  • 1. MODEL: Q-150BLOOD PRESSURE MONITOR Automatic Upper Arm Type 1
  • 2. TABLE OF CONTENTS OF CONTENTS Before using the Monitor 1. Introduction ............................................... 3 2. Safety Information ..................................... 4 3. Device Description .................................... 5 4. Device Function ........................................ 6 5. How to Install the Battery .......................... 7 6. Specifications............................................ 8 7. Applying The Cuff .................................... 9 8. Before Taking a Measurement ............... 11 9. Measure the Blood Pressure .................. 12 10. Setting the Date and Time ...................... 14 11. How to view average value ..................... 15 12. How to Maintain and store the unit ......... 16 13. Limited Warranties .................................. 17 14. Troubleshooting tips................................ 18 15. State of Health ........................................ 192
  • 3. 1. INTRODUCTIONThank you for purchasing the QS HEALTH Q-150Automatic Upper Arm Blood Pressure Monitor.Fill in for future reference. DATE PURCHASED: SERIAL NUIMBER: Staple your purchase receipt here. Register your product on-line athttp://www.qs-health.com/product-registerBlood Pressure Monitor CE ROHS FDA certificate 3
  • 4. 2. SAFETY INFORMATION To assure the correct use of the product, basic safety measures should always beFollowed including the warnings and cautions listed in this ins- truction manual. OPERATING THE DEVICE: Follow the Instructions of your physician or licensed healthcare provider.Self-diagnosis and treatment using measured results may be dangerous. DO NOT adjust medication based on measurement results from this blood pressure monitor take medication asprescribed by your physician. Only a physician is qualified to diagnose and treat High Blood Pressure.The monitor is not intended to be a diagnostic device. Read all of the information in the instruction manual and any other literature in the box before operatingthe unit. This device is intended for use in measuring blood pressure and pulse rate in the adult population. Do not use this device on infants or persons who cannot express their inten- tions.Do not use a cellular phone near the device. It may result in an operational failure.Use only “AA” alkaline batteries with this device. Do not use other types of Batteries This may damage the unit.4
  • 5. 3. DEVICE DESCRIPTION 5
  • 6. 4. DEVICE FUNCTION Automatic determines ideal cuff inflation. 4 -person memory zones, each 99-reading with measuring date/time. Average value for last three records. Auto shut-off in 1 minute, power saving. Clear Memory Function 4 alarm setting for blood pressure measuring or for the medicine time intake 6 Color Light sign to display the State of Health Completely automatic, operates with the push of a button Time and Date setting Selectable between mmHg and Kpa. High & Low pulse rate sign when the measured pulse rate value is too high ortoo low Guide Standards for Assessment of High Blood Pressure Irregular Heart Beating sing Warning heart blinker. A. When the user measures the blood pressure and intend to delete any record. Please take the following step.*press button twice to enter into memory display, then press button around 3 seconds that will show “Clr” on LCD, then press button. B. If the user intend to clean all records, please take the following step. *Press button twice to enter into memory display, then press button around 3 seconds that will show “CLr ” on LCD, then press button, it will show “ ALL ”, then press button.6
  • 7. Pulse FunctionThere are two warning signs (PH and PL) to display with the “Pulse rate” ifthe user’s pulse rate is high or low. The warning sign “PH” means “Pulse”High The warning sign “PL” means “Pulse” Low Battery Indicator:Display 3 level battery capacitive indicator during measurement.After setting, display will show “Kpa” “ON” or “OF”, you can select pressureunit with mmHg or Kpa. ON means Kpa, OF means mmHg.If you selected the Kpa for pressure unit, but you want to return mmHg, youshould move the batteries around 3 minutes, then re-start the device, it willbe revert mmHg or you repeat the above step 1 to 3, you can set it again 5. HOW TO INSTALL BATTERYA. Remove the battery cover in accordance with thedirection of arrow.B. Insert 4 x “AA” batteries in accordance with the indication of polarities on the batteries.C. Slide the battery cover to close. 7
  • 8. 6. SPECIFICATIONS Type: Upper Arm Blood Pressure Monitor 4 Person Memory: 1 ~ 99 with data and time stamp Measurement Method: Oscillometric Method Technology Measurement Range - Pressure: 30 ~ 280 mmHg Pulse Rate: 40 ~ 195 beats / min Accuracy Pressure: within ±3 mmHg Pulse rate: within ±5% of reading (Subject to the temperature, humidity listed below) Inflation: Automatic inflation via pump Selectable between mmHg and Kpa Auto Power off: 1 minute after last key operation Operating Temperature: 10 ~ 40 °C (50 ~ 104 °F) Operating Humidity: 30 ~ 85% R.H. Storage Temperature: -20 ~ 60 °C (-4 ~ 140 °F) Storage Humidity: 10 ~ 95% R.H. Power Supply: 4 x AA batteries Arm Circumference: 22 ~ 32 cm (8.7 ~ 12.6 inch) Item Dimension: approx. 137 (L) x 81 (W) x 42 (H) mm (5.39’ x 3.19’ x 1.65’ inch) Item Weight: approx. 210 g (7.41 oz) NOTE: These specifications are subject to change without notice Accessories included: Upper Arm Blood Pressure Monitor Upper Arm Cuff with Air Tubing User Manual and Guide Spanish & English Protective Leather Pouch 4 alkaline AA batteries8
  • 9. 7. APPLYING THE CUFFCAUTIONRead all of the information in the instruction manual and any other literaturein the box before operating the unit.Do not use this device on infants or persons who cannot express theirintentions.APPLYING THE CUFF ON THE LEFT ARM 1 Make sure the air plug 2 Remove tight-fitting clothing is securely inserted in the from your upper left arm. main unit. 3 Sit on a chair with Place your left arm on a table so your feet flat on the cuff is level with your heart. the floor. 9
  • 10. 4 Hold the thumb grip on the 5 Turn the palm of your left cuff securely with your right hand upward. hand. 6 Apply the cuff to your left upper 7 Wrap the cuff firmly in place arm so the white stripe is on the around your arm using the inside of your arm and aligned clothfastener. with your middlefinger. The air tube runs down the inside of your arm. The bottom of the cuff should be approximately 1/2” aboveyour elbow. APPLYING THE CUFF ON THE RIGHT ARM Apply the cuff to your right upper arm so the thumb grip is centeredon the inside of your inner arm. The bot- tom of the cuff should beApproxima- tely 1/2’’ above your elbow. NOTE: Be careful not to rest your arm onthe air tube. This will restrict the flow of airto the cuff.10
  • 11. 8. BEFORE TAKING A MEASSUREMENTFollow these recommendations1. Avoid eating, drinking alcohol, and smoking, exercising, and bathing for30 minutes before taking ameasurement. Rest for at least 15 minutes beforetaking the measurement.2. Stress raises blood pressure. Avoid taking measurements during stressfultimes.3. The cuff can be applied to your left or right wrist.4. Measurements should be taken in a quiet place.5. Position the unit at heart level throughout the measurement.6. Remain still and do not talk during the measurement.7. Wait 2-3 minutes between measurements you may need to increase thewait time depending on yourindividual physiological characteristics.8. A single measurement does not provide an accurate indication of your trueblood pressure. You need totake and record several readings over a periodof time. Try to measure your blood pressure at the same timeeach day forconsistency. 11
  • 12. 9. MEASURE THE BLOOD PRESURE The American Heart Association1 recommends the following guideline For upper limit of normal for home blood pressure readings: Home Guideline for Upper Limit of Normal BP Systolic Blood Pressure 135 mmHg, Diastolic Blood Pressure 85 mmHg The monitor is not intended to be a diagnostic device. QS HEALTH recommends you contact your physician if the MORNING HYPERTENSION SYMBOL and/or the HEARTBEAT SYMBOL appear on the display with your measurement values. CONNECT THE TUBE OF ARM CUFF TO THE UNIT PROPERLY. 1. To apply the arm cuff. A. Turn the palm of your hand upward. Place mark cloth inside your arm, and match the end of the arm cuff to approximately 1-2 cm from the flexed side of the elbow joint of your arm (nipple level) and wrap the arm cuff on the arm. B. Hold the end of the Arm Cuff and wrap it up by gently pulling so that there is no gap between the arm cuff and the arm. C. Press button to measure blood pressure arm cuff.12
  • 13. 2. To start the measurementNote: In order to get correct measurement results and to avoid unnecessaryerroneous measurement, arm, hand and fingers should be still during mea-surement, also you should be seated and keep silence.A. Press button,the cuff will inflate automatically.B. After main unit detecting Systolic/Diastolic and Pulse Values etc.all of the measured values will be displayed on the LCD panel.C. Release Arm Cuff from your arm while the cuff deflates completely.TO SEE THE PREVIOUS MEASURED VALUES3. The main unit automatically stores the previous measuredvalues.A. Press button, then you can read the average value forlast three records.B. If you press button successively, then you can read theprervious test record for the first persen’s measured.C. Press button, then you can read second, third or forthperson’s measured values.D. Stored capacity is up to 99 sets each page (person).If you press “view” button, you can read the all 99 sets records automaticallyone by one.4. To finish the measurementAfter deflation, the measurement of blood pressure is completed, If you wantto measure again,please press button again, the unit will shut-off then,you should press button again to restart the power. Generally, LCDpanel displays Month/Date/Hour/Minute. 13
  • 14. 10. HOW TO SET TIME AND DATE ONCE YOU INTEND TO SET THE TIME. PLEASE OPERATE AS BELOW: 1. Press button at least 5 seconds, then the YEAR “2008” will appear and blinking. 2. Then press to forward YEAR setting,also you can press to backword YEAR setting. 3 After finishing it, you can press button again. Now you can adjust Month/Date/Hour and Minute as you desired.14
  • 15. HOW TO SER ALARM CLOCKAlarm clock design provides the convenience for user medicine takenorblood pressure measuring. The operation of alarm setting is sameas previoustime and date setting.1. Pressing button at least 5 seconds and wait “ YEAR “ blinking, adjusting Date & Time in correct.2. Then LCD display “ C - - - - - - 0:07”, and “ ON “ , press button again, you can set the first alarm time, If you qant to adjust the time, please press or .3. Keep LCD display in the “ ON ” position ,and press button to continue 2 - 4 alarm setting.4. If you need to stop Alarm setting, you may press or button to let LCD display transfer to the “ OFF “ position. Then the Alarm setting will stop immediately.5. During the setting, if there is any confused, you may simply to stop by pressing button, then this unit will re-display TIME function.6. This unit offers 4- Alarm setting, the Alarm time is 1 minute. 11. HOW TO VIEW AVERAGE VALUEThis unit could offer average value for user reference automatically.The average is decided by user’s latest 3 measurements.1. Press , you can view the value for latest 3 measurements.2. You can press to read the other person’s average values.• If you selected the Kpa for pressure unit, but you want to return mmHg,you should move thebatteries around 3 minutes, then re-start the device, itwill be revert mmHg or you repeat the abovestep 1 to 3, you can set it again. 15
  • 16. 12. How to maintain and store the unit 1. Use a dry, soft cloth to clean the unit, if desired, or use a cloth lightly dampened with tap water. 2. Do not use alcohol, benzene thinner or other harsh chemicals to clean the device or cuff. 3. Do not wash or wet the cuff. 4. Keep away from direct sunlight, high humidity or dust for storage. 5. Do not subject the monitor to strong shocks, such as dropping the unit on the floor. 6. Do not submerge the device or any of the components in water. 7. Changes or modification not approved by the manufacturer will void the user 8. Warranty. Do not disassemble or attempt to repair the unit or components. Note: When the unit will not be used for a long time, be sure to remove the batteries. Otherwise, the batteries may leak damage to the unit.16
  • 17. 13. LIMITED WARRANTIESYour QS HEALTH Q-150 Automatic blood pressure monitor , iswarranted tobe free from defects in materials and workmanshipappearing within 2 yearsfrom the date of purchase, when used inaccordance with the instructionsprovided with the monitor.Theabove warranties extend only to the original retail purchaser.To obtainwarranty service contact Customer Service by calling1-786) 519 1206 forthe address of return shipping and handling fee.Enclose the Proof of Purcha-se. Include a letter, with your name,address, phone number, and descriptionof the specific problem.QS HEALTH SHALL NOT BE LIABLE FOR LOSS OF USE OR ANYOTHER SPE-CIAL, INCIDENTAL, CONSEQUENTAIL OR INDIRECTCOSTS, EXPENSES ORDAMAGES.FOR CUSTOMER SERVICEVisit our web site at:http://www.qs-health.com/technical-supportCall 1-786-519-1206 17
  • 18. 14. TROUBLESHOOTING TIPS Phenomenon Problems Solutions Nothing is 1.Batteries exhausted. Replace the batteries displayed when 2.The polarities of with new ones. you press the batteries in the wrong Power position “ Err P ” shown Fail to inflate 1.Please check the connection on on LCD display main unit and cuff. If loosened, then connect them tightly. 2.Please check if air tube isbroken, then change it “ Err 1 ”shown 1.The cuff is loosened. 1.Fasten the cuff, stant it again. on LCD display 2.Deflate too fast 2.Please send it to service station “ Err 2 ”shown 1.Movements occur on 1.Keep still after you press on LCD display the wrist. Power button. 2.Interference signal 2.Please remove interfering sources around the BPM. from Magnetic material or cell phone “ Err 3 ”shown Incorrect measured 1.Take off your watch or cuff on LCD display result might be placed incorrectly.Please re-wrap cuff in correct position. 2.You might use it incorrectly, such as measure blood pressure on child. • If correct measurement is impossible even after checking the above mentioned points, consult at the store where you have purchased the unit or the nearest dealer without touching the internal mechanism. • In some very rare cases, there may be error due to the physical condition of the person. In such cases, please consult your doctor. • If LCD display doesn’t show anything, you may release Battery and install it again.18
  • 19. 15. STATE OF HEALTHStandards for assessment of high blood pressure, without regard tobage,have been established by theWorld health Organization (WHO), as shown inthe chart below. It will help users to find out their healthsituation. Differentlaves shows different color lights. Optimal Normal High-normal Grade 1 hypertension Grade 2 hypertension Grade 3 hypertension 19