross lines is useful to encourage stable fixation. The patient should be asked to fixate upon the point where they imagine the centre of the cross to be. In cases of poor performance with magnification, noting the position of the scotoma relative to the centre is useful. A scotoma shifted towards the right of the field will have a greater impact upon the ability to scan from left to right when reading. • History of poor photo stress recovery. Reports of persistent after images after exposure to, for example, a flashlight or a difficulty in adapting to changes in ambient light levels may indicate early macular disease.A simple procedure might be as follows; • Correct the patient to adequately see the target at 28cm. Illumination should be good with out constituting a glare source, and one eye should be occluded. • The patient should be directed to fixate upon the dot in the centre of the grid. As poor fixation is perhaps the main source of error in any field assessment, this instruction cannot be repeated enough throughout the test • While looking at the central target, the patient should report if any of the four corners of the grid are missing and, if so, the missing area should be shaded on the record sheet (a replica grid only in black on a white background) • The patient should then report if any of the grid is missing and, if so, the position and size of the blind spot noted • The patient should then report if any of the lines are wavy or distorted and again this recorded if found. If distorted, It is useful to note if the distortion is static (as might be the case with an old atrophic scar or a heavy concentration of drusen around the fovea) or moving or shimmering. The latter, described as metamorphopsia, might be indicative of an active exudative process (such as prior to choroidal neovascularisation) and might warrant an urgent referral. Page 93
  • 91. In cases where one eye has already suffered macular disease, or where there is evidence of macular disturbance yet to affect vision, many practitioners give a copy of the grid to the patient who may then self-monitor there central field on a regular basis at home and in the knowledge that they must report any new disturbance they might detect. Sensitivity versus specificity So far, the assessments considered have been fairly crude. Their ability to detect minor changes in the field of vision are limited (so have a low sensitivity). On the other hand, any defect detected is likely to be actual and related to an abnormal process (so their specificity is high). The balance of the ability of a test to correctly identify disease (sensitivity) while on the other hand ensuring that patients free of disease do not fail the test (specificity) is an important one for fields assessment and the relationship between the two may be changed according to the desired outcome. An bright target should be seen by all normally sighted people and only those with a significant defect will miss it. A test using a bright target would have maximum (100%) specificity. On the other hand, the test would have low sensitivity because even those patients with a mild field defect would still report seeing the light and many early disease presentations would be missed. The highest sensitivity would be gained by using stimuli of an intensity just visible to the point of the retina to which it is presented and which would not be seen if of a lower intensity. This form of presentation is described as a full threshold assessment as it employs stimuli of an intensity set at the threshold of the retinal ability to detect them. This maximum sensitivity should allow even the mildest partial scotoma to be detected. However, many normals might also struggle to see some of the stimuli and report them as missing. This therefore would represent low specificity. Generally speaking, a screening programme should be sensitive enough to detect most eye disease while specific enough for normals to pass the assessment. A suprathreshold assessment uses stimuli of a set intensity brighter than threshold such that all normals should see them and yet defects of a depth greater than the intensity level above threshold should fail the assessment. Generally for screening purposes, a suprathreshold assessment is preferred as it is quicker, sensitive enough to detect most defectsPage 94 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 92. while allowing most normals to pass. For investigation of a suspiciouspatient presentation, such as someone with glaucomatous discappearance or a strong family history of glaucoma, a full thresholdassessment, though more demanding of the patient, is more likelyto detect even the first evidence of reduced retinal function.Suprathreshold fields are also poor at monitoring the progression offield defects, so in secondary care, full threshold fields are preferred.Automated fields screenersMost automated field screeners assess the central field to within25 to 30 degrees from fixation. Whilst it is true that around 85 percent of field defects fall within this space, the remaining minorityaffecting the intermediate and far periphery can be clinicallysignificant. Such defects might result from retinal damage, as withfor example chorioretinitis, retinal detachment, a neoplasm or earlyretinitis pigmentosa. It is also important to remember that manydefects due to damage to the visual pathway may also start in the farperiphery and only extend into the central field at a later stage. Anexample might be the superior bitemporal loss due to compressionof the chiasma from below by a pituitary adenoma (remember abitemporal hemianopia does not just suddenly appear and the veryfirst compression will result in far superior bitemporal loss, oftenasymmetric). A gross perimetry assessment is always worthwhileto support a more accurate automated static central assessment.Alternatively, some screeners, such as the Dicon or the ZeissHumphrey Visual Field Analyser, allow the practitioner to select aprogramme which can assess more peripheral fields as well.Most modern screeners, even suprathreshold screeners such as theHenson, are capable of full threshold testing and can establish thethreshold sensitivity at each point to which a stimulus is presented.The Humphrey Visual Field Analyser may be programmed to assessthe field faster by incorporating the Swedish Interactive ThresholdingAlgorithm (SITA). It was designed as an attempt to decrease the timetaken for a traditional staircase method assessment (a 4-2dB algorithmtypically takes 15 minutes) yet maintaining good sensitivity in glaucomadetection. Test duration is reduced in several ways. The algorithmspecifies the appropriate stimulus brightness to be presented at eachpoint, “guessing” the value from the thresholds already established.It also monitors the time taken to respond to a stimulus as anindication of how close the stimulus is to threshold and is able to Page 95
  • 93. stop once sufficient information has been taken. The presentation of stimuli at levels around that at which fifty per cent are seen greatly reduces the speed of testing. So by better pacing of the test, and the useful interpretation of false positives and false negatives, the SITA appears to correlate well with standard perimetric assessment yet takes less time. The reduced duration is likely to help accuracy of patient response. Longer does not necessarily mean better with field assessment. Poor compliance, fatigue and loss of fixation all play an increasing part in reducing the accuracy of field assessment. This is one of the reasons why it is essential, if a field defect is suspected, for the test to be repeated to ensure that any defect is not artefactual. is able to stop once sufficient information has been taken. The presentation of stimuli at levels around that at which fifty per cent are seen greatly reduces the speed of testing. So by better pacing of the test, and the useful interpretation of false positives and false negatives, the SITA appears to correlate well with standard perimetric assessment yet takes less time. The reduced duration is likely to help accuracy of patient response. Longer does not necessarily mean better with field assessment. Poor compliance, fatigue and loss of fixation all play an increasing part in reducing the accuracy of field assessment. This is one of the reasons why it is essential, if a field defect is suspected, for the test to be repeated to ensure that any defect is not artefactual. Reliability indices As already stated, one of the major problems with fields assessment is the detection of abnormal data that is due to patient ‘error’. Poor fixation or fatigue can lead to points being reported as seen when they should have been missed (a false positive) and not seen when they shuld have been (false negative). Many modern screeners, such as the Humphrey VFA or the Oculus, record these errors such that the reliability of a plot might be interpreted by a practitioner. False positives are recorded when a stimulus which has not been shown or has been presented at an intensity known to be below the threshold value for the individual is still reported as seen. A false negative is recorded when a stimulus known to have been previously seen is now missed. Any significant score in the reliability indices (over 20%) is an indication that the assessment needs to be repeated, as the plot cannot be usefully interpreted and acted upon (figure 2).Page 96 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 94. Fig 3. Reliability indicesA common method of assessing fixation loss is by using the Heijl-Krakau Technique of monitoring fixation throughout a field assessmentby means of projecting a bright stimulus into the blind spot fromtime to time. If fixation is not present, the patient will respond tothis stimulus and the machine will note the number of errors. Thetechnique is open to error as it is only a sampling technique, testingfixation a handful of times during the field test and unless the blindspot is accurately plotted first, the assumed location of the blindspot may be incorrect. Gaze tracking is an alternative which allowsa continual assessment of fixation throughout by monitoring thealignment of the purkinje images from the ocular surfaces. A moreerratic gaze tracking plot is an indication of poor fixation.Total and pattern deviationSome instruments, notably the Humphrey VFA, incorporate softwareallowing analysis of any variation in the assessed field. Any thresholdvalue found to be below that expected in an age-matched populationwill show up as a defect in the total deviation plot. Fogging a twenty-year-old, normal patient will reduce their threshold value across theentire field and this will give an extensive total deviation. A similareffect is seen with cataract.The pattern deviation plot represents significant variation in retinalsensitivity within the field, as might be the case with a scotoma. If apatient has a cataract and also glaucoma, it is possible that they havean extensive total deviation defect while just the earliest signs of anarcuate defect in the pattern deviation plot. In this way, the screenercan help to separate out diseases of causing a true visual fieldanomaly from those causing artefacts, thus increasing its diagnosticvalue. Page 97
  • 95. Fig 4. Total & pattern deviation plots To understand probability plots it is necessary to understand the normal age-related decline in sensitivity which occurs across the visual field. The normal sensitivity of the peripheral visual field declines at a faster rate than centrally, i.e. the hill of vision changes shape with advancing age, becoming steeper in profile. There is a range in sensitivity, which is wider at more peripheral locations. The 95 per cent confidence interval predicts that 95 per cent of normal values fall within this region. Therefore, if a sensitivity value falls outside this range, it will gain a P<5 per cent probability level, i.e. the sensitivity is normal in less than 5 per cent of the normal population. Thus, probability plots graphically illustrate the level of statistical significance associated with a given visual field abnormality compared with the normal reference field. As stated, there are two types of probability plots, total deviation and pattern deviation. In the total deviation plot, the difference between the measured visual field sensitivity and the expected normal visual field sensitivity are plotted as a numeric map. A second plot illustrates those locations where the deviation is significantly different from the normal population, either at the P <0.5 per cent, P <1 per cent, P <2 per cent or P <5 per cent levels. The pattern deviation probability map separates the general reduction in sensitivity that may arise through media opacities, optical defocus or pupillary miosis (the total deviation) from any localised reduction in sensitivity. To calculate the pattern deviation probability map, all locations within 24 degrees of fixation are ranked according to the deviation in sensitivity compared to the age-matched normal population. The general height of the hill of vision is calculated from the measured value of the seventh highest deviation (85th percentile) in sensitivity.Page 98 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 96. The glaucoma hemifield testThe glaucoma hemifield test (GHT) was introduced in the Humphreyfield analyser with the aim of deciding whether visual field loss wascompatible with a diagnosis of glaucoma. Ten anatomical sectors in thevisual field are superimposed on the Program 30-2 test grid, selectedaccording to the normal arrangement of retinal nerve fibres. Fivesectors in the upper hemifield mirror five sectors in the inferior field.Probability scores are calculated for each location within the 10sectors according to the pattern deviation probability map. Withineach sector, the sum of the probability scores is calculated and thedifference compared to the mirror image sector. A visual field isclassed as ‘outside normal limits’ if the difference in any of the fivecorresponding pairs of sectors falls outside the 0.5 per cent or 99.5per cent confidence limits for that pair of sectors. Visual fields areclassified as ‘borderline’ if any sector-pair difference exceeds the 3 percent confidence limit. If the general height of the field is below the 0.5per cent limit, the GHT evaluates the field as a ‘general reduction insensitivity’.A classification of ‘abnormally high sensitivity’ is associated with ahigh level of false-positive responses. The general height test is notperformed if the visual field has already been classified as ‘outsidenormal limits’. A sensitivity and specificity of 80.8 per cent and 81.4per cent respectively has been reported for the GHT. Page 99
  • 97. Interpreting field defects Every field screener produces its own particular plot and interpretation varies depending upon the model. This is covered in much greater detail in the Visual Fields Eye Essentials book by Dr Robert Cubbidge. However, a few general points should be outlined; • Check the field plot reliability indices first. A high number of fixation losses, false positives or false negatives makes the field plot worthless • Any repeatable defect needs to be taken seriously but consider possible artefacts (such as lens rim or eyelid) first. • Overall threshold values between the eyes should be similar. If not, this may indicate reduced light levels reaching the retina, as with media opacification • Single points of reduced sensitivity close to the disc often represent the obscuration of photoreceptors by an overlying retinal blood vessel (an angioscotoma) • Defects found in the area between the blind spot and fixation (centrocoecal defects) usually are associated with damage to the papillomacular bundle of retinal nerve fibres as can occur with toxic responses to drugs or excessive alcohol consumption • Field loss in one eye is due to a lesion anterior to the optic chiasma • Irregular, asymmetric loss in both eyes which does not seem to be limited by the horizontal or vertical midlines is often related to focal retinal damage and can be confirmed by ophthalmoscopy • Progressive optic neuropathy, such as glaucoma, will tend to affect the arcuate fibres initially (usually inferior first as there is a greater susceptibility to damage here) resulting in an arcuate (usually superior) defect. The arcuate fibres extend temporally to the fovea to end at the horizontal raphe. This represents the boundary between the superior and inferior nerve tissues from embryological development. Superior and inferior arcuate loss often ends at different positions at this temporal boundary,Page 100 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 98. and the resultant step in the nasal field is significant evidence ofarcuate fibre damage (figure 5)Fig 5. Nasal step • Any defect clearly respecting the horizontal midline is likely to be retinal in cause. If sudden onset, this is most likely a vascular incident, as with the altitudinal defect in (figure 6)Fig 6. Altitudinal defect • Any defect respecting the vertical midline is likely to be due to damage within the visual pathway from the chiasma or behind Page 101
  • 99. Fig 7. Pituitary adenoma • Heteronymous loss is due to chiasmal damage, either due to pressure on the nasal fibres as with the bitemporal loss due to pituitary adenoma (figure 7), or on the temporal fibres, as with the binasal loss due to an internal carotid aneurysm • Damage to the visual pathway posterior to the chiasma results in homonymous loss. This will be more symmetrical (or congruous) the further back in the pathway the lesion is, increasing significantly in congruence posterior to the lateral geniculate nuclei. • The optic radiations spread through two major lobes of the brain, the parietal and temporal. A common site for damage due to a cerebrovascular accident, damage to the radiations in a patient who survives stroke often leaves a homonymous quadranopic defect • Damage to the occipital lobe may result in a homonymous hemianopia As a significant number of nerve fibres are damaged before detectable field loss results, any repeatable field defect must be treated as important and would usually warrant referral. In the case of a lesion in the visual pathway, this may be an urgent or emergency referral (as with the binasal defect mentioned above).Page 102 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 100. Adapted assessment for early stage glaucomaMuch research has suggested that there is selective nerve fibredamage in glaucoma, with the magnocellular pathway being affectedbefore the parvocellular pathway. Attempts have been made to usestimuli to which the receptors attached to magnocellular ganglioncell fibres are particularly receptive, in an attempt to detect field lossin glaucoma at an earlier stage. Short wave perimetry (or SWAP)is one such attempt (available as an extra on the Humphrey VFAor the Octopus) which uses short wavelength stimuli thought tobe magnocellular responsive. Whilst effective, this technique canbe difficult for the patient to perform and is more susceptible toextraneous factors such as media opacities.Another method employs the presentation to various parts of thefield a sine wave grating that alternates repeatedly such that thespatial frequency is seen to double. The ability to detect this so-calledfrequency doubling is again thought to reduce in early glaucomaand so the Zeiss Humphrey FDT may be used in specific glaucomascreening programmes. Some excitement exists currently at machinessensitive enough to detect minute relative afferent pupil defectsindicative of ganglion cell loss.Useful readingCubbidge R (2005). Eye Essentials; Visual Fields. Elsevier Science Page 103
  • 101. Useful References Practical Viewing of the Optic Disc Butterworth Heinemann 2003, by Kathleen Digre & James Corbett Eye Essentials: Assessment & Investigative Techniques Butterworth Heinemann, 2005, by Sandip Doshi, William Harvey Eye Essentials, Investigative Techniques and Ocular Examination Butterworth Heinemann, 2003, by Sandip Doshi & Bill Harvey College of Optometrists, Guidance on the Re-Use of Contact Lenses and Ophthalmic Devices Clinical Gonioscopy: A Text and Stereoscopic Atlas, Mosby 1972, by Stanley C. Becker New Color Atlas of Gonioscopy, American Academy of Ophthalmology, 2006, by L. Wallace and M. Alward DOCET Clinical Skills DVD, August 2006 I Can See It But What Is It? Common Retinal Conditions Virtual Lecture 2008, Ian Cameron Angle Closure Glaucoma Virtual Lecture 2008, Nick Rumney Interpreting Visual Fields Virtual Lecture 2007, Peter CharlesworthPage 104 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 102. Page 105
  • 103. NotesPage 106 DEVELOPING KEY OPTOMETRIC SKILLS WORKSHOP MANUAL.
  • 104. Page 107
  • 105. Written BySandip Doshi and Bill HarveySandip Doshi PhD MCOptom is an optometrist working in privatepractice in Hove, East Sussex. He is an examiner for the College ofOptometrists and was formerly Clinical Editor of Optician Journal. Heis the co-author of several optical publications including “Eye Essentials:Assessment and Investigative Techniques”.Bill Harvey MCOptom is a visiting clinician and director of the visualimpairment clinic, City University and professional programme tutor forBoots opticians. He is also Clinical Editor of OPTICIAN journal. He isthe co-author of several optical publications including “Eye Essentials:Assessment and Investigative Techniques”. Page 109
  • 106. Knowledge and best practice in this field are constantly changing. Best practice procedures may change as new research and experience broaden our knowledge. Readers are advised to check the most current advice from the professional bodies and manufacturers of the equipment used. It is the practitioner’s responsibility to make diagnoses, and determine treatment for each individual patient and to ensure appropriate safety precautions are adhered to. To the fullest extent of the law, with NHS Education for Scotland, Replay Learning, the editor and the authors accept no liability for any injury or damage to persons or property arising from this publication. Replay Learning Ltd No part of this publication may be reproduced, stored in a retrieval system or tramsmitted in any form by any means, electronic, mechanical, photocopying recording or otherwise, without the prior permission of the Replay Learning Ltd and NHS Education for ScotlandPage 110